Hach LBOD101 User manual

DOC022.53.80025
LBOD101
05/2021, Edition 4
User Manual


Table of Contents
Section 1 Product overview................................................................................. 3
Section 2 Specifications........................................................................................ 3
Section 3 Safety information............................................................................... 4
3.1 Intended use ....................................................................................................... 4
3.2 Use of hazard information..................................................................................... 4
3.3 Precautionary labels............................................................................................. 5
3.4 Product hazards.................................................................................................... 5
Section 4 Preparation for use.............................................................................. 5
Section 5 Calibration.............................................................................................. 6
5.1 Calibration notes................................................................................................... 6
5.2 LBOD water-saturated air (100%) calibration procedure...................................... 6
5.3 Zero-point calibration procedure........................................................................... 7
Section 6 Sample measurement ...................................................................... 8
6.1 Sample measurement notes................................................................................. 8
6.2 Sample measurement procedure......................................................................... 8
6.3 Low-level measurements...................................................................................... 9
Section 7 Maintenance........................................................................................... 9
7.1 Clean the probe.................................................................................................... 9
7.2 Replace the sensor cap and iButton..................................................................... 9
7.3 Replace the stirrer.............................................................................................. 12
7.4 Storage............................................................................................................... 12
Section 8 Troubleshooting................................................................................. 13
Section 9 Consumables....................................................................................... 14
1

Table of Contents
2

Section 1 Product overview
The Intellical LBOD101 probe is a digital, luminescent dissolved oxygen sensor that measures the
dissolved oxygen concentration of BOD (biochemical oxygen demand) samples. The probe has
temperature and absolute air pressure sensors for accurate dissolved oxygen measurements. The
probe stays on standard BOD bottles and stirs the sample during measurements. Refer to Figure 1.
Figure 1 Probe overview
1 LBOD sensor cap 6 Stirrer on/off button
2 Thermistor 7 Power indicator LED
3 Stirrer assembly 8 Air pressure sensor module
4 Probe lens 9 iButton® compartment1
5 Probe body with stirrer assembly
Section 2 Specifications
Specifications are subject to change without notice.
Specifications Details
Probe type Luminescent dissolved oxygen (LDO) sensor with integrated stirring
system
Dissolved oxygen range 0.05 to 20.0 mg/L (ppm); 1 to 200% saturation
Dissolved oxygen accuracy ±0.05 mg/L for concentrations less than 10 mg/L O2
±0.1 mg/L for concentrations more than 10 mg/L O2
1iButton is a registered trademark of Maxim Integrated Products, Inc.
English 3

Specifications Details
% saturation resolution 0.1%
% saturation accuracy ±0.59% of reading
Stabilization time T90% at 10 seconds (when the solution is stirred)
Temperature resolution 0.1 °C (0.18 °F)
Temperature accuracy ±0.3 °C (±0.54 °F)
Pressure resolution 5 mbar (0.5 kPa)
Pressure accuracy ±3 mbar (0.3 kPa)
Operating temperature 0 to 50 °C (32 to 122 °F)
Storage temperature 5 to 40 °C (41 to 104 °F)
Minimum immersion depth 10 mm (0.394 in.)
Dimensions Diameter: 15.875 mm (0.625 in.)
Length: 215 mm (8.46 in.)
Cable length: 1.8 m (5.9 ft)
Cable connection M12 digital output and connector
Warranty 1 year on the probe. This warranty covers manufacturing defects, but
not improper use or wear.
Certifications CE, FCC/ISED
Section 3 Safety information
3.1 Intended use
The Intellical probes are intended for use by individuals who measure water quality parameters in the
laboratory or in the field. The Intellical probes do not treat or alter water.
3.2 Use of hazard information
D A N G E R
Indicates a potentially or imminently hazardous situation which, if not avoided, will result in death or
serious injury.
WARNING
Indicates a potentially or imminently hazardous situation which, if not avoided, could result in death
or serious injury.
CAUTION
Indicates a potentially hazardous situation that may result in minor or moderate injury.
N O T I C E
Indicates a situation which, if not avoided, may cause damage to the instrument. Information that
requires special emphasis.
4 English

3.3 Precautionary labels
Read all labels and tags attached to the instrument. Personal injury or damage to the instrument
could occur if not observed. A symbol on the instrument is referenced in the manual with a
precautionary statement.
Electrical equipment marked with this symbol may not be disposed of in European
domestic or public disposal systems. Return old or end-of-life equipment to the
manufacturer for disposal at no charge to the user.
3.4 Product hazards
CAUTION
Chemical exposure hazard. Obey laboratory safety procedures and wear all of the
personal protective equipment appropriate to the chemicals that are handled. Refer to
the current safety data sheets (MSDS/SDS) for safety protocols.
CAUTION
Chemical exposure hazard. Dispose of chemicals and wastes in accordance with local,
regional and national regulations.
Section 4 Preparation for use
Prepare the probe for calibration and measurement as follows. Do not touch the protective black
layer on the LBOD sensor cap.
1. Open the pressure sensor module cap and make sure that the label on the iButton is on top.
2. Rinse the probe with deionized water. Blot dry with a lint-free cloth.
3. Make sure that the meter has the correct date and time settings. The service-life time stamp in
the probe comes from the date and time settings in the meter.
Note: Some meters automatically open the date and time settings when the meter starts for the first time, or
after battery replacement.
4. Connect the probe to the meter.
English 5

