T6555 Online Residual Chlorine Meter
Membrane
Operating Manual


Perface
Features
Specifications
Installations
Instrument connection
Keypad descriptions
Display description
Menu structure
Calibration
Curve query
1
2
3
4
4
5
6
7
8
10
12
13
20
21
21
22
23
Contents
wall-mounted
MODBUS RTU General Information
pH buffer and temperature values
Daily Maintance
Package set
Warranty
Notes

1
Preface
Thank you for your support. Please read this manual carefully before use.The correct use
will maximize the performance and advantages of the product, and bring you a good
experience.
When receiving the instrument, please open the package carefully, check whether the
instrument and accessories are damaged by transportation and whether the accessories
are complete. If any abnormalities are found, please contact our after-sales service
department or regional customer service center, and keep the package for return
processing.
This instrument is an analytical measurement and control instrument with highly
precision.Only skilled,trained or authorized person should carry out installation, setup and
operation of the instrument.Ensure that the power cable is physically separated from the
power supply when connection or repair.Once the safety problem occurs, make sure that
the power to the instrument is off and disconnected.
For example, it may insecurity when the following situations occur:
1) Apparent damage to the analyzer
2) The analyzer does not work properly or provides specified measurements.
3)The analyzer has been stored for a long time in an environment where the temperature
exceeds 70 °C.
●Color LCD display
●Intelligent menu operation
●Multiple automatic calibration function
●Manual and automatic temperature compensation
●Three relay control switches
●High & low alarm and hysteresis control
●4-20mA & RS485,Multiple output modes
●Multi parameter display simultaneously shows –Free Chlorine, Temp,
current,etc.
●Password protection function to prevent misoperation by non-staff.

●Color LCD display
●Intelligent menu operation
●Multiple automatic calibration function
●Manual and automatic temperature compensation
●Three r elay control switches
●High & low alarm and hysteresis control
●4-20mA & RS485,Multiple output modes
●Multi parameter display simultaneously shows –Free Chlorine, Temp,
current,etc.
●Password protection function to prevent misoperation by non-staff.
The analyzer must be installed by professionals in accordance with relevant local
specifications, and instructions are included in the operation manual.
Comply with the technical specifications and input requirements of the analyzer.
Features
Online Free chlorine meter is a microprocessor-based water quality online monitoring
control instrument.It is widely used in drinking water treatment plants, drinking water
distribution networks, swimming pools, water treatment projects, sewage treatment, water
disinfection and other industrial processes.It continuous monitoring and control Free
chlorine and temperature value in aqueous solution.
2

3
Technical Specification
0.005~20.00mg/L ; 0.005~20.00ppm
Membrane
0.001mg/L ; 0.001ppm
±1%F.S
-2։16.00pH
pH
0.001pH
±0.01pH
-10։150.0˫(Based on sensor)
0.1˫
±0.3˫
2 groups: 4։20mA
RS485 Modbus RTU
Data record &Curve display
3 groups:5A 250VAC,5A 30VDC
85~265VAC,9~36VDC,power consumption≤3W
No strong magnetic field interference around except the geomagnetic field.
-10։60˫
≤90%
IP65
1.5kg
235×185×120mm
Wall mounted
Measurement range
Measurement unit
Resolution
Basic error
Measurement range
Measurement unit
Resolution
Basic error
Temperature
Temperature Resolution
Temperature Basic error
Current output
Signal output
Other functions
Three relay control contacts
Optional power supply
Working conditions
Working temperature
Relative humidity
Waterproof rating
Weight
Dimensions
Installation methods
1

Installations
Installation size
Installation completion drawingFix the fixed piece of the instrument
185 mm
35 mm
235 mm
190 mm
4
120 mm
Instrument shape, wall - mounted fixed piece size
Hole:8*12.5mm

