7
ENTER THE METER’S K-FACTOR UNIT – Directly af-
ter the METER size is selected, the display’s K-factor unit
must be chosen. Use the UP arrow key to select your K-factor
unit. For meters calibrated in gallons, use PUL/GAL (pulses
per gallon), for meters calibrated in cubic meters, use PUL/
M3 (pulses per cubic meter), etc. Press ENTER to save the K-
factor unit and advance to the next parameter.
ENTER THE METER’S K-FACTOR – To change the K-
FACTOR value, use the RIGHT arrow key and select the po-
sition of the number that you wish to change. Using the UP
arrow key, increment the display digit until it matches the me-
ter’s K-factor digit. Repeat this process until all K-factor dig-
its have been entered. Press ENTER once to save the K-factor
and advance to the RATE/TOTAL Units selection.
SELECT THE RATE/TOTAL UNITS OF MEASURE –
The monitor allows the choice of five common rate/total
units. The monitor shows the rate/total unit that the display is
currently set for. If the current selection is current, press the
ENTER key once to advance to the next parameter. To change
to an alternate unit, use the arrow keys to scroll to the desired
rate unit and press ENTER to save the choice.
Note: The meter connection size and the bore size are dif-
ferent. For example, many of the 1” NPT turbines have
bore sizes that range from 3/8” up to 1”. Be sure to use
the correct bore size or the meter will report incorrect
flows and totals.
Note: Unless otherwise specified, Blancett turbine flow
meters are supplied with K-factors measured in pulses per
gallon (PUL/GAL) which will automatically convert to
your desired units of measure.
Note: The K-factor supplied with the meter or calculated
from calibration data will be needed to complete next step.