Thermo orion 135A User manual

Instruction Manual
Models 135A and 136S
Waterproof and Intrinsically Safe
Portable Conductivity Meters

Software Version: 1.x
TA–193.204–ORE02 260201
AQUAfast, Cahn, ionplus, KNIpHE, ORION, PerpHecT, perpHecTion, Sens-
ing the future, SensorLink, Sure-Flow, Titrator PLUS and TURBO2 are reg-
istered trademarks of Thermo Orion.
1-888-pHAX-ION, All in One, AssuredAccuracy, AUTO-BAR, AUTO-CAL,
AUTO DISPENSER, AUTO-LOG, AUTO-STIR, AUTO-READ, Cable-Free,
CERTI-CAL, CISA, DataCOLLECT, digital LogR, DirectCal, DuraProbe,
Extra Easy/Extra Value, FAST QC, GLPcal, GLPcheck, GLPdoc, Ionalyzer,
KAP, KNIpHE, LogR, Minimum Stir Requirement, MSR, NISS, One-Touch,
One-Touch Calibration, One-Touch Measurement, Optimum Results, PEN-
Pal, pHISA, pHix, Phuture, Pure Water, QuicKcheK, rf link, ROSS, ROSS
Resolution, Sage, SAOB, Stat Face, The Enhanced Lab,ThermaSense,
Triode, TRIUMpH, Unbreakable pH, Universal Access and Wine Master are
trademarks of Thermo Orion.
Guaranteed Success and The Technical Edge are service marks of
Thermo Orion.
PerpHecT meters are protected by U.S. patent 4,321,544. Other patents
pending.
ROSS and PerpHecT ROSS are protected by U.S. patent 4,495,050. Other
patents pending.
ORION Series A meters and 900A printer are protected by U.S. patents
5,108,578, 5,198,093, D334,208, D346,753.
ORION 81, 82, 91, and 92 series glass electrodes are protected by U.S.
patents 4,661,236 and 4,687,500.
Sure-Flow electrodes are protected by European patent 278,979 and Cana-
dian patent 1,286,720. Other patents pending.
ionplus electrodes and Optimum Results solutions have patents pending.
ECopyright 1999, Thermo Orion. All rights reserved.
The specifications, descriptions, drawings, ordering information and part
numbers within this document are subject to change without notice.
This publication supersedes all previous publications on this subject.

Models 135A and 136S Portable Conductivity Meters
Information III
Safety Precautions
Be sure to read and follow these instructions !
The Model 136S Conductivity Meters may only be opened
to change the batteries outside hazardous areas. If repairs
are necessary, the meter must be sent in to the factory.
Never operate the remote interface or printer within haz-
ardous areas.
When using the meter in hazardous areas, watch for elec-
trostatic charges! For example, never wipe off the meter
with a dry cloth. Observe the relevant regulations concern-
ing ESD.
Whenever it is likely that the protection has been impaired, the meter shall
be made inoperative and secured against unintended operation.
The protection is likely to be impaired if, for example:
❏the meter shows visible damage
❏the meter fails to perform the intended measurements
❏after prolonged storage at temperatures above 70 °C
❏after severe transport stresses
Before recommissioning the meter, a professional routine test according to
EN 61 010-1 shall be performed. This test should be carried out at our fac-
tory.

cal
Note
Warning
Information IV
Information on this Instruction Manual
ITALICS are used for texts which appear in the display of
Thermo Orion Model 135A or 136S.
Bold print is used to represent keys, e.g. cal.
Display examples
or
keys whose functions are explained are frequently shown
in the left-hand column.
Notes provide important information which should always
be observed when using the meter.
Warning means that the instructions given must always be
followed for your own safety. Failure to follow these
instructions may result in injuries.

