Metreg MQM User manual

livinggas.
Metreg Technologies GmbH
MQM
Quantometers
Manual
Quantometers

www.metreg-technologies.de
MQM_Manual_EN_16.01.2015
©2014MetregTechnologiesGmbH•Design:www.reidelsoltaugrakdesign.de
All rights reserved. Subject to technical changes.

5
MQM
Quantometers
Manual
Quantometers

www.metreg-technologies.de
Table of Content
Manual MQM Quantometers
1. Purpose of this Manual ................................................................................................ 5
2. Basic safety instructions ............................................................................................... 5
3. Declaration................................................................................................................. 6
4. Use and Scope of Quantometers ................................................................................... 6
5. DesignoftheMQMQuantometer................................................................................. 6
6. TechnicalData......................................................................................................... 7-8
7. Installation .............................................................................................................. 8-9
8. MountingPosition/DirectionofFlow........................................................................... 9
9. Preliminarywork.................................................................................................. 10-11
10. Lubrication and Maintenance of the MQM Quantometer .............................................. 12
11. Maintenance of MQM Quantometers ......................................................................... 13
12. Commissioning of the MQM Quantometer .................................................................. 13
13. DecommissioningprocedureoftheMQMQuantometer.............................................. 13
14. Pulse Generators .................................................................................................. 14-15
15. Pressure Measuring Point ........................................................................................... 16
16. Temperature Measuring Point ................................................................................... 16
17. Cleaning ................................................................................................................... 17
18. Transport and Storage ............................................................................................... 17
19. Annex A ................................................................................................................... 18
20. Annex B .................................................................................................................... 19

5
1. Purpose of this Manual
The instruction manual is dedicated to technical sta with adequate training and technical
knowledgeinthesectorofgasdistributionindustryaccordingtotheapplicabletechnical
lawsandrequirements.ForexamplestawhoareeducatedaccordingDVGWcodesof
practice G 492 and / or G 495.
2. Basic safety instructions
Requirements for the personnel
Theoperationalstamustbequalied,instructedandauthorizedaccordingtheassigned
functions.Allserviceandmaintenanceworkshallbeperformedexclusivelybypersonswith
provenexpertknowledge,includingknowledgeofoccupationalsafetyandenvironmental
protection.
Protectiveactionshavetobedenedaccordingthelegalregulations.
Designated use
The MQM Quantometer is designed for in-plant volume measurement of combustible gases
ofthe1.,2.and3.gasfamily(naturalgas,towngas,propane,butaneandothers),puriedand
driedbiogasaswellasotherinert,non-aggressiveandnon-explosivegasesandgasmixtures.
The permissible operating pressure range reaches up to 20 bar dependent on the design of
themeter,thetemperaturerangerunsfrom-20°Cto+60°C.
The manufacturer is not liable for damage caused by improper or not-designated use.
Safety at work
Weartherequiredpersonalprotectiveequipmentaccordingtofederal/nationalregulations.
Operational safety
The device is permitted to operate in technical faultless and fail-safe condition only.
Responsible is the operator.
Product safety
Thismeasuringdeviceisdesigned,manufacturedandtestedaccordingthestate-of-art
demands. It meets all legal requirements and complies with the EC directives.
Themanufacturerconrmsthedevice-specicECDeclarationofConformityby
xingtheCEmarktothedevice.
The device has left the factory in perfect condition.
Warrantycouldbeextendedonlyincaseofproperinstallationandoperationaccording
to this manual.

