Hach 2100A User manual

DOC022.97.80041
2100Q and 2100Qis
01/2010, Edition 1
Basic User Manual
Manuel d'utilisation de base
Manual básico del usuario
Manual básico do usuário
ᴀ⫼᠋ݠ
ၮᧄ࡙ࠩࡑ࠾ࡘࠕ࡞
₆⽎G㌂㣿G㍺ⳛ㍲

English..................................................................................................................................................................................................3
Français.............................................................................................................................................................................................16
Español..............................................................................................................................................................................................30
Português..........................................................................................................................................................................................44
中文......................................................................................................................................................................................................58
日本語.................................................................................................................................................................................................70
한글......................................................................................................................................................................................................83
2

Specifications
Specifications are subject to change without notice.
Specification Details
Measurement
method
Ratio turbidimetric determination using a primary
nephelometric light scatter signal (90°) to the transmitted light
scatter signal.
Regulatory 2100Q: Meets EPA Method 180.1
2100Qis: Meets ISO 7027
Lamp source 2100Q: Tungsten filament lamp
2100Qis: Light-emitting diode (LED) at 860 nm
Range 0–1000 NTU (FNU)
Accuracy ±2% of reading plus stray light from 0–1000 NTU (FNU)
Repeatability ±1% of reading or 0.01 NTU (FNU), whichever is greater
Resolution 0.01 NTU on lowest range
Stray light ≤ 0.02 NTU (FNU)
Signal averaging Selectable on or off
Detector Silicon Photodiode
Reading modes Normal (Push to Read), Signal Averaging or Rapidly Settling
Turbidity™
Calibration
options
Single step RapidCal™ for Low-Level Regulatory Reporting
from 0–40 NTU (FNU)
Full range calibration from 0–1000 NTU (FNU)
Calibration to degrees of turbidity
Calibration logger Records the last 25 successful calibrations
Verification logger Logs the last 250 successful verifications
Data logger 500 records
Specification Details
Power
requirement
AC 100–240 V , 50/60 Hz (with power or USB/power module)
4 AA alkaline batteries
Rechargeable NiMH (for use with USB/power module)
Operating
conditions
Temperature: 0 to 50 °C (32 to 122 °F)
Relative Humidity: 0–90% at 30 °C, 0–80% at 40 °C, 0–70%
at 50 °C, noncondensing
Storage
conditions
–40 to 60 °C (–40 to 140 °F), instrument only
Interface Optional USB
Sample required 15 mL (0.5 oz.)
Sample cells Round cells 60 x 25 mm (2.36 x 1 in.) borosilicate glass with
screw caps
Dimensions 22.9 x 10.7 x 7.7 cm (9.0 x 4.2 x 3.0 in.)
Weight 530 g (1.17 lb) without batteries
620 g (1.37 lb) with four AA alkaline batteries
Meter enclosure
rating
IP67 (closed lid, battery and module compartment excluded)
Protection class Power supply: Class II
Certification CE certified
Warranty 1 year (EU: 2 years)
General Information
In no event will the manufacturer be liable for direct, indirect, special,
incidental or consequential damages resulting from any defect or omission
in this manual. The manufacturer reserves the right to make changes in
this manual and the products it describes at any time, without notice or
obligation. Revised editions are found on the manufacturer’s website.
English 3

