Metrix MX2033 User manual

Digital Proximity System
Operaon and Maintenance Manual
100576 • REV C
The DPS (Digital Proximity System) devices
consists of two modules: the MX2033 and
MX2034. The MX2033 is a convenonal
3-wire driver while the MX2034 is a two-
wire current loop transmier. Users can con-
gure either unit to measure peak-to-peak
vibraon, gap distance, or rotaonal speed.
These products are used in conjuncon with
a proximity probe and extension cable to
measure the vibraon levels, gap distances,
or rotaonal speed of machinery shas. Us-
ers can congure the units via a USB cable
and conguraon soware.

Doc# 100576 • REV C (Sept 2017) Page 2 of 42
TABLE OF CONTENTS
........................................................3
1.1 Safety Terms and Symbols ..................................................3
1.2 General Safety Summary. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
2. Overview .........................................................................4
2.1 MX2033 Three-Wire Proximity Driver..............................................5
2.2 MX2034 Two-Wire Transmier....................................................5
2.3 For more informaon. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
.................................................................6
3.1 Equipment Required .......................................................6
3.2 Soware Installaon .......................................................7
3.3 Connecng the DPS to a Computer ..........................................11
3.3.1 Remove the Base .......................................................11
3.3.2 Apply Power ...........................................................12
3.4 Launching the DPS Conguraon Soware ....................................12
3.5 Main Screen .............................................................13
3.5.1 Command Line Opons ..................................................13
3.5.2 Device Informaon ......................................................13
3.5.3 Device Communicaon Status .............................................14
3.5.4 Congurable Device Sengs ..............................................14
3.5.5 Conrming the Conguraon .............................................16
3.5.6 Prinng a Label .........................................................16
3.5.7 Installing the Label ......................................................18
3.6 Using the MX2034 Buered Output .........................................18
.................................................19
4.1 Scale Factor Vericaon ...................................................20
4.2 Scale Factor Custom Calibraon .............................................24
4.2.1 Current Loop Vericaon (MX2034 only) ....................................26
..........................................................29
5.1 MX2033 and MX2034 .....................................................29
5.2 MX2034 Only ............................................................30
.......................................................31
6.1 Driver/Transmier Temperature Response ....................................31
6.1.1 MX2030/BN 3300 5 Meter System .........................................31
6.1.2 MX2030/BN 3300 9 Meter System .........................................34
6.1.3 Metrix/BN 7200 5 Meter System ..........................................36
6.1.4 Metrix/BN 7200 9 Meter System ..........................................39
6.2 Phase Delay .............................................................42
.................................................43

Doc# 100576 • REV C (Sept 2017) Page 3 of 42
HIGH VOLTAGE
PRESENT
PROTECTIVE
EARTH NOTE
DANGER or
CAUTION
FUNCTIONAL
GROUND
1. GENERAL INFORMATION
1.1 SAFETY TERMS AND SYMBOLS
Terms that appear in this manual requiring special aenon include:
• WARNING: Warning statements idenfy condions or pracces that could result in injury
or loss of life.
• CAUTION: Cauon statements idenfy condions or pracces that could result in dam-
age to the product, loss or corrupon of data, or damage to the environment or other
property.
• NOTE: Notes idenfy material of special interest or importance to the user, not including
cauons or warnings.
Symbols that may appear on the product and/or in this manual include:
1.2 GENERAL SAFETY SUMMARY
Review the following safety precauons to avoid injury and prevent damage to this product
or any products connected to it.
• USE ONLY AS SPECIFIED
To avoid potenal hazards, use this product only as specied. Only qualied personnel
should perform installaon and uninstallaon procedures.
• OBSERVE ALL TERMINAL RATINGS
To avoid re or shock hazard, observe all rangs and markings on the product. Consult
the individual secons of this manual for further rangs informaon before making con-
necons to the product.
• AVOID EXPOSURE TO CIRCUITRY
Do not touch exposed electrical connecons and components when power is present.
• DO NOT OPERATE WITH SUSPECT FAILURES
If you suspect there is damage to this product, have it inspected by qualied personnel.

Doc# 100576 • REV C (Sept 2017) Page 4 of 42
1.3 RECEIVING, INSPECTING AND HANDLING THE SYSTEM
Metrix ships the probe, extension cable, and driver as separate units that the user intercon-
nects at the installaon site. Carefully remove all equipment from the shipping containers
and inspect it for shipping damage. If you see shipping damage, le a claim with the carrier
and submit a copy to Metrix Instrument Co. Include part numbers and serial numbers on all
correspondence. If no damage is apparent and the equipment is not going to be used imme-
diately, return the equipment to the shipping containers and reseal unl ready for use. Store
the equipment in an environment that is free from potenally damaging condions such as
extreme temperature, excessive humidity, or a corrosive atmosphere.
2. OVERVIEW
The Metrix Digital Proximity System module comes in two versions:
• MX2033 – 3-wire proximity driver
• MX2034 – 2-wire, loop powered proximity transmier
A complete DPS system requires a DPS module, a proximity probe, and an extension cable.
You can congure the MX2033 and MX2034 modules to operate with the probe and cable
combinaons shown in Table 1.
Metrix MX2030 5 mm or 8mm p 5m
Metrix MX2030 5 mm or 8 mm p 9 m
Metrix 7200 5 mm or 8 mm p 5 m
Metrix 7200 5 mm or 8 mm p 9 m
Bently Nevada 3300 or 3300 XL 5 mm or 8
mm p
5 m
Bently Nevada 3300 or 3300 XL 5 mm or 8
mm p
9 m
Bently Nevada 7200 5 mm or 8 mm p 5 m
Bently Nevada 7200 5 mm or 8 mm p 9 m
Standard target material is AISI 4140. Other materials available upon request.

Doc# 100576 • REV C (Sept 2017) Page 5 of 42
The secons below provide more module informaon.
2.1
The MX2033 signal output is proporonal to the distance between the probe p and the
target material. The signal output follows API Standard 670 and is compable with most
connuous vibraon monitoring systems. The MX2033 uses -24Vdc excitaon and provides
an output scale factor of 7.87 mV/mm (200mV/mil) for 5 mm and 8mm probes.
2.2
The MX2034 provides an ISA standard 4-20 mA signal proporonal to vibraon, axial posi-
on sha speed (RPM) for direct connecon to a PLC, DCS, SCADA system, or other instru-
mentaon without requiring a separate monitor system. The MX2034 is powered by +24
Vdc, supplied within the current loop. The MX2034 is user congurable to funcon as either
a radial vibraon transmier (where the 4-20 mA signal is proporonal to peak-peak vibra-
on amplitude), as an axial posion transmier (where the 4-20 mA signal is proporonal to
average probe gap) or as a speed (RPM) transmier (where the 4-20 mA signal is propor-
onal to the speed of a machine). For convenience when connecng to signal analyzers,
portable data collectors, and test instrumentaon, the raw vibraon signal is available at
a short-circuit protected BNC connector. Refer to secon 3.6 of this manual for important
informaon and cauons regarding proper use of this connector.
2.3 FOR MORE INFORMATION
Refer to these documents for more informaon:
1087015 MX2030 Digital Proximity Transducer System Datasheet
1232961 DPS Hazardous Area Installaon Manual
100528 DPS Label Kit Manual

Doc# 100576 • REV C (Sept 2017) Page 6 of 42
3. CONFIGURING A DPS
• This secon lists the procedure for conguring a DPS:
• Installing the DPS Conguraon Soware
• Connecng a DPS to your computer
• Retrieving the conguraons from the DPS
• Changing the DPS conguraon for probe/cable type
• Changing the DPS conguraon for measurement type (MX2034 only)
• Changing the DPS full scale range (MX2034 only)
• Prinng DPS labels
3.1 EQUIPMENT REQUIRED
You will need the equipment listed below to congure a DPS.
Item Notes
Computer With Windows XP, or Windows 7 or Windows 10, Operang
System
24V Power
Supply
Posive or Negave. Refer to datasheet 1087015 for power
supply voltage tolerance and DPS current requirements.
USB Cable USB 2.0 A to mini B
Labels Avery 6570
Label Overlay Included with the DPS

Doc# 100576 • REV C (Sept 2017) Page 7 of 42
3.2 SOFTWARE INSTALLATION
Metrix provides the DPS Conguraon and Ulity Soware on a USB memory sck that you
request with the DPS (Refer to the DPS datasheet for soware ordering informaon.)
Insert the USB sck into your computer and open explorer to view the les.
Double-click the version for your operaon system according to Table 2 and follow the
instrucons listed on the following pages.
Version
Windows XP DPSUser.exe (32 bit)
Windows 7 DPSUser_x64.exe (64 bit)
Click Next to connue

Doc# 100576 • REV C (Sept 2017) Page 8 of 42
Figure 2 shows the end user license agreement. Click ‘Accept” to connue
Click Install to begin the installaon

Doc# 100576 • REV C (Sept 2017) Page 9 of 42
If you are running Windows 7, Windows may present a User Account Control noce as
shown in Figure 4. Click “Yes” to connue.
You may addionally see a Windows Security noce as shown in Figure 5. Click ”Install this
driver soware anyway” to connue.

Doc# 100576 • REV C (Sept 2017) Page 10 of 42
Click the Finish buon as shown in Figure 6 to close the installaon process.
Other manuals for MX2033
2
This manual suits for next models
1
Table of contents
Other Metrix Measuring Instrument manuals

Metrix
Metrix MX 409 User manual

Metrix
Metrix MTX 1032-B User manual

Metrix
Metrix VM2800 User manual

Metrix
Metrix AG 900 User manual

Metrix
Metrix GX-1030 User manual

Metrix
Metrix MX 406B User manual

Metrix
Metrix 6850-001 User manual

Metrix
Metrix IX 3131B User manual

Metrix
Metrix MX 650 User manual

Metrix
Metrix SA6200A User manual