EGAmaster 51251 User manual

INSULATION RESISTANCE
MULTIMETER
OPERATING INSTRUCTIONS
GUARANTEE...........................23
COD. 51251

2
ENGLISH
INSULATION RESISTANCE MULTIMETERS
INTRODUCTION
The code 51251 is a battery-powered, true RMS insulation multimeter (hereafter “the Meter”) with 6000
count display.
THE METER MEASURES OR TESTS THE FOLLOWING:
• AC/DCvoltageandcurrent
• Resistance
• Voltageandcurrentfrequency
• Diode
• Continuity
• Capacitance
• Temperature
• InsulationTesting
SAFETY INFORMATION
ThisMetermeetsCATIII1000VandCATIV600VIEC61010standards:inpollutiondegree2anddouble
injection.
CATIII:Distributionlevel,xedinstallation,withsmallertransientovervoltagesthanCATIV.
CATIV:Primarysupplylevel,overheadlines,cablesystemsetc.
UsetheMeteronlyasspeciedinthisusersmanual.OtherwiseprotectionprovidedbytheMetermaybe
impaired. See Table 1 for a list of symbols used on the Meter and in this manual.
Warningidentieshazardousconditionsandactionsthatcouldcausebodilyharmordeath.
CautionidentiesconditionsandactionsthatcoulddamagetheMeter,theequipmentundertest,orcause
permanent loss of data.
Warning
Toavoidpossibleelectricshockorpersonalinquiry,followtheseguidelines:
• UsetheMeteronlyasspeciedinthismanualortheprotectionprovidedbytheMetermightbe
impaired.
• DonotusetheMeterortestleadsiftheyappeardamaged,oriftheMeterisnotoperatingproperly.If
indoubt,havetheMeterserviced.
• Alwaysusetheproperterminal,switchposition,andrangeformeasurementsbeforeconnecting
Meter to circuit under test.
• VerifytheMeter’soperationbymeasuringaknownvoltage.

3
• DonotapplymorethantheratedvoltageasmarkedontheMeter,betweentheterminalsorbetween
anyterminalandearthground.
• Usecautionwithvoltagesabove33Vacrms,46.7VacPeakor70Vdc.Thesevoltagesposeashock
hazard.
• Replacethebatteryassoonasthelowbatteryindicatory( ) appears.
• Disconnectcircuitpoweranddischargeallhigh-voltagecapacitorsbeforetestingresistance,
continuity, diodes, or capacitance.
• DonotusetheMeteraroundexplosivegasorvapor.
• Whenusingthetestleads,keepyourngersbehindthengerguards.
• Inordertocomplywiththesafetystandard,theMetermustbeusedalongwiththeincludedtestleads.
Anydamagesofthetestlead,mustreplaceitwiththesamemodelnumberandspecication.
• RemovetestleadsfromtheMeterbeforeopeningtheMetercaseorbatterydoor.Neveroperatethe
Meterwiththecoverremovedorthebatterydooropen.
• Complywithlocalandnationalsafetyrequirementswhenworkinginhazardouslocations.
• UndertheinuenceofRadio-FrequencyElectromagneticField&ConductedRadio-Frequency
ElectromagneticFieldphenomenon.Thecaptionedmodelhaveamagnicenterrorinfunctionmeasurement.
Itwillbebacktonormalwhentheinterferenceisremoved.
• Useproperprotectiveequipment,asrequiredbylocalornationalauthoritieswhenworkingin
hazardousareas.
• Avoidworkingalone.
• Useonlythereplacementfusespeciedortheprotectionmaybeimpaired.
• Checkthetestleadsforcontinuitybeforeuse.Donotuseifthereadingsarehighornoisy.
• Donotusecurrentterminalsorcurrentrangetotestvoltage.
TABLE 1. SYMBOLS
DoubleInsulated EarthGround
AC(AlternatingCurrent) DC(DirectCurrent)
Diode Battery (Low battery when shown on
display)
ContinuityBuzzer ACorDC(AlternatingCurrentor
DirectCurrent)
ConformstoStandardofEuropeanUnion
UNPACKING INSPECTION
OpenthepackagecaseandtakeouttheMeter.Checkthefollowingitemscarefullytoseeanymissingor
damagedpart:
Table 2. Unpacking Inspection
Users Manual 1 piece
Test Lead 1 pair
Short Lead 1 pair
Alligator Clip 1 pair
K type thermocouple 1 piece
Batteries AA (AM3/LR6) 6 pieces
Intheeventyoundanymissingordamage,pleasecontactyourdealerimmediately.

