YSI 900 User manual

900 Chlorine Colorimeter
OPERATIONS MANUAL
INSTRUCTION MANUAL: PAGES 2 21
MODE D‘EMPLOI: PAGES 22 41
INSTRUCCIONES: PÁGINA 42 61

Contact Information
ORDERING & TECHNICAL SUPPORT
Telephone: (800) 897-4151
(937) 767-7241
Monday through Friday, 8:00 AM to 5:00 PM ET
Fax: (937) 767-9353 (orders)
(937) 767-1058 (technical support)
Email: [email protected]
Mail: YSI Incorporated
1725 Brannum Lane
Yellow Springs, OH 45387
USA
Internet: www.ysi.com
When placing an order please have the following information available:
SERVICE INFORMATION
YSI has authorized service centers throughout the United States and Internationally. For the nearest
Support directly at 800-897-4151.

3
Table of Contents
CAUTION
• General notes.............................................................4
.......................................4
Method notes..............................................................4
.....................................................5
• Functional description .....................................................6
Operation.................................................................6
...........................................................7
........................................................7
...............................................................7
• Methods .................................................................8
...................................................8
..............................................10
.............................................12
.............................................14
• Menu options ............................................................16
Menu selections...........................................................16
.......................................................16
Transmitting stored data ....................................................16
Setting date and time.......................................................17
• Calibration Mode .........................................................17
...........................................................17
....................................................19
• Technical data ...........................................................20
Operating messages .......................................................21
Error codes ..............................................................21
Battery disposal ...........................................................21

4
Guidelines for photometric measurements
vial aligned with the
shielding ring.
Method notes
General Notes

5
Replacement of batteries:
Correct position of the vial (Ø 24 mm):
CAUTION:
AAA/LR 03
AAA/LR 03
AAA/LR03
AAA/LR03
(E) seal ring
(A) screws
Correctllingofthevial:
correct
10 ml
5 ml
10 ml
5 ml
wrong
correct correct wrongwrong

6
Operation
Scroll Memory (SM)
-
The result appears in the display.
The result is saved automatically.
Repeating the test:
Repeating the zero:
Press the
METHOD
METHOD
METHOD
RESULT
On
Off
Mode
Zero
Test
Zero
Test
Zero
Test
0.0.0
Zero
Test
METHOD
Functional Description

7
Display backlight
-
Recall of stored data
4 seconds to access the recall menu.
Countdown / reaction period
-
surement starts immediately.
Caution:
An incomplete reaction period can lead to incorrect test
results.
!
!
!
Zero
Test

8
Methods
!
Zero
Test
Zero
Test
CL 6
Tablet Powder
CL 6
Tablet Powder
CL 6
Tablet Powder
0.0.0
Tablet Powder
RESULT
Tablet Powder
RESULT
Tablet Powder
Chlorine with Tablet: 0.01 – 6.0 mg/l
a) Free Chlorine
Fill a clean vial (24 mm Ø) with 10 ml of the water sample
empty it, leaving a few drops
remaining in the vial.
Add one DPD No. 1 tablet
dissolved.
b) Total Chlorine
Add one DPD No. 3 tablet
dissolved.
Wait for a reaction period of 2 minutes.
c) Combined Chlorine
Tolerances:
!
Zero
Test
Zero
Test
CL 6
Tablet Powder
CL 6
Tablet Powder
CL 6
Tablet Powder
0.0.0
Tablet Powder
RESULT
Tablet Powder
RESULT
Tablet Powder

9
Notes:
1. Vial cleaning:
Preparation:
then rinse all glassware thoroughly with deionized water.
3. Preparing the sample:
5. Exceeding the measuring range:
6. Turbidity (can lead to errors):

10
Methods
!
Zero
Test
Zero
Test
CL 10
Tablet Powder
CL 10
Tablet Powder
0.0.0
Tablet Powder
CL 10
Tablet Powder
RESULT
Tablet Powder
RESULT
Tablet Powder
Chlorine HR with DPD Tablet: 0.1 – 10 mg/l
a) Free Chlorine
Fill a clean vial (24 mm Ø) with 10 ml of the water sample
empty it, leaving a few drops
remaining in the vial.
Add one DPD No. 1 HR tablet
dissolved.
.
.
b) Total Chlorine
Add one DPD No. 3 HR tablet
.
dissolved.
Wait for a reaction period of 2 minutes.
.
c) Combined Chlorine
Tolerances:
Table of contents
Languages:
Other YSI Measuring Instrument manuals

YSI
YSI 550A User manual

YSI
YSI MultiLab 4010-3W User manual

YSI
YSI Pro 1020 Manual

YSI
YSI MultiDO 5400 User manual

YSI
YSI EcoSense ORP15A User manual

YSI
YSI 550 DO User manual

YSI
YSI Professional Plus User manual
YSI
YSI SonTek FlowTracker User manual

YSI
YSI YSI 556 MPS User manual

YSI
YSI EcoSense pH1000A User manual