Twinno T4030 User manual

T4030 Online Conductivity Meter
Conductivity/Resistivity/TDS/Salinity
Operating Manual
Shanghai Chunye Instrument Technology Co.,Ltd
上海淳业仪表科技有限公司


Table of Contents
Preface
Features
Technical Specifications
Installation
Installation size
Embedded installation
Wall mounted installation
Electrical connection
Keypad descriptions
Display descriptions
Menu
Calibration
MODBUS RTU
Daily Maintenance
Package Set
FAQ
Warranty
Notes
1
1
2
3
3
4
4
5
6
7
8
10
12
19
20
21
22
23

Features
Industrial online conductivity meter is a microprocessor-based water quality online
monitoring control instrument.The meter is equipped with conductivity electrodes with
different constants This instrument is widely used in power plants, petrochemical industry,
mining industry, paper industry,semiconductor industry , medicine, food and beverage
and other industries.It is suitable for softening water, raw water, steam condensate water,
seawater distillation and deionized water, etc. It can continuously monitor and control
the conductivity, resistivity, TDS, salinity and temperature of aqueous solutions.
Ӫ
Ӫ
Ӫ
Ӫ
Ӫ
Ӫ
Ӫ
Preface
Thank you for your support. Please read this manual carefully before use.The correct use
will maximize the performance and advantages of the product, and bring you a good
experience.
When receiving the instrument, please open the package carefully, check whether the
instrument and accessories are damaged by transportation and whether the accessories
are complete. If any abnormalities are found, please contact our after-sales service
department or regional customer service center, and keep the package for return
processing.
This instrument is an analytical measurement and control instrument with highly
precision.Only skilled,trained or authorized person should carry out installation, setup and
operation of the instrument.Ensure that the power cable is physically separated from the
power supply when connection or repair.Once the safety problem occurs, make sure that
the power to the instrument is off and disconnected.
For example, it may insecurity when the following situations occur:
1) Apparent damage to the analyzer
2) The analyzer does not work properly or provides specified measurements.
3)The analyzer has been stored for a long time in an environment where the temperature
exceeds 70 °C.
The analyzer must be installed by professionals in accordance with relevant local
specifications, and instructions are included in the operation manual.
Comply with the technical specifications and input requirements of the analyzer.
1
Color LCD display
Intelligent menu operation
Multiple automatic calibration function
Manual and automatic temperature compensation
Two relay control switches
High limit,low limit and hysteresis control
4-20mA & RS485,Multiple output modes
Password protection function to prevent misoperation by non-staffӪ
Ӫ

Technical Specifications
Conductivity
Resolution
Basic error
Resistivity
Resolution
TDS
Resolution
Salinity
Resolution
Stability
Temperature
Temperature resolution
Temperature accuracy
Temperature compensation
Temperature compensation
Electrode residual signal
Response time
Stability
Current output
Communication output
Relay control set-points
Optional power supply
Working conditions
Working temperature
Relative humidity
Waterproof rating
Weight
Dimensions
Installation opening size
Installation methods
0~500mS/cm
0.01μS/cm;0.01mS/cm
±0.5%F.S
0~18.25MΩ/cm
0.01KΩ/cm;0.01MΩ/cm
0~9999mg/L;0~250g/L
0.01mg/L;0.01g/L
0~700ppt
0.01ppt
pH:≤0.02pH/24h;ORP: ≤2mV/24h
0~50.0°C
0.1°C
±0.3°C
0~60.0°C
Manual or automatic
<1‰
25°C<60S; 35°C<30S (To attain 90%)
At constant pressure and temperature, the weekly drift<2%F•S;
Two:4~20mA,20~4mA,0~20mA(load resistance<750Ω)
RS485 MODBUS RTU
Two:3A 250VAC,3A 30VDC
85~265VAC,9~36VDC,power consumption≤3W
No strong magnetic field interference except the geomagnetic field.
-10~60°C
≤90%
IP65
0.6kg
98×98×130mm
Panel & wall mounting
2
92.5×92.5mm

