FEATURES:
The DP-15K differential voltage probe provides a safety means of
measuring floating potentials for all models of oscilloscopes incomplete
safety.
It converts the high differential voltage (≦15 KV peak) into a low voltage
(≦8.0 V) with reference to the earth for display on the oscilloscopes.
The MAX. bandwidth is 75MHz, is suitable for big power testing and
research using.
The BNC output is designed to operate on an input with an impedance
of 1 MΩ. It is 2 times of the 50 Ω.
DP-15K is able to connect to DMM to observe directly. It is able to get
more accurate testing real value. (Oscilloscope accuracy is 3%, DMM
is about 10 times accuracy).
DP-15K is a design for high sensitivity module and high dynamic range.
Attenuation x 100, x 1000 is multiple of 10, which is easy for calculation
without mistake. Maximum voltage is 15KVp-p. It is a model designed
for ultra high voltage.
SPECIFICATIONS:
(1) Bandwidth:
DC - to 75 MHz (-3 dB) for x 100
DC - to 20 MHz (for attenuation x 1000)
(2) Attenuation: x 100, or x 1000
(3) Accuracy: ±2%
(4) Voltage Input Ranges (DC + AC peak to peak):
≦1.5 KVp-p for x 100, (i.e about 550 V RMS or ±750 V DC)
≦15 KVp-p for x 1000, (i.e about 5.5 KV RMS or ±7.5 KV DC)
(5) Permitted Max Input Voltage
Max differential voltage: 15 KV (DC + AC peak to peak)
Max voltage between each input terminal and ground: 7.5 KV RMS