6
FEATURES:
The DP-35 differential voltage probe provides a safety means of
measuring floating potentials for all models of oscilloscopes incomplete
safety.
It converts the high differential voltage (≦1600 peak) into a low
voltage (≦8.0 V) with reference to the earth for display on the
oscilloscopes.
The BNC output is designed to operate on an input with an impedance
of 1 MΩ. It is 2 times of the 50 Ω.
Recommend to use PINTEK PL-10 with DP-35, and expand the
measuring ranges. From DMM can observe more exact measured
testing voltage. (Oscilloscope accuracy is 3%, and DMM is 10 times).
DP-35 is a design for high sensitivity module and high dynamic range.
Attenuation x10, x100 is multiple of 10, which is easy for calculation
without mistake. Maximum voltage is 1600Vp-p. It is for high sensitivity
as well as for high voltage.
SPECIFICATIONS:
(1) Bandwidth: DC - to 35 MHz
(2) Attenuation: x 10, or x 100
(3) Accuracy: 2%
(4) Voltage Input Ranges (DC + AC peak to peak):
≦160 Vp-p for x 10, (i.e about 56 V RMS or 80 V DC)
≦1600 Vp-p for x 100, (i.e about 560 V RMS or 800 V DC)
(5) Permitted Max Input Voltage
Max differential voltage: 1600V (DC + AC peak to peak)
Max voltage between each input terminal and ground: 600 V RMS
(6) Input Impedance:
Differential: 8 MΩ// 2.5 pF
Between terminals and ground: 4 MΩ// 5 pF
(7) Output: ≦8.0 V