Section 5 Calibration
The procedure that follows is applicable to meters that can connect to Intellical LBOD probes. Refer
to the applicable meter documentation for meter operation and probe-specific settings.
5.1 Calibration notes
Read the notes that follow before calibration.
• Measure samples as soon as possible after collection.
• Use the single display mode for calibration when more than one probe is connected to the meter (if
applicable).
• Calibrate the probes and verify the calibration regularly for best results. Use the meter to set
calibration reminders.
• The calibration data is stored in the probe. When a calibrated probe is connected to a different
meter with the same calibration options, a new calibration is not necessary.
• Air bubbles below the sensor when in solution can cause a slow response or error in the
calibration. Make sure to remove air bubbles during calibration.
• The meter uses the slope value shown at the end of the calibration to monitor the condition of the
sensor cap.
5.2 LBOD water-saturated air (100%) calibration procedure
1. Fill a BOD bottle
approximately ¾ full
with water.
2. Put a stopper on
the bottle and shake
for 30 seconds.
3. Rinse the probe
with deionized water.
Dry the probe with a
lint-free cloth.
4. Carefully dry the
LBOD sensor cap
with a non-abrasive
cloth.
5. Put the probe in
the bottle. Wait
approximately
10 minutes for the
contents to adjust to
ambient
temperature.
6. Go to the
calibrate menu.
Select the probe, if
applicable.
7. Read the
dissolved oxygen
value. The display
shows 100% when
the reading is stable.
8. Save the
calibration.
6 English

5.3 Zero-point calibration procedure
A zero-point calibration can increase the measurement accuracy of samples that have less than
1 mg/L O2. A zero point calibration is not recommended for samples that have more than 1 mg/L O2.
Set the calibration mode to "100% with 0". The meter starts the calibration with the 100%
measurement, then continues to the zero-point measurement.
1. Fill a BOD bottle
with deionized water.
2. Add 300 mg of
sodium sulfite to the
bottle.
3. Add 2 mL of
cobalt chloride
solution to the bottle.
4. Put a stopper on
the bottle and fully
mix for 30 seconds.
5. Remove the
stopper. Put the
probe in the bottle.
6. Tap the probe to
remove air bubbles.
7. Wait
approximately
10 minutes for the
contents to adjust to
ambient
temperature.
8. Start the stirrer.
9. Read the
dissolved oxygen
value. Wait for the
measurement to
become stable.
10. Save the
calibration.
11. Rinse the probe
with deionized water.
Dry the probe with a
lint-free cloth.
12. Fill a BOD bottle
with deionized water.
Put the probe in the
bottle and stir for
10 minutes to fully
remove the sulfite
from the probe.
English 7

Section 6 Sample measurement
The procedure that follows is applicable to meters that can connect to Intellical LBOD probes. Refer
to the applicable meter documentation for meter operation and probe-specific settings.
6.1 Sample measurement notes
Read the notes that follow before sample measurements.
• Salinity changes the solubility of oxygen in water. Measure the sample salinity and enter the value
in the probe settings of the meter.
• High concentrations (more than 1 molar) of acids or bases will decrease the service life of the
LBOD sensor cap.
• Rinse the probe with deionized water and dry with a lint-free cloth between measurements to
prevent contamination.
• If complete traceability is necessary, enter a sample ID and operator ID before measurement.
Refer to the meter manual for instructions.
• The meter automatically saves the measurement data when the user manually reads each data
point and when the meter is set to read at regular intervals. The user must manually save each
data point when the meter is set to read continuously.
• Air bubbles below the sensor can cause a slow response or error in the measurement. Make sure
to remove air bubbles before and during measurements.
6.2 Sample measurement procedure
1. Rinse the probe
with deionized water.
Dry the probe with a
lint-free cloth.
2. Put the probe in
the BOD sample.
3. Tap the probe to
remove air bubbles.
4. Start the stirrer.
5. Read the
dissolved oxygen
value of the sample.
The display shows
the dissolved oxygen
value when the
reading is stable.
8 English
Table of contents
Other Hach Measuring Instrument manuals

Hach
Hach Polymetron 9523sc pH Quick manual

Hach
Hach 500sc SiO2 Instruction Manual

Hach
Hach orbisphere C1100 User manual

Hach
Hach POLYMETRON 9135 User manual

Hach
Hach Hach SL1000 Quick manual

Hach
Hach TU5300 sc Quick manual

Hach
Hach 9186 User manual

Hach
Hach TL2360 Quick manual

Hach
Hach ISENO3181 User manual

Hach
Hach BioTector B3500e User manual

Hach
Hach DOC026.52.00769 User manual

Hach
Hach LANGE DR 900 User manual

Hach
Hach sensION+ EC5 User manual

Hach
Hach POLYMETRON 8810 ISE Quick manual

Hach
Hach CL17sc User manual

Hach
Hach BioTector B3500s User manual

Hach
Hach Lange APA 6000 Operating instructions

Hach
Hach sensION+ DL User manual

Hach
Hach 911 User manual

Hach
Hach PHOSPHAX sigma High Range Quick manual