Safety information
Please read this entire manual before unpacking, setting up or operating
this equipment. Pay attention to all danger and caution statements. Failure
to do so could result in serious injury to the operator or damage to the
equipment.
Make sure that the protection provided by this equipment is not impaired,
do not use or install this equipment in any manner other than that specified
in this manual.
Use of hazard information
D A N G E R
Indicates a potentially or imminently hazardous situation which, if not avoided, will
result in death or serious injury.
WARNING
Indicates a potentially or imminently hazardous situation which, if not avoided,
could result in death or serious injury.
C A U T I O N
Indicates a potentially hazardous situation that may result in minor or moderate
injury.
N O T I C E
Indicates a situation which, if not avoided, may cause damage to the instrument.
Information that requires special emphasis.
Precautionary labels
Read all labels and tags attached to the instrument. Personal injury or
damage to the instrument could occur if not observed. A symbol, if noted
on the instrument, will be included with a danger or caution statement in
the manual.
This is the safety alert symbol. Obey all safety messages that follow
this symbol to avoid potential injury. If on the instrument, refer to the
instruction manual for operation or safety information.
This symbol indicates that a risk of electrical shock and/or
electrocution exists.
Electrical equipment marked with this symbol may not be disposed
of in European public disposal systems after 12 August of 2005. In
conformity with European local and national regulations (EU
Directive 2002/96/EC), European electrical equipment users must
now return old or end-of life equipment to the Producer for disposal
at no charge to the user.
Note: For return for recycling, please contact the equipment producer or supplier
for instructions on how to return end-of-life equipment, producer-supplied
electrical accessories, and all auxiliary items for proper disposal.
Certification
Canadian Radio Interference-Causing Equipment Regulation,
IECS-003, Class A:
Supporting test records reside with the manufacturer.
This Class A digital apparatus meets all requirements of the Canadian
Interference-Causing Equipment Regulations.
Cet appareil numèrique de la classe A respecte toutes les exigences du
Rëglement sur le matériel brouilleur du Canada.
FCC Part 15, Class "A" Limits
Supporting test records reside with the manufacturer. The device complies
with Part 15 of the FCC Rules. Operation is subject to the following
conditions:
1. The equipment may not cause harmful interference.
2. The equipment must accept any interference received, including
interference that may cause undesired operation.
Changes or modifications to this equipment not expressly approved by the
party responsible for compliance could void the user's authority to operate
the equipment. This equipment has been tested and found to comply with
the limits for a Class A digital device, pursuant to Part 15 of the FCC rules.
These limits are designed to provide reasonable protection against
harmful interference when the equipment is operated in a commercial
4 English

environment. This equipment generates, uses and can radiate radio
frequency energy and, if not installed and used in accordance with the
instruction manual, may cause harmful interference to radio
communications. Operation of this equipment in a residential area is likely
to cause harmful interference, in which case the user will be required to
correct the interference at their expense. The following techniques can be
used to reduce interference problems:
1. Disconnect the equipment from its power source to verify that it is or
is not the source of the interference.
2. If the equipment is connected to the same outlet as the device
experiencing interference, connect the equipment to a different outlet.
3. Move the equipment away from the device receiving the interference.
4. Reposition the receiving antenna for the device receiving the
interference.
5. Try combinations of the above.
Product overview
The 2100Q and 2100Qis portable turbidimeters measure turbidity from 0
to 1000 NTU (FNU). Primarily for field use, the portable meter operates on
four AA batteries. Data can be stored and transferred to a printer, computer
or USB storage device.
Figure 1 Product overview
1Power on or off 5Alignment arrow
2Backlight keys (+ and -) 6Module
3Sample cell holder with lid 7Lamp compartment
4Attachment for lanyard 8Battery compartment
Product components
Refer to Figure 2 to make sure that all components have been received.
If any of these items are missing or damaged, contact the manufacturer
or a sales representative immediately.
English 5

Figure 2 2100Q and 2100Qis components
12100Q or 2100Qis turbidimeter 6Silicone oil
2Carrying case 720, 100 and 800 NTU StablCal
calibration standards
3User manual, Quick reference guide
and CD-ROM
8AA alkaline batteries (pk/4)
4Oiling cloth 9StablCal 10 NTU verification
standard
51" sample cell (10 mL) with cap (pk/
6)
Installation
C A U T I O N
Personal injury hazard. Only qualified personnel should conduct the tasks
described in this section of the manual.
Install the battery
WARNING
Potential fire hazard. Use only alkaline or nickel metal hydride batteries (NiMH) in
the meter. Other battery types or incorrect installation can cause a fire. Never mix
battery types in the meter.
N O T I C E
The battery compartment is not waterproof. If the battery compartment becomes
wet, remove and dry the batteries and thoroughly dry the interior of the
compartment. Check the battery contacts for corrosion and clean them if
necessary.
N O T I C E
When using nickel metal hydride (NiMH) batteries, the battery icon will not indicate
a full charge after freshly charged batteries have been inserted (NiMH batteries
are 1.2 V versus 1.5 V for alkaline batteries). Even though the icon does not indicate
complete charge, 2500 mAH NiMH batteries will achieve 90% of instrument
operation lifetime (before recharge) versus new alkaline batteries.
The meter can be powered with AA alkaline or rechargeable NiMH
batteries. To conserve battery life, the meter will power off after 10 minutes
of inactivity, the backlight powers off after 30 seconds. This time can be
changed in the Power Management menu.
Note: Rechargeable batteries will only be recharged with the USB/power module.
Refer to the module documentation for further information.
For battery installation refer to Figure 3.
1. Remove the battery cover.
2. Install 4 AA alkaline or 4 AA nickel metal hydride (NiMH) batteries.
Make sure that the batteries are installed in the correct orientation.
3. Replace the battery cover.
6 English

