Hach 5500sc User manual

DOC023.98.80473
5500sc
Ammonia/Monochloramine
05/2017, Edition 4
Installation
Installation
Instalación
Монтаж

English..............................................................................................................................3
Français......................................................................................................................... 18
Español.......................................................................................................................... 35
Русский..........................................................................................................................52
2

Table of contents
Specifications on page 3
General information on page 4
Installation on page 8
Specifications
Specifications are subject to change without notice.
Table 1 General specifications
Specification Details
Dimensions (W x D x H) 452 x 360 x 804 mm (17.8 x 14.2 x 31.7 in.)
Enclosure Rating: NEMA 4x/IP56
Material: PC/ABS case, PC door, PC hinges and latches, 316 SST hardware
Indoor use only. Keep away from direct sunlight.
Weight 20.5 kg (45.2 lb) without reagents and standards, 30 kg (66.1 lb) with reagents,
standards and cleaning solution
Mounting Wall, panel or table
Protection class I
Pollution degree/installation
category
2/II
Power requirements AC: 100–240 VAC, 50/60 Hz
Instrument: 0.5 A nominal, 8.3 A maximum; accessory output 100–240 Vac,
5.0 A maximum
Connection: 1.0 to 1.3 mm2 (18 to 16 AWG) wire, 1.0 mm2 (18 AWG) stranded
recommended; field wiring insulation must be rated to 65°C (149 ºF) minimum
Operating temperature 5 to 45 °C (41 to 113 °F)
Operating humidity 5 to 95% non-condensing
Storage temperature –20 to 60 °C (–4 to 140 °F)
4–20 mA outputs Four; load impedance: 600 Ω maximum
Connection: 0.644 to 1.29 mm2 (22 to 16 AWG) wire, 0.644 to 0.812 mm2
(22 to 20 AWG) recommended, twisted pair shielded wire
Alarm relay outputs Four; type: not powered SPDT relays, each rated at 5 A resistive, 240 VAC
maximum
Connection: 1.0 to 1.29 mm2 (18 to 16 AWG) wire, 1.0 mm2 (18 AWG)
stranded recommended
Digital inputs Four; connection: 0.644 to 1.29 mm2 (22 to 16 AWG) wire, 0.644 to 0.812 mm2
(22 to 20 AWG) stranded (isolated DC voltage input or an open-collector/relay
contact closure input) recommended
Fuses Input power—AC: T 1.6 A, 250 VAC; DC: T 6.3 A, 250 VAC
Output power—AC: T 5.0 A, 250 VAC; DC: T 1.6 A, 250 VAC
Alarm relay outputs: T 5.0 A, 250 V
Fittings Sample line and sample bypass drain: 6 mm OD push-to-connect fitting for
plastic tubing
Chemical and case drains: 11 mm (7/16 in.) ID slip-on fitting for soft plastic
tubing
English 3

Table 1 General specifications (continued)
Specification Details
Sample pressure, flow rate, and
temperature
Pressure: 17.2 to 600 kPa (2.5 to 87 psi) to preset pressure regulator
Flow rate: 100 to 1000 mL/minute
Temperature: 5 to 50 °C (41 to 122 °F)
Number of sample streams 1 or 2 with programmable sequence
Certifications ETL certified to UL and CSA standards, CE marked
Table 2 Measurement specifications
Specification Details
Light source LED (light emitting diode) with a peak wavelength of 650 nm
Measurement range 0.01 to 2 ppm as N, 0.05 to 10 ppm as Cl2
Accuracy ± 5% or 0.01 ppm as N or 0.05 ppm as Cl2 (the larger value) for 5 to 40 °C
(41 to 104 °F)
±10% or 0.02 as N or 0.10 ppm as Cl2 (the larger value) for 40 to 50 °C
(104 to 122 °F)
Precision/Repeatability 3% or 0.01 ppm as N or 0.05 ppm as Cl2 (the larger value)
Response time Within 5 minutes
Stabilization time After initial startup or annual maintenance: 10 measurement cycles
After standby: 1 measurement cycle
After calibration: 1 measurement cycle
Calibration time Automatic slope and offset calibration: 50 minutes
Minimum detection limit 0.01 ppm as N or 0.05 ppm as Cl2
Reagent usage Usage: 1 L of reagent every month with a 5-minute cycle time
Container: 1 L, HDPE with polypropylene caps
Standard usage and cleaning
solutions
Usage: 2 L of each standard and cleaning solution monthly
Container: 2 L, PETE with polypropylene caps
General information
In no event will the manufacturer be liable for direct, indirect, special, incidental or consequential
damages resulting from any defect or omission in this manual. The manufacturer reserves the right to
make changes in this manual and the products it describes at any time, without notice or obligation.
Revised editions are found on the manufacturer’s website.
Safety information
N O T I C E
The manufacturer is not responsible for any damages due to misapplication or misuse of this product including,
without limitation, direct, incidental and consequential damages, and disclaims such damages to the full extent
permitted under applicable law. The user is solely responsible to identify critical application risks and install
appropriate mechanisms to protect processes during a possible equipment malfunction.
Please read this entire manual before unpacking, setting up or operating this equipment. Pay
attention to all danger and caution statements. Failure to do so could result in serious injury to the
operator or damage to the equipment.
Make sure that the protection provided by this equipment is not impaired. Do not use or install this
equipment in any manner other than that specified in this manual.
4 English

