MRU-120HD
ERP-1
Saving a result to the memory
After the measurement press ENTER.
Select memory cell using buttons ▲▼.
Select memory bank using buttons ◄►.
Connect the meter to the leg of the measured pole.
Select method 3P . Connect the exible clamp to
ERP-1 adapter. Wrap the clamp around the object's
leg and it's earthing tape. Select measuring voltage.
Press START.
Earth resistivity measurement
Connect the meter to the measured earth.
Enter settings
▪F1 measuring voltage
Using button ENTER run the measurement.
Press START.
Using buttons ▲▼enter the distance between electrodes.
Main result
Additional
results
Earth resistance measurement RE
3P
3P
3P
4P
4P
Measurement of RE of poles using ERP-1 adapter
Target cell empty Target cell occupied
Press ENTER to save the result.
Enter settings
▪ F1 voltage/pulse shape
Using button START run the measurement.
Conguration and RE measurement
UN........voltage on the measurement points
fN..........interference frequency
IL..........interfering current
RH........resistance of current electrode
RS.........resistance of voltage electrode
δ..........additional uncertainty caused by the resistance
of the electrodes
Main result
Additional
results
Similarly, connect the next legs of the object. Having
the results of each leg measurement, you can calculate
the resultant resistance of the entire object from the
formula:
RE= 1
1+ ... + 1
R1Rn
Using FLEX button select the type of exible clamps
connected to the device.
Using TURNS button select the number of exible
clamp wraps around the pole leg.