Figure 3 Battery installation User interface and navigation
User interface
Figure 4 Keypad description
1SETTINGS key: select menu
options for setting up the meter
5UP key: scroll through menus, enter
numbers and letters
2CALIBRATION key: shows
calibration screen, start calibration,
select cal options
6RIGHT key (contextual): read
turbidity sample, selects or confirms
options, opens/jumps to sub-menus
3DOWN key: scroll through menus,
enter numbers and letters
7DATA MANAGEMENT key:
view, delete or transfer stored data
4LEFT key (contextual): access for
calibration verification, cancels or
exits the current menu screen to the
previous menu screen
English 7

Display description
The measurement screen shows the turbidity, unit, calibration status, date
and time, operator ID (if setup) and sample ID (if setup). Refer to
Figure 5.
Figure 5 Single screen display
1Operator identification 9NTU (Nephelometric Turbidity Unit)
or FNU (Formazin Turbidity Unit)
2Sample identification 10 Reading mode: Rapidly Settling
Turbidity (Target icon)
3Stability or display lock indicator 11 Reading mode: Signal Average (X-
bar icon)
4Calibration status indicator
(Calibration OK=pass)
12 Time
5Calibration status indicator
(Calibration ?=fail)
13 Date
6Parameter title 14 Read (contextual: OK, Select)
7AC power icon 15 Options (contextual)
8Battery icon 16 Verification calibration
Navigation
The meter contains a Settings menu, Reading Options menu, Calibration
Options menu and Calibration Verification Options menu to change
various options. Use the UP and DOWN keys to highlight different options.
Push the RIGHT key to select an option. There are two ways to change
options:
1. Select an option from a list: Use the UP and DOWN keys to select an
option. If check boxes are shown, more than one option can be
selected. Push the LEFT key under Select.
Note: To deselect check boxes, push the LEFT key under Deselect.
2. Enter an option value using the arrow keys:
Push the UP and DOWN keys to enter or change a value.
3. Push the RIGHT key to advance to the next space.
4. Push the RIGHT key under OK to accept the value.
Start-up
Turn the meter on and off
Push the ON/OFF key to turn on or turn off the meter. If the meter does
not turn on, make sure that the batteries, or the module, are properly
installed or that the AC power supply is properly connected to an electrical
outlet.
Note: The Auto-Shutoff option can also be used to turn off the meter. For additional
information, refer to the CD for an expanded version of this manual.
Change the language
There are three options to set the language:
• The display language is selected when the meter is powered on for the
first time.
• The display language is selected when the power key is pushed and
held.
• The language can be changed from the Settings menu.
1. Select a language from the list. Confirm with OK.
2. Push Done when the update is complete.
8 English

