FEATURES:
The DP-16K series differential voltage probes provides a safety means
of measuring floating potentials for all models of oscilloscopes
incomplete safety.
It converts the high differential voltage (≦16 KVDC peak) into a low
voltage (≦8.0 V) with reference to the earth for display on the
oscilloscopes.
The BNC output is designed to operate on an input with an impedance
of 1 MΩ. It is 2 times of the 50 Ω.
Recommend to use PINTEK PL-10 with DP-16K series, and expand
the measuring ranges. From DMM can observe more exact measured
testing voltage. (Oscilloscope accuracy is 3%, and DMM is 10 times).
DP-16K series are designed for high sensitivity modules and high
dynamic range. Attenuation x100, x1000 are multiple of 10, which is
easy for calculation without mistake. Maximum voltage is 16 KVp-p.
These are designed for ultra high voltage.
SPECIFICATIONS:
(1) Bandwidth:
DC to 100 MHz (-3 dB) (DP-16K)
DC to 70 MHz (-3 dB) (DP-16KM)
DC to 50 MHz (-3 dB) (DP-16KL)
(2) Attenuation: x100, or x1000
(3) Accuracy: ±2%
(4) Voltage Input Ranges (DC + AC peak to peak):
≦1.6 KVp-p for x100, (i.e about 565 VRMS or ±800 VDC)
≦16 KVp-p for x1000, (i.e about 5650 VRMS or ±8 KVDC)
(5) Permitted Max Input Voltage
Max differential voltage: 16 KV (DC + AC peak to peak)
Max voltage between each input terminal and ground: 3500 VRMS
(6) Input Impedance:
10 MΩ// 10 pF each side to ground