WIKA 632.50 User manual

DE
EN
Operating instructions
Betriebsanleitung
Example: Model 632.50.100 per ATEX
Capsule pressure gauge, stainless steel
Models 632.50 and 633.50 per ATEX
Kapselfedermanometer, CrNi-Stahl
Typen 632.50 und 633.50 nach ATEX

DE
EN
40433177.01 11/2020 EN/DE
Operating instructions capsule pressure gauge, model 63x.50 per ATEX
2
© 11/2020 WIKA Alexander Wiegand SE & Co. KG
All rights reserved. / Alle Rechte vorbehalten.
WIKA®is a registered trademark in various countries.
WIKA®ist eine geschützte Marke in verschiedenen Ländern.
Prior to starting any work, read the operating instructions!
Keep for later use!
Vor Beginn aller Arbeiten Betriebsanleitung lesen!
Zum späteren Gebrauch aufbewahren!
Operating instructions
models 632.50 and 633.50 per ATEX
Page 3 - 28
Betriebsanleitung
Typen 632.50 und 633.50 nach ATEX
Seite 29 - 53

EN
Operating instructions capsule pressure gauge, model 63x.50 per ATEX 3
40433177.01 11/2020 EN/DE
Contents
Declarations of conformity can be found online at www.wika.com.
1. General information 4
1.1 Explanation of symbols ................................................................. 5
2. Safety 6
2.1 Intended use ................................................................................. 6
2.2 Responsibility of the operator........................................................ 7
2.3 Personnel qualification .................................................................. 9
2.4 Safety instructions for hazardous locations ................................. 10
2.5 Labelling / Safety marks .............................................................. 15
2.6 Ignition hazard analysis............................................................... 16
3. Specications 17
4. Design and function 18
5. Transport, packaging and storage 18
5.1 Transport ..................................................................................... 18
5.2 Packaging and storage................................................................ 19
6. Commissioning, operation 19
6.1 Mechanical connection ............................................................... 19
6.2 Requirements for the installation point ........................................ 20
6.3 Installation................................................................................... 21
6.4 External zero point setting (if available)....................................... 22
6.5 Permissible ambient and operating temperatures ....................... 22
6.6 Permissible vibration load at the installation site ......................... 23
6.7 Level check ................................................................................. 23
6.8 Commissioning ........................................................................... 23
7. Faults 24
8. Maintenance and cleaning 26
8.1 Maintenance ............................................................................... 26
8.2 Cleaning...................................................................................... 26
9. Dismounting,returnanddisposal 27
9.1 Dismounting................................................................................ 27
9.2 Return ......................................................................................... 27
9.3 Disposal ...................................................................................... 27
Annex: EU Declaration of conformity 54

EN
Operating instructions capsule pressure gauge, model 63x.50 per ATEX
4
40433177.01 11/2020 EN/DE
1. General information
The pressure gauge described in the operating instructions has been
designed and manufactured using state-of-the-art technology.
All components are subject to stringent quality and environmental
criteria during production. Our management systems are certied to
ISO 9001 and ISO 14001.
These operating instructions contain important information on
handling the instrument. Working safely requires that all safety
instructions and work instructions are observed.
Observe the relevant local accident prevention regulations and gener-
al safety regulations for the instrument's range of use.
The operating instructions are part of the product and must be kept
in the immediate vicinity of the instrument and readily accessible to
skilled personnel at any time.
Skilled personnel must have carefully read and understood the
operating instructions prior to beginning any work.
The manufacturer's liability is void in the case of any damage caused
by using the product contrary to its intended use, non-compliance
with these operating instructions, assignment of insuciently qualied
skilled personnel or unauthorised modications to the instrument.
The general terms and conditions contained in the sales documenta-
tion shall apply.
Subject to technical modications.
1. General information

EN
Operating instructions capsule pressure gauge, model 63x.50 per ATEX 5
40433177.01 11/2020 EN/DE
Further information:
- Internet address: www.wika.de / www.wika.com
Model NS Data sheet
632.50, 633.50 63, 100, 160 PM 06.03
1.1 Explanation of symbols
WARNING!
... indicates a potentially dangerous situation that can result in
serious injury or death, if not avoided.
Information
... points out useful tips, recommendations and information
for ecient and trouble-free operation.
WARNING!
... indicates a potentially dangerous situation in the hazard-
ous area that results in serious injury or death, if not avoided.
1. General information

EN
Operating instructions capsule pressure gauge, model 63x.50 per ATEX
6
40433177.01 11/2020 EN/DE
2. Safety
WARNING!
Before installation, commissioning and operation, ensure
that the appropriate pressure gauge has been selected in
terms of measuring range, design and specic measuring
conditions.
Check the compatibility with the medium of the materials
subjected to pressure!
In order to guarantee the measurement accuracy and
long-term stability specied, the corresponding load limits
must be observed.
Non-observance can result in serious injury and/or damage
to property.
Further important safety instructions can be found in the
individual chapters of these operating instructions.
2.1 Intended use
These pressure gauges are used for measuring pressure in hazardous
areas of industrial applications.
ClassicationperEuropeanpressureequipmentdirective
Instrument type: Pressure accessory without safety function
Media: Gaseous, group 1 (hazardous)
Maximum permissible pressure PS, see chapter 2.5 “Labelling /
safety marks”
Volume of wetted parts: < 0.1 L
2. Safety