Change the date and time
The date and time can be changed from the Date & Time menu.
1. Push the SETTINGS key and select Date & Time.
2. Update the time and date information:
Option Description
Format Select one of the formats for the date and time:
yyyy-mm-dd 24h
yyyy-mm-dd 12h
dd-mm-yyyy 24h
dd-mm-yyyy 12h
mm/dd/yyyy 24h
mm/dd/yyyy 12h
Date Enter the current date
Time Enter the current time
The current date and time will be shown on the display.
After the date and time setup, the meter is ready to take a reading.
Standard operations
Use a sample ID
The sample ID tag is used to associate readings with a particular sample
location. If assigned, stored data will include this ID.
1. Select Sample ID in the Settings menu.
2. Select, create or delete a sample ID:
Option Description
Current ID Select an ID from a list. The current ID will be
associated with sample data until a different ID is
selected.
Option Description
Create a New
Sample ID
Enter a name for a new sample ID.
Delete Sample ID Delete an existing sample ID.
Use an operator ID
The operator ID tag associates readings with an individual operator. All
stored data will include this ID.
1. Select Operator ID in the Settings menu.
2. Select, create or delete an operator ID:
Option Description
Current ID Select an ID from a list. The current ID will be
associated with sample data until a different ID is
selected.
Create a New
Operator ID
Enter a name for a new operator ID (maximum 10
names can be entered).
Delete Operator ID Delete an existing operator ID.
Advanced operations
Calibrate the turbidimeter with StablCal® Standards
Note: For best accuracy use the same sample cell or four matched sample cells for
all readings during calibration. Insert the sample cell in the instrument cell
compartment so the diamond or orientation mark aligns with the raised orientation
mark in front of the cell compartment.
English 9

1. Push the
CALIBRATION key to
enter the Calibration
mode. Follow the
instructions on the
display.
Note: Gently invert each
standard before
inserting the standard.
2. Insert the 20 NTU
StablCal Standard and
close the lid.
Note: The standard to
be inserted is bordered.
3. Push Read. The
display shows
Stabilizing and then
shows the result.
4. Repeat Step 2 and 3
with the 100 NTU and
800 NTU StablCal
Standard.
Note: Push Done to
complete a 2 point
calibration.
5. Push Done to review
the calibration details.
6. Push Store to save
the results.
After a calibration is
complete, the meter
automatically goes into
the Verify Cal mode. For
additional information,
refer to the CD for an
expanded version of this
manual.
Reading modes
1. Push the UP or DOWN key to enter the Reading Options menu.
2. Select Reading Mode to select one of the following options:
Option Description
Normal (Default
setting)
The normal mode reads and averages three readings.
The result is shown after the reading.
Signal Average The Signal Average mode compensates for reading
fluctuations caused by drifting of sample particles
through the light path.
The X-bar icon is shown on the display when signal
averaging is on.
The Signal Average mode measures 12 times and starts
to show the average after three readings. The final result
is the average of all 12 readings.
Rapidly Settling
Turbidity™ (RST)
The Rapidly Settling Turbidity (RST) mode calculates
and continuously updates the turbidity reading of the
sample to a confidence of 95%, based on the
accumulated trend of the real time measured values.
The RST mode is best used on samples that settle
rapidly and continuously change in value. The reading
is based on a correctly prepared sample that is
homogeneous at the beginning of the reading. It is best
applied to samples that are greater than 20 NTU. The
sample must be mixed thoroughly by inversion
immediately before inserting it into the meter.
The target icon is shown on the display when the Rapidly
Settling Turbidity is on.
The Rapidly Settling Turbidity reads and calculates five
readings while showing intermediate results.
Maintenance
C A U T I O N
Personal injury hazard. Only qualified personnel should conduct the tasks
described in this section of the manual.
10 English
This manual suits for next models
2
Table of contents
Languages:
Other Hach Measuring Instrument manuals

Hach
Hach DR 1900 Quick manual

Hach
Hach sension ph31 User manual

Hach
Hach NA5600 sc Na+ Instruction Manual

Hach
Hach FilterTrak 660 User manual

Hach
Hach AF7000 Quick manual

Hach
Hach HQ440d Quick manual

Hach
Hach 8520600 User manual

Hach
Hach TOC-X5 User manual

Hach
Hach H125 User manual

Hach
Hach CLF10sc User manual

Hach
Hach Lange ORBISPHERE 3100 User manual

Hach
Hach 5500sc SiO2 User manual

Hach
Hach 5500sc User manual

Hach
Hach sensION3 Reference guide

Hach
Hach LANGE LOC 100 User manual

Hach
Hach POLYMETRON 9125 User manual

Hach
Hach AMTAX sc User manual

Hach
Hach CLF10sc User manual

Hach
Hach Lange APA 6000 User manual

Hach
Hach REFO 3 User manual