4
THE METER STRUCTURE
Figure 1. The Meter Structure
1. FrontHousing
2. LCDDisplay
3. FunctionalButtons
4. RotarySwitch
5. InputTerminals
ROTARY SWITCH POSITIONS
TurntheMeteronbypressingthePOWER button. Use the BLUE button to select any rotary switch
alternatefunctions(labeledwithblueletters).RotaryswitchselectionsareshowninFigure2anddescribedin
Table3.
Figure 2. Rotary Switch

5
Table 3. Rotary Switch Selections
Switch Position Measurement Function
ACvoltagefrom60mVto1000V
DCvoltage1mVto1000V
DCmV0.01mvto600mV
code 51251 Temperaturefrom-40°Cto+537°C(-40°Cto+998°F).Celsiusisthedefault
temperature measurement unit
Ohmsfrom0.1to40M.
Capacitancefrom0.01nFto100µF
Continiutytest.Beeperturnsonat<30
Diodetest.Thereisnoranginginthisfunction
DCmAfrom0.01mAto600mA(600mAoverloadfor2minutesmaximum)
ACmAfrom3mAto600mA(600mAoverloadfor2minutesmaximum)
insulation Presstoperforminsulationtest
From0.1Mto2G.Select50V,100V,250V,500Vand1000Vtestingvoltage
BUTTONS
Usethebuttonstoactivatefeaturesthataugmentthefunctionselectedwiththerotaryswitch.Thebuttons
areshowninFigure3describedinTable4.
Figure 3. Buttons
Table 4. Buttons
Button Description
POWER PressPOWER to turn on the Meter
PressandholdPOWERtoturnotheMeter
HOLD
IninsulationTestmode,thisschedulesatestlockthenexttimeyoupressTESTonthe
Meter.ThetestlockactstoholddownthebuttonuntilyoupressHOLDorTESTagain
toreleasethelock.
Inordermeasurementmode:Presstofreezethedisplayedvalue.
Pressagaintoreleasethedisplay.
Pressandholdthebuttontoclearthestoreddata.
Pressthebuttontoturnthebacklightonando.
SAVE
Pressthisbuttontostorethesinglerecord.
Pressandholdthisbuttontostoredatacontinuously.PressSAVEagaintoexit
continuousdatastoringmode.
Themaximumnumberofdatastoreis99sets.TheMeterdisplaysFULLwhenitreaches
themaximum.PressCLRtoclearthestoreddatainordertostorenextsetofdata

