
MA-5510_EN.doc Page 6 on 14 Rev: 27-03-19
3.4. Calibration
It must be done after the sensor has been turned on for
a while (10-15 minutes) to obtain a uniform
temperature of the installation. The cells do not usually
need to be adjusted with each other. However, when
greater precision is needed, it is sometimes necessary
to stabilise the cells individually with the resistances in
the junction box. Those resistances are of several
ohms (±10) and are installed in the supply circuit.
A parallel adjustable resistance is mounted with a fixed
resistance. The most sensitive cell will have its input
resistance increased and the least sensitive will have
its lowest input resistance. You will see that it is
preferable to work on both supply cables: schematic
mounting is given for your information and allows a
variation of 0 to 20 ohms in series on the input
impedance (2x10 Ω).
Note: A well known weight of more than 20% of the
nominal load of the system can be expected. The
calibration error is always much higher than the error
made on the evaluation of the load.
3.5. Measurement errors
When the calibration is difficult and measurement errors are observed, it is necessary
to check the installation. Mechanically, the cells must be free in the direction of the
load and well positioned. Electrically, the connections must be securing, the junction
boxes exempt from humidity and the cables intact.
If there is no fault to be seen, it is necessary to verify the internal circuit. SENSY can help to diagnose on the
basis of the associated diagnosis sheet provided in the appendix and filled in beforehand.
3.6. Insulation test
The measuring of the insulating resistance is done with
a multimeter. The standardized testing voltage is 10 V.
It is applied to a conductor. It can be determined by
disconnecting the measuring instrument and applying
voltage between one of the conductors and the metallic
mounting structure, or individually, cell by cell to situate
the leakage with precision. The insulation must not, in
any case, be lower than 2 GΩfor a 10 V voltage. This
insulation default will generate measurement errors if
the insulation resistance is lower than several hundred
MΩ. Insulation default can also be generated by
environmental conditions (temperature, humidity).