Use of hazard information
D A N G E R
Indicates a potentially or imminently hazardous situation which, if not avoided, will result in death or serious injury.
WARNING
Indicates a potentially or imminently hazardous situation which, if not avoided, could result in death or serious
injury.
CAUTION
Indicates a potentially hazardous situation that may result in minor or moderate injury.
N O T I C E
Indicates a situation which, if not avoided, may cause damage to the instrument. Information that requires special
emphasis.
Precautionary labels
Read all labels and tags attached to the instrument. Personal injury or damage to the instrument
could occur if not observed. A symbol on the instrument is referenced in the manual with a
precautionary statement.
This is the safety alert symbol. Obey all safety messages that follow this symbol to avoid potential
injury. If on the instrument, refer to the instruction manual for operation or safety information.
This symbol indicates the need for protective eye wear.
This symbol identifies a risk of chemical harm and indicates that only individuals qualified and trained
to work with chemicals should handle chemicals or perform maintenance on chemical delivery
systems associated with the equipment.
This symbol indicates that a risk of electrical shock and/or electrocution exists.
This symbol indicates that the marked item can be hot and should not be touched without care.
This symbol indicates that a risk of fire is present.
This symbol identifies the presence of a strong corrosive or other hazardous substance and a risk of
chemical harm. Only individuals qualified and trained to work with chemicals should handle chemicals
or perform maintenance on chemical delivery systems associated with the equipment.
Electrical equipment marked with this symbol may not be disposed of in European domestic or public
disposal systems. Return old or end-of-life equipment to the manufacturer for disposal at no charge to
the user.
Certification
Canadian Radio Interference-Causing Equipment Regulation, IECS-003, Class A:
English 5

Supporting test records reside with the manufacturer.
This Class A digital apparatus meets all requirements of the Canadian Interference-Causing
Equipment Regulations.
Cet appareil numérique de classe A répond à toutes les exigences de la réglementation canadienne
sur les équipements provoquant des interférences.
FCC Part 15, Class "A" Limits
Supporting test records reside with the manufacturer. The device complies with Part 15 of the FCC
Rules. Operation is subject to the following conditions:
1. The equipment may not cause harmful interference.
2. The equipment must accept any interference received, including interference that may cause
undesired operation.
Changes or modifications to this equipment not expressly approved by the party responsible for
compliance could void the user's authority to operate the equipment. This equipment has been tested
and found to comply with the limits for a Class A digital device, pursuant to Part 15 of the FCC rules.
These limits are designed to provide reasonable protection against harmful interference when the
equipment is operated in a commercial environment. This equipment generates, uses and can
radiate radio frequency energy and, if not installed and used in accordance with the instruction
manual, may cause harmful interference to radio communications. Operation of this equipment in a
residential area is likely to cause harmful interference, in which case the user will be required to
correct the interference at their expense. The following techniques can be used to reduce
interference problems:
1. Disconnect the equipment from its power source to verify that it is or is not the source of the
interference.
2. If the equipment is connected to the same outlet as the device experiencing interference, connect
the equipment to a different outlet.
3. Move the equipment away from the device receiving the interference.
4. Reposition the receiving antenna for the device receiving the interference.
5. Try combinations of the above.
Product overview
D A N G E R
Chemical or biological hazards. If this instrument is used to monitor a treatment process and/or
chemical feed system for which there are regulatory limits and monitoring requirements related to
public health, public safety, food or beverage manufacture or processing, it is the responsibility of the
user of this instrument to know and abide by any applicable regulation and to have sufficient and
appropriate mechanisms in place for compliance with applicable regulations in the event of malfunction
of the instrument.
The analyzer measures total ammonia and monochloramine in drinking water and calculates free
ammonia concentration. The chemical analysis uses a modified phenate method to measure
monochloramine values by colorimetry. An excess of hypochlorite at the correct pH generates total
ammonia values measured by colorimetry. Then, the analyzer calculates the free ammonia values
out of the difference between the measured parameters.
Refer to Figure 1 for the analyzer overview. The doors can be easily removed for better access.
Refer to Figure 2. Make sure that the doors are installed, closed and securely latched to keep the
enclosure environmental rating.
6 English

Figure 1 Product overview
1 Upper and lower doors 5 Display and keypad 9 Analytics panel
2 Funnel cover 6 SD card slot 10 Reagent bottle tray
3 Grab sample input funnel 7 Power switch111 Grab sample out fitting
4 Status indicator light 8 Power LED (on = analyzer is on) 12 Grab sample out cover
Figure 2 Door removal
1Open the upper door and the analytics panel. The power switch is inside on the far right side on
the back of the analyzer.
English 7

