8 TX6353-UM-EN-01 Draft r
Operating temperature
range
-10 to +40°C
Storage temperature -20 to +60°C
Humidity 90% non-condensing
Housing materials Reinforced polymer. EMC protected and proof
against surface electrostatic charge
Protection classification Housing dust and waterproof to IP65
Gas port protected to IP54
Weight 550 g
Mounting Two 6.5 mm diameter fixing holes with Unistrut
6946 indents
Cable entries 2 x M20
Electrical connections 4 mm barrier/clamp terminals with circuit
segregation barriers
Conductors A maximum of two per terminal
Conductor size A maximum of 2 x 2.5 mm²
Terminal torque 2.4 Nm maximum
Information display 128 x 65 dot graphic LCD screen (Backlit)
Vibration limits (EN 61779)
10 to 200 Hz, 7g peak
Impact limits 20 joules (housing)
Calibration and setup
Digitally controlled Zero and Span pushbutton setting
Signal fix The transmitted output signal of the sensor is Fixed
during calibration to prevent false alarms from being
initiated
Fault indication Under-range signal transmitted and Fault display for:
• Loss of communication between sensing
eModule and transmitter
• Critical reduction in sensor sensitivity
• Sensor over-range protection
Alarms Inbuilt adjustable visual alarm LEDs for General and
High alarms. Programmable setpoints and options
selectable from the setup menus
3
www.trolex.com