Instrument installation
Instrument dimensions Instrument dimensions(Side view)
Insert the instrument into the square hole
and fix it with the collocated clip.
Insert mounting hole size
Embedded installation
93 mm
93 mm
(± 0.5 mm)
(± 0.5 mm)
3
92.5 mm
10 mm
120 mm
Installation of back
cover with
waterproofing seal
98 mm
98 mm
MENU
ESC

4
53 mm
18.4 mm
ħ5mm
a. Fixed bracket (M4 screw) is
installed on the wall first.
b. Insert the instrument into the
fixture along the chute
Fixture Dimension
Wall mounting
Fixed bracket (M4 screw) is
installed on the wall first.
Insert the instrument into the
fixture along the chute

Instrument connection
5
Electrical connection
Connection between instrument and sensor: Connection of power supply, output signal, relay
contacts and instrument baseplate. The cable length of sensors is usually 5-10 meters. There are
labeled inserts at the end of the cable, which can be inserted into the terminal with the same
digital symbols on the instrument roof and tightened.
Terminal
CELL1
CELL2
CELL3
CELL4
Temp
Temp
Cure1+
Com-
Cure2+
RS485 B
RS485 A
Function
Conductivity Sensor Input 1
Conductivity Sensor Input 2
Conductivity Sensor Input 3
Conductivity Sensor Input 4
Temperature sensor Input
Temperature sensor Input
Current output 1+
Current output common
Current output 2+
Communication Terminal
Communication Terminal
Function
High set relay working position
Alarm relay common
High set relay resting position
Low set relay working position
Alarm relay common
Low set relay resting position
DC power +(9-36V)
DC power -
Earth wire
Live wire
Neutral wire
Earth wire
Terminal
NO 1
COM 1
NC 1
NO 2
COM 2
NC 2
Pow DC +(9-36V)
Pow DC-
Pow PE
Pow ACL (85-265V)
Pow AC N
Pow PE
CELL1
CELL2
CELL3
CELL4
Temp
Temp
N/C
N/C
N/C
Cur 1+
Com -
Cur 2+
RS485 B
RS485 A
COND
NO 1
COM 1
NC 1
NO 2
COM 2
NC 2
Pow DC+
(9~36V)
Pow DC-
Pow PE
Pow ACL
(85~265V)
Pow AC N
Pow PE

Keypad descriptions
Keypad operation tips:
Short Press: Short Press means to release the key immediately after pressing. ((Default to
short presses if not indicated below)
Long Press: Long Press is to press the button for 3 seconds and then release it.
Press and hold: Press and hold means to press the button, and accelerate after a certain
time until the data is adjusted to the user's required value before releasing the button.
Confirmation options
Enter menu setting mode
Return to previous menu
Long press to return the measurements screen
Menu setting mode: press this key to loop down the menu options
Value input mode: current bit value change (loop)
In measurement mode, long press this button to set parameters for browsing
Press this key to move menu options left (circle)
6
MENU
ESC

Relay 1
[blue is off
and red is on]
Instrument type
Display descriptions
All pipe connections and electrical connections should be checked before use. After the power is
switched on, the meter will display as follows.
Setting mode
Measurement mode Calibration mode
Main measurment+ Unit
Relay 2
[blue is off
and red is on]
Current 1
Current 2
Switch display
Automatic
temp. comp.
Temp. + Unit
The percentage
corresponding
to the main
measurement
7
Configure
Calibration
Set Points
Output
System
Conductivity / TDS / Salinity / Resistivity
Conductivity / TDS / Salinity / Resistivity
83.01μS/cm
25.0 °C
14.51 mA
C1
SP1 SP2
Voltage:
Point 1
Conductivity / TDS / Salinity / Resistivity
0Done
μS/cm
-5 mV
Conductivity / TDS / Salinity / Resistivity
83.01μS/cm
25.0 °C
14.51 mA
C1
SP1 SP2
Table of contents
Other Twinno Measuring Instrument manuals

Twinno
Twinno T4050 User manual

Twinno
Twinno T6000 User manual

Twinno
Twinno T4055 User manual

Twinno
Twinno RM785 User manual

Twinno
Twinno T4046 User manual

Twinno
Twinno T4058 User manual

Twinno
Twinno 16540 User manual

Twinno
Twinno T6555 User manual

Twinno
Twinno T6570 User manual

Twinno
Twinno T6500 User manual