Below are numbered steps to follow for operating the GV2510. Each step is described in
detail in section 5.2.
WARNING: The GV2510 is NOT certified for use in hazardous environments.
5.1
Operating steps
Please refer to Figure 2 for details.
1. The ReferenceMate is provided with a mounting head that accepts a 1/4-28 mounting
stud. Attach the unit to be tested (UUT- accelerometer, piezovelocity sensor, etc.) to the
ReferenceMate mounting platform, making certain to only hand-tighten the
accelerometer to no more than 10 in-lb (112 N-cm). If the accelerometer has something
other than a 1/4-28 mounting thread, use the appropriate adaptor outlined in Table 2.
The appropriate mounting torques for adaptors used with the GV2510 are also shown in
Table 2. An open end wrench is provided to stabilize the shaker head and make certain
the mounting platform (shaker head) is not excessively torqued. Over torqueing the
shaker head can result in permanent damage to the shaker suspension system. To
stabilize the shaker head during accelerometer attachment, slip the open end of the
wrench over the flats provided on the shaker head.
2. Once the UUT (and any mounting accessory) is mounted to the ReferenceMate head,
connect all cables from the UUT to the readout equipment. You are now ready to test
your device.
3. Press the ON/OFF button to initiate the 1 g test cycle. The OOR indicator (out of
range indicator) will come on momentarily as the control circuit stabilizes to a 1 g test
level. The shaker will shut off after approximately 90 seconds.
a.
Allow at least three seconds for the GV2510 to achieve the controlled vibration level.
Observe the BAT and OOR LEDs and make sure they are not illuminated. If the BAT
indicator remains on, or if the unit will not stay on after releasing the ON/OFF switch, replace
the batteries and try again. If the OOR LED illuminates, the maximum load has been
exceeded. The excessive weight must be removed, otherwise a 1 g test level