Installation
WARNING
Multiple hazards. Only qualified personnel must conduct the tasks described in this section of the
document.
Mechanical installation
D A N G E R
Risk of injury or death. Make sure that the wall mounting is able to hold 4 times the weight of the
equipment.
WARNING
Personal injury hazard.
Instruments or components are heavy. Use assistance to install or move.
The object is heavy. Make sure that the instrument is securely attached to a wall, table or floor for a
safe operation.
This instrument is rated for an altitude of 2000 m (6562 ft) maximum. Use of this instrument at an
altitude higher than 2000 m can slightly increase the potential for the electrical insulation to break
down, which can result in an electric shock hazard. The manufacturer recommends that users with
concerns contact technical support.
Install the analyzer indoors, in a non-hazardous environment. Refer to the supplied mounting
documentation.
Plumbing
D A N G E R
Fire hazard. This product is not designed for use with flammable liquids.
N O T I C E
Do not install reagents until all plumbing is complete.
Make sure to use the specified tubing size.
Sample line guidelines
Select a good, representative sampling point for the best instrument performance. The sample must
be representative of the entire system.
To prevent erratic readings:
• Collect samples from locations that are sufficiently distant from points of chemical additions to the
process stream.
• Make sure that the samples are sufficiently mixed.
• Make sure that all chemical reactions are complete.
Connect the sample stream
Install each sample line into the center of a larger process pipe to minimize interference from air
bubbles or bottom sediment. Figure 3 shows examples of good and bad installation.
Keep the sample lines as short as possible to prevent the accumulation of bottom sediment. The
sediment can absorb some of the analyte from the sample and cause low readings. The sediment
can later release the analyte and cause high readings. This exchange with the sediment also causes
a delayed response when the analyte concentration in the sample increases or decreases.
8 English

Figure 3 Sampling methods
1 Air 2 Sample flow
Plumbing access ports
Make plumbing connections through the plumbing access ports. Refer to Figure 4. To keep the
enclosure rating, make sure that plugs are installed in the plumbing ports that are not used.
Figure 4 Ports for one or two sample streams
1 Sample 2 bypass drain for two
sample stream analyzers
5 Chemical drain 9 Case drain for leaks or spills
2 Not used 6 Rear of analyzer 10 Sample 1 bypass drain
3 Sample 2 inlet for two sample
stream analyzers
7 Sample holder drain
4 Sample 1 inlet 8 Sample holder overflow
Plumb the sample and drain lines
CAUTION
Explosion hazard. Use only the supplied regulator from the manufacturer.
English 9

CAUTION
Chemical exposure hazard. Dispose of chemicals and wastes in accordance with local, regional and
national regulations.
N O T I C E
Do not connect the drain lines to other lines because backpressure or damage to the analyzer can occur. Make
sure that the drain lines are open to air.
N O T I C E
To prevent backpressure and damage to the analyzer, make sure that the analyzer is higher than the facility
drain(s) used and that the drain line has a constant downward slope.
Use the supplied tubing (6 mm), shut-off valve, Y-strainer with filter and pressure regulator to plumb
the drain and the sample to the analyzer. Refer to Figure 5 and Figure 6. The sample line tubing that
goes into the plumbing access ports must be 6 mm. Tubing of 1/4 in. can be used for the sample line
up to the valve or to the y-strainer, but not into the plumbing access ports of the analyzer. For correct
drainage, install the drain lines at least with a 1-inch decrease in height per 1 foot of tubing length.
Figure 5 Plumb the sample
1 Sample in (single-stream) 3 Shut-off valve 5 Adjustable pressure regulator
2 Rear side of the analyzer 4 Y-strainer with filter
10 English
Other manuals for 5500sc
3
Table of contents
Languages:
Other Hach Measuring Instrument manuals

Hach
Hach K1100 Quick manual

Hach
Hach TU5300 sc Quick manual

Hach
Hach ISENA381 Series User manual

Hach
Hach sensION3 User manual

Hach
Hach 911 User manual

Hach
Hach sc200 User manual

Hach
Hach TL2350 Quick manual

Hach
Hach DR 3900 Quick manual

Hach
Hach POCKET COLORIMETER II User manual

Hach
Hach TL2360 Quick manual

Hach
Hach LANGE LICO 500 User manual

Hach
Hach A1000 Quick manual

Hach
Hach sensION+ EC71 User manual

Hach
Hach Micro Dist User guide

Hach
Hach POCKET COLORIMETER II LCK 360 Zinc User guide

Hach
Hach HQ440d Quick manual

Hach
Hach DR 1900 Quick manual

Hach
Hach CLF10sc User manual

Hach
Hach HQd Quick manual

Hach
Hach ORBISPHERE G1100 User manual