5
Instrument connection
The connection between the instrument and the sensor: the power supply, output signal, relay alarm contact and the
connection between the sensor and the instrument are all inside the instrument, and the wiring is as shown above. The
length of the cable lead fixed by the electrode is usually 5-10 Meters, insert the corresponding label or color wire on the
sensor into the corresponding terminal inside the instrument and tighten it.
RLY1 RLY2 RLY3
G,L,N
V+,V-,RX,TX
V+,V-,A,B
V+,V-,A,B
A1,B1
+,-
3 groups of relay
G-groundwire, L-live, N-neutral
RS232 communication output
Digital input channel 1
Digital input channel 2
RS485 communication output
DC power supply
EC1,EC2,EC3,EC4
T+,T-
INPUT
REF
S+,S-
CE,RE,WE
I1,G,I2
I3,G,I4
Electrical conductivity/resistivity wiring
Temperature connection
pH/ORP/lon measure
pH/ORP/lon reference
Membrane dissolved oxygen, FCL
Constant voltage residual chlorine/
chlorine dioxide/ozone
Output current
Input current
Tips: If the user dials the wire by himself, lengthens the cable or shortens
the cable,
Take care to remove the black conductive skin outside the transparent
wire.
To prevent short circuit from the reference end of the electrode, please
use a multimeter to check
Use the pH dedicated cable connection.
RLY1 RLY2 RLY3
G N L
AC220V
EC1 EC2 EC3 EC4 T+ T- INPUT REF S+ S- WE RE CE I1 G I2 I3 G I4
COND
V+ V- RX TX
RS232
V+ V- A B
SENSOR1
V+ V- A B A1 B1 + -
SENSOR2 485
DC24
pH
DO/FCL
FCL/DOZ
mA1 mA2 mA3 mA4
TEMP

Keypad descriptions
Keypad operation tips:
Short Press: Short Press means to release the key immediately after pressing. ((Default to
short presses if not indicated below)
Long Press: Long Press is to press the button for 3 seconds and then release it.
6
Enter menu setting mode
Return measurements
Two modes switching
Enter standard solution calibration mode
Calibrated: check the calibration status
Re-calibration: press “ENT” again
Return to
previous menu
Confirmation
options
Menu setting mode:
press this key to loop
down the menu options
Menu setting mode:
press this key to
rotate menu options
Press the key in
measurement mode to
display the trend chart
In measurement mode,
press this button to set
parameters for browsing

7
25.0 °C7.01 pH
8.025 mg/L
2019-01-09 12:53:17
Free Chlorine
Relay 1
[blue is off
and red is
on
Relay 2
[blue is off
and red is on
Cleaning
Temp. + Unit
Automatic
temp. comp.
Current 1
Current 2
Switch display
Setting modeMeasurement Mode
Trend Chart DisplayCalibration Mode
20
16
12
8
4
050 100 150 200 250 300 350 400
2019-01-09 12:53:17
2019-01-09 12:53:17
14.001 25.0 °C
Free Chlorine
2019-01-09 12:53:17
Free Chlorine
Configure
Calibration
Set points
Output
History
System
Relay 3
[blue is off
and red is
on
2019-01-09 12:53:17
Free Chlorine
Voltage: -5 mV
Calibrating
Point 1
0.01 mg/L
25.0 °C7.01 pH
8.025 mg/L
2019-01-09 12:53:17
Free Chlorine
Sensor Communication
Warning
pH Compensation
+Unit
Display descriptions
All pipe connections and electrical connections should be checked before use. After the power is
switched on, the meter will display as follows.
Main measurment+ Unit
Instrument type
Year-Month-Date Time
Table of contents
Other Twinno Measuring Instrument manuals

Twinno
Twinno T4050 User manual

Twinno
Twinno T6570 User manual

Twinno
Twinno T6055 User manual

Twinno
Twinno RM785 User manual

Twinno
Twinno T4058 User manual

Twinno
Twinno T4030 User manual

Twinno
Twinno T4055 User manual

Twinno
Twinno T4046 User manual

Twinno
Twinno T6000 User manual

Twinno
Twinno T6075 User manual