6
LOAD
• Pressthisbuttononcetorecallthestoredvalue.Pressortorecallthepreviousornext
storedvalue.PressLOADagaintoexitdatarecallmode.
• Pressandholdthisbuttontorecallthestoreddatacontinuously.PressLOADagainto
exit.
• AfterenteringtheLoadmode,itisnotallowedtoturntherotaryswitchtoenterto
measuringmode.YoucanonlyenterthemeasuringmodeafterexitingtheLoadmode.
InMeasurementmode,exceptbuzzer,diodeandinsulationtest,presstostartretaining
maximumvalue.
InLoad,eachpresstogobacktothepreviousstoredreading
InMeasurementmode,exceptbuzzer,diodeandinsulationtest,presstostartretaining
minimumvalue.
InLoad,eachpresstorecallthenextstoredreading.
RANGE
InInsulationTestmode,presstoselectdierenttestingvoltage.
Inothermode,exceptbuzzeranddiode,presstochangeRangingmodefromAutoto
ManualRangingmode.PressandholdtoreturntoAutoRangingmode.
SELECT Functionsasashiftkey.Presstoaccessorangefunctionsontherotaryswitch.
InitiatesaninsulationtestwhentherotaryswitchisontheINSULATIONposition.Causes
theMetertosource(output)ahighvoltageandmeasureinsulationresistance.
InMeasurementmode,activatefrequencymeasurement.
UNDERSTANDING THE DISPLAY
DisplayindicatorsareshowninFigure4anddescribedinTable5.
Figure 4. Display indicators
Table 5. Display Indicators
Indicator Description
MAX Indicatesmaximumreadinghasbeenselected.
MIN Indicatesminimumreadinghasbeenselected.
H Indicatesdisplayholdisactive
Diodetestfunctionisselected
DCAC AC/DCVoltageorCurrentmeasurement
trms True RMS measurement
Continuitytestfunctionisselected
Lowbattery.Indicateswhenitistimetoreplacethe
battery.
Warning
Toavoidfalsereadings,whichcouldleadtopossible
electricshockorpersonalinjury,replacethe
battery as soon as the low battery indicator
appears

7
Highvoltagewarning
AutoRange Autorangingmode
ManualRange Manualrangingmode
61000mV Displayrangeinuse
-Negativesymbol
>Greaterthansymbol
Primarydisplay
MEM Datastoresymbol
MR Datarecallsymbol
The number of stored data
FULL
Whenthenumberofstoreddatareaches99,the
displaywillshowFULL.
Whenthenumberofstoreddataisfull,pressCLR
toclearupthestoreddatabeforestoringanother
set of data.
CLR Clearupthestoreddata
Indicatesatestlockwillbeappliedthenexttime
youpressTESTontheMeter.Thetestlockacts
toholddownthebuttonuntilyoupressHOLDor
TESTagain.
2500V
1000V
Sourcevoltageratingforinsulationtest.
50V,100V,250V,500Vand1000Vrangesavailable.
SecondaryDisplay
Measurement units
MAKING BASIC MEASUREMENTS
MEASURING AC VOLTAGE (see gure 5)
Figure 5. Measuring AC Voltage

8
Warning
To avoid harm to you or damages to the Meter from electric shock, please do not attempt to
measure voltages higher than 1000V although readings may be obtained.
TheACVoltagerangesare:6.000V,60.00V,600.0Vand1000V.Ifovertherange,itdisplays“OL”.To
measureACvoltage,connecttheMeterasfollows:
1. InserttheredtestleadintotheVΩ terminalandtheblacktestleadintotheCOMterminal.
2. Set the rotary switch to
3. Connectthetestleadsacrosswiththeobjectbeingmeasured.Themeasuredvalueshowsonthedisplay.
4. ThedisplayedvalueisTrueRMS.
Note:
• Ifthevalueofvoltagetobemeasuredisunknown,settheMetertoautorangingorusethemaximum
measurementposition1000Vandreducetherangestepbystepuntilasatisfactoryreadingisobtained.
• Ineachrange,theMeterhasaninputimpedanceof10MΩ.Thisloadingeectcancause
measurementerrorsinhighimpedancecircuits.Ifthecircuitimpedanceislessthanorequalto10kΩ,the
errorisnegligible(0.1%orless).
• WhenDCvoltagemeasurementhasbeencompleted,disconnecttheconnectionbetweenthetesting
leadsandthecircuitundertest,andremovethetestingleadsawayfomtheinputterminalsoftheMeter.
MEASURING DC VOLTAGE (see gure 6)
Figure 6.Measuring DC Voltage
Warning
To avoid harm to you or damages to the Meter from electric shock, please do not attempt to
measure voltages higher than 1000V although readings may be obtained.
TheDCVoltagerangesare:6.000V,60.00V,600.0Vand1000V.Ifovertherange,itdisplays”OL”.To
measureDCvoltage,connecttheMeterasfollows:
1. Inserttheredtestleadintothe terminalandtheblacktestleadintotheCOM terminal.
2. Set the rotary switch to
3. Connectthetestleadsacrosswiththeobjectbeingmeasured.Themeasuredvalueshowsonthe
display.
Note:
• Ifthevalueofvoltagetobemeasuredisunknown,settheMetertoautorangingorusethemaximum
measurementposition1000Vandreducetherangestepbystepuntilasatisfactoryreadingisobtained.