Models 135A and 136S Portable Conductivity Meters
Contents V
Contents
Safety Precautions III. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Information on this Instruction Manual IV. . . . . . . . . . . . . . . . . . . . . . . . . .
1 The Models 135A and 136S Conductivity Meters 1. . . . . . . . . . . . . . .
Package Contents 1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Short Description of Meters 1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2 Operation 3. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Meter Design 3. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Display 4. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Keypad 4. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Connection and Start-up 6. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Configuration 8. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Calibration 11. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Measurement 15. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Data Log Memory 17. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Data Log 18. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Clock Mode 20. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Serial Interface 21. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Standard Setting for PRT300 Printer 22. . . . . . . . . . . . . . . . . . . . . . .
Configuring the PRT300 Printer 22. . . . . . . . . . . . . . . . . . . . . . . . . . . .
Printing Measured Values and GLPdoctReport 23. . . . . . . . . . . . .
3 Troubleshooting and Maintenance 24. . . . . . . . . . . . . . . . . . . . . . . . . . .
Error Messages 24. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Maintenance 27. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Changing batteries 27. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Cleaning the meter 28. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Contents VI
Appendix 29. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Declarations of Conformity 29. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Certificate of Conformity 31. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Control Drawing 33. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Ordering Information 35. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Specifications 39. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Specifications for PRT300 Printer 40. . . . . . . . . . . . . . . . . . . . . . . . . .
Warranty 41. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Technical Terms 44. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Index 46. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Models 135A and 136S Portable Conductivity Meters
Models 135A and 136S 1
1 The Models 135A and 136S
Conductivity Meters
Package Contents
Please check the completeness of the shipment after un-
packing.
The package should include:
❏Thermo Orion Model 135A or 136S Conductivity Meter
(ready for operation)
❏Neck strap
❏This instruction manual
❏Quickstart instructions
❏Interface cable with adapter for printer and PC
❏PC software
Short Description of Meters
❏The Models 135A and 136S measure conductivity, sali-
nity, TDS and temperature in industry, the environment,
food processing and waste-water treatment.
❏Operation of the Model 136S is also permitted in Zone 1
hazardous areas.
❏The meters meet the European EMC regulations
(89-336-EEC) and the recommendations of NAMUR
NE 21.
❏The meters are IP 66 protected to EN 60 529.

Warning
Models 135A and 136S 2
❏Temperature compensation is automatic with a conducti-
vity probe with integral temperature sensor or a sepa-
rate ATC probe, or the temperature may be manually
entered.
❏Calibration can be carried out by directly entering the
cell constants, by calibrating with Thermo Orion Con-
ductivity Standards, or with any other buffer solutions.
❏The data log records up to 100 measured values with
the temperature, date and time. Recording takes place
either manually, interval or event-controlled.
❏To minimize battery consumption, the meter switches off
after either one or twelve hours when it is not operated.
❏Only three alkaline AA batteries are required for uninter-
rupted operation for approx. 1,000 hours.
❏PC software allows complete remote control of the me-
ter via PC. All measured values and parameters can be
read out and easily processed further (e.g. using Micro-
soft Excel).
❏Measured values and GLPdoctcan also be sent di-
rectly to a printer via the serial interface.
Never use the remote interface to PC or printer in hazard-
ous areas!

Models 135A and 136S Portable Conductivity Meters
Operation 3
2 Operation
Meter Design
1
2
3
4
5
1 Sensor connection
2, 3 Separate temperature probe connection
4 PC/printer interface connection
5 Probe holder, removable
meas cal
STO
RCL
Y
B
print
clock
power

power
meas
Note
cal
Operation 4
Display
Battery
empty Press
cal key Wait
Manual
temperature
detection
Temperature
detection
using NTC
Write
memory
Read
memory
Slope
(cell
constant)
Clock
Main
display
Secondary
display
Measure-
ment units
Error
message
Keypad
Pressing the power key switches the meter on or off.
After switching on, the meter automatically performs a
self-test and checks for the presence of a temperature
probe. Then it goes into the measuring mode.
Pressing the meas key returns the meter to the measur-
ing mode from any function. Pressing the meas key in the
measuring mode displays the following parameters:
Cond measuring mode: temperature compensation
tdS measuring mode: TDS factor
You can also power the meter up using the meas key.
However, only an abbreviated self-test is performed.
Pressing the cal key starts calibration.
This manual suits for next models
1