EN
Operating instructions capsule pressure gauge, model 63x.50 per ATEX 7
40433177.01 11/2020 EN/DE
2. Safety
The instrument must only be used with media which are not harmful
to the wetted parts over the entire operating range of the instrument.
Any change in the state of the matter or any decomposition of unstable
media is not permitted.
Only use the instrument in applications that lie within its technical perfor-
mance limits (e.g. max. ambient temperature, material compatibility, ...).
→ For performance limits see chapter 9 “Specications”.
Suitability for use
Application
For gaseous, dry and aggressive media, that are not highly viscous or crystallising, also
in aggressive environments
Process industry: Chemical industry, petrochemical industry, oil and gas, power genera-
tion, water and wastewater technology, machine building and general plant construction
High dynamic pressure loads and vibrations (only with optional case liquid lling)
The instrument has been designed and built solely for the intended use
described here, and may only be used accordingly.
The manufacturer shall not be liable for claims of any type based on
operation contrary to the intended use.
2.2 Responsibility of the operator
The legibility of the marking must be observed during time in use but at
least during inspection periods of three years. If any harm of the legibility
is found please contact the manufacturer to renew the marking.
For the safety of the system, the operator is obliged to carry out an
ignition source analysis. The responsibility for classication of zones
lies with the plant manager and not the manufacturer/supplier of the
equipment.

EN
Operating instructions capsule pressure gauge, model 63x.50 per ATEX
8
40433177.01 11/2020 EN/DE
2. Safety
These ignition sources must be taken into account for the instrument:
1. Hot surfaces
The surface of the instrument can heat up due to the temperature of
the process medium.This depends on the installation situation and
must be taken into account by the operator.
2. Mechanically generated sparks
Mechanically generated sparks are a potential ignition source. If the
materials used exceed a total mass percentage of 7.5 % magnesium,
titanium and zirconium, the operator must take appropriate protective
measures.
3. Static electricity
- To avoid electrostatic charging, the instrument must be included in
the equipotential bonding of the system.This can be done via the
process connection or other suitable measures.
- The instrument can optionally contain components with a non-con-
ductive surface coating or lining. In such cases, the operator must
take appropriate measures to prevent electrostatic charging.
- Metal components of the instruments (e.g.TAG plates) must be
included in the equipotential bonding of the system during installa-
tion and operation.
4. Adiabaticcompressionandshockwaves
With gaseous media, the temperature may increase as a result of
compression warming. In these cases it may be necessary to throttle
the rate of change of pressure or reduce the permissible medium
temperature.

EN
Operating instructions capsule pressure gauge, model 63x.50 per ATEX 9
40433177.01 11/2020 EN/DE
5. Chemical reactions
The operator must ensure that chemical reactions between wetted
parts, process medium and environment are excluded.The materi-
als used can be found in the instrument marking. See chapter 2.5
„Labelling / Safety marks“.
On the wetted parts of the instrument, small residual amounts
of the adjustment medium (e.g. compressed air, water, oil)
can adhere from production.With increased requirements for
technical cleanliness, suitability for the application must be
checked by the operator before commissioning.
Liquid media with the property of changing the volume during
solidication can damage the measuring system (e.g. water if
it falls below the freezing point).
2.3 Personnelqualication
WARNING!
Riskofinjuryshouldqualicationbeinsucient!
Improper handling can result in considerable injury and
damage to property.
▶The activities described in these operating instructions
may only be carried out by skilled personnel who have the
qualications described below.
2. Safety

EN
Operating instructions capsule pressure gauge, model 63x.50 per ATEX
10
40433177.01 11/2020 EN/DE
Skilled personnel
Skilled personnel are understood to be personnel who, based on their
technical training, knowledge of measurement and control technology
and on their experience and knowledge of country-specic regulations,
current standards and directives, are capable of carrying out the work
described and independently recognising potential hazards.
2.4 Safety instructions for hazardous locations
WARNING!
Non-observance of these instructions and their contents may
result in the loss of explosion protection.
WARNING!
It is imperative that the application conditions and safety
requirements of the operating instructions are followed.
▶Pressure gauges must be grounded via the process
connection.
For use in ambient temperatures below the freezing point of
water, lled instruments are recommended. The case lling
prevents the formation of and freezing of condensation in the
case.
Permissible ambient temperature
Model 632.50: -40 ... +60 °C (unlled)
Model 633.50: -20 ... +60 °C (glycerine lling)
-40 ... +60 °C (silicone oil lling)
Attention! With gaseous media, the temperature may increase as a
result of compression warming. In these cases it may be necessary to
throttle the rate of change of pressure or reduce the permissible medium
temperature.
2. Safety
This manual suits for next models
1
Table of contents
Languages:
Other WIKA Measuring Instrument manuals

WIKA
WIKA 230.15 User manual

WIKA
WIKA DMSU21SA User manual

WIKA
WIKA GDM-100 User manual

WIKA
WIKA FLC-2300 User manual

WIKA
WIKA Delta-comb User manual

WIKA
WIKA 3 Series User manual

WIKA
WIKA DG-10-E User manual

WIKA
WIKA GDM-RC-100-T User manual

WIKA
WIKA TR12 Series Operating instructions

WIKA
WIKA 700.01.080 User manual