3. Declaration
Declaration of Conformity (AnnexA)
4. Use and Scope of Quantometers
TheQuantometerMQMisdesignedfornonescalorin-plantmeteringofnaturalgasand
approved according the European guidelines of 97/23/EG. The MQM Quantometer is suitable
tobeinstalledinhazardousareaofCategory2(Zone1)II2GcIICT4X.
TheMQMQuantometerissuitableforthefollowinggases:
•Naturalgas,towngas,propane,butane,ethylene,air,nitrogen,noblegases
•Furthergasesonrequest.
TheMQMQuantometercannotbeusedfor:Oxygen,Acetyleneandotheraggressivegases
as well as for explosive gases and gas mixtures. The meter should not be used in applications
with the presence of strong pulsations. These conditions will result in an increased error and
maysignicantlydecreasethemeter’slifetime.
5. Design of the MQM Quantometer
Zähleraufbau
1 Meter housing
2 Turbine rotor
3 Meter index head
4 Magnet coupling
5 Flowconditioner
6 Measuring unit housing
7Measuring unit
Note: In case of failure or damage:
Repairs are only allowed to be done
by qualied and authorized repair centers.
Fig. 1: Sectional view of MQM Quantometer
6
3
1
2
4
5
6
7

7
6. Technical Data
Meter type MQM Quantometer
Approval PED97/23/EC
Size G 10 to G 1000
Nominal diameter DN25(2“)toDN150(6“)
Pressure rating PN 16 and ANSI CLASS 150
Rangeability 20:1orhigherdependingontheoperationpressure
Temperature ranges
• gastemperature -20°Cto+60°C
• ambienttemperature -25°Cto+55°C
• storagetemperature -25°Cto+55°C
Housing material Aluminum
Max. operation pressure 20 barg
Accuracy Qmin ≤ Q < Qt:±3.0%
Qt ≤ Q ≤ Qmax:±1.5%
Qt=0,2Qmax
Meter index
Number of indexes 1
Number of digit rollers per index 8
Meterindexhead Composite(optional:Aluminum)
Protection class IP 65
Pulse generators
LF-pulse generator 1LF(reedcontact),add.1LFasoption
Contact type makecontact(normallyopen)
Max. contact load 10W
Max. switching voltage 50VDC/200VDC
Max. switching current 0.1 A / 0.5 A
Resistance 100 Ω in series to reed contact
Max. switching frequency fmax 500Hz
HF-pulse generator: 1HF(inductivesensor,acc.NAMURspecs)
Supplyvoltage 8.2VDC(internalresistanceof1kΩ)
Currentconsumption(sensorcovered) <1mA
Currentconsumption(sensorfree) >2.1mA
Max. switching frequency fmax 1500Hz
Anti tampering contact 1 Reed contact in series with 100 Ω resistor
Contacttype makecontact(normallyclosed)
Max.contactload 10W
Max.switchingvoltage 175VDC
Max. switching current 0.25 A
Resistance 100 Ω in series to reed contact

Rangeability and pulse weighting data
*)Theabsolutenumberofthepulsesdependsonthemetersizeandtheindividualmeteritself.
Thestatedvaluesareoftypicalsize.Exactvaluesdeterminedbycalibrationofthemeterare
located on the nameplate.
7. Installation
Recommendations and requirements for the installation of the MQM Quantometer:
•Preferably,themetershouldbeinstalledhorizontally,withindexheadupward.
•Straightinletpipe≥5DNandoutletpipe≥1DNinthenominalpipediameter
is recommended for high accuracy.
•Ensure that the gas ow is in the same direction as shown by the arrow
at the meter housing.
•Avoid heavy vibrations and pulsations in the point of measurement.
Installationofasuitablecompensatorcaneectivelydampdownthiskindofdisturbance.
•Drasticchangesinthegasowaswellasstronglypulsatingowwillincreasethe
measuring error. They arise for example during frequently switching on-o of
solenoid valves. The pressure impulses by opening the valve very fast may damage
the ball bearings of the turbine rotor and cause failure of the Quantometer.
•Filter(smallmeshsize)shouldbeinstalledupstreamofthemetertoprotectitfrom
anycontamination(suspendedparticles>50µm)duringstart-upandoperation.
8
DN [mm] G-Type Qmin [m3/h] Qmax [m3/h] HF *) [imp/m3] NF [imp/m3]
25 10 2 16 140000 10
25 16 2,5 25 140000 10
25 25 4 40 140000 10
50 40 6,5 65 10400 10
50 65 10 100 10400 10
80 100 8 160 27000 1
80 160 12,5 250 27000 1
80 250 20 400 27000 1
100 160 12,5 250 13500 1
100 250 20 400 13500 1
100 400 32 650 13500 1
150 400 32 650 5400 1
150 650 50 1000 5400 1
150 1000 80 1600 5400 1