9
• Ineachrange,theMeterhasaninputimpedanceof10MΩ.Thisloadingeectcancause
measurementerrorsinhighimpedancecircuits.Ifthecircuitimpedanceislessthanorequalto10kΩ,the
errorisnegligible(0.1%orless).
• WhenDCvoltagemeasurementhasbeencompleted,disconnecttheconnectionbetweenthetesting
leadsandthecircuitundertest,andremovethetestingleadsawayfromtheinputterminalsoftheMeter.
MEASURING DC MILLIVOLTS (see gure 7)
Figure 7.Measuring DC Millivolts
Warning
To avoid harm to you or damages to the Meter from electric shock, please do not attempt to
measure voltages higher than 500V, otherwise you cannot obtain correct reading.
TheDCMillivoltsrangesare:60.000mVand600.0mV.Ifovertherange,itdisplays“OL”.TomeasureDC
Millivolts,connecttheMeterasfollows:
1. Inserttheredtestleadintothe terminalandtheblacktestleadintotheCOMterminal.
2. Set the rotary switch to
3. Connectthetestleadsacrosswiththeobjectbeingmeasured.
Themeasuredvalueshowsonthedisplay.
Note:
• Themaximuminputimpedanceis4000MΩ.
• WhenDCMillivoltsmeasurementhasbeencompleted,disconnecttheconnectionbetweenthetesting
leadsandthecircuitundertest,andremovethetestingleads

10
MEASURING RESISTANCE AND CAPACITANCE
MEASURING RESISTANCE (see gure 8)
Figure 8. Measuring Resistance
Warning
To avoid electric shock, injury, or damage to the Meter, disconnect circuit power and discharge all
high-voltage capacitors before testing resistance.
TheResistancerangesare600.0Ω,6.000kΩ,60.00kΩ,600.0kΩ,6.000MΩand40.00MΩ
Tomeasureresistance,connecttheMeterasfollows:
1. Inserttheredtestleadintothe terminalandtheblacktestleadintotheCOM terminal.
2. Set the rotary switch to Resistance measurement is default or press SELECTbuttontoselectΩ
measurement mode
3. Connectthetestleadsacrosswiththeobjectbeingmeasured.Themeasuredvalueshowsonthe
display.
Note:
• WhenResistancemeasurementhasbeencompleted,disconnecttheconnectionbetweenthetesting
leadsandthecircuitundertest,andremovethetestingleadsawayfromtheinputterminalsoftheMeter.
MEASURING CAPACITANCE (see gure 9)
Figure 9. Measuring Capacitance
Table of contents
Other EGAmaster Measuring Instrument manuals

EGAmaster
EGAmaster 57542 User manual

EGAmaster
EGAmaster 79502 User manual

EGAmaster
EGAmaster 51266 Series User manual

EGAmaster
EGAmaster 57666 User manual

EGAmaster
EGAmaster EGAtronik 51265 User manual

EGAmaster
EGAmaster COD.66265 User manual

EGAmaster
EGAmaster 65189 User manual

EGAmaster
EGAmaster 51248 User manual

EGAmaster
EGAmaster EGATRONIK User manual

EGAmaster
EGAmaster EGAtronik 51719 User manual