•The gas should be dry. If condensate may be occur then the installation
of a condensation trap upstream of the meter is recommended
(insomeltertypesitisintegrated).Inthiscasechoose
o Flowdirectionfromtoptobottom
o DoNotinstallthemeteratthelowestpointofthepiping
Attention: The MQM Quantometer is not suitable for oxygen and other highly
corrosive gases as well as for explosive gases and gas mixtures.
8. Mounting Position / Direction of Flow
There are two positions of installation possible for the turbine gas meter:
Horizontal installation or vertical installation
Figure 2: Typical horizontal installation of the MQM Quantometer (recommended)
Figure 3: Typical vertical installation of the MQM turbine gas meter (only DN ≤ 150)
Notice: Ensure that the gas ow is in the same direction as shown by the arrow at the meter housing.
9
By-pass
Isolation valve
Isolation valve
Isolation
valve
Filter
MQM
Ventil
By-pass
Isolation valve
Isolation
valve
Filter
MQM
Isolation valve

10
9. Preliminary work
Before the installation of the meter please make sure that the following steps
has been observed:
1. Remove the plastic cover sheets the inlet and outlet of the meter.
2.Inspectthemeterforcompleteness(e.g.plugs,providedoilvolume)
and for absence of any transport damages.
3. Cleantheangesurfaces(e.g.withpetroleum).
4. Checkbyblowingintothemeterthatturbinerotorrotateseasilyandstopssmoothly.
5. Priortothemeter’sinstallationcheckthatnocontaminations,condensate,dirtanddebris
are in the pipe. Especially the cleaning of elbows in vertical installations is important
topreventdamageofthemeterbycontaminations,dirtanddebrispassing.Acone-shaped
start-upsievemaybeinstalledforinitialoperation.After4-6weeksitshouldberemoved
toavoidperturbationsoftheowprolewithresultingdecreaseoftheaccuracy.
6. Makesurethatthereisnointrusionofthesealingelementintothepipeduringinstallation
andoperation(concentricmounting).
7. Checkthatinadditiontothemeter-lengththereisenoughinstallationspacefor
the two seals between inlet and outlet anges.
8. The meter must be installed stress-free. Therefore in addition to point 7 observe strictly the
properalignmentoftheQuantometer’sinlettooutletaxiswiththeconnectingpipingaxis.
Necessary equipment fort he meter installation:
1.Suitable and certied gaskets:
•
DimensionsaccordingDINEN1514-1orDIN2690
2. ScrewsaccordingDINISO4014andASMEB1.1
3. NutsaccordingDINENISO4032andASMEB1.1
DN Pressure Flanges according ANSI B 16.5
[mm] ANSI Class Stud bolts with continuous thread and nuts
Min. Operating Number Strength Strength
temperatur of of screws class class
the screws screws screws
and nuts
25 150 -25°C 4xM12 ASTMA193 ASTMA194
Grade B7 Grade 2H
50 150 -25°C 4xM16 ASTMA193 ASTMA194
Grade B7 Grade 2H
80 150 -25°C 8xM16 ASTMA193 ASTMA194
Grade B7 Grade 2H
100 150 -25°C 8xM16 ASTMA193 ASTMA194
Grade B7 Grade 2H
150 150 -25°C 8xM20 ASTMA193 ASTMA194
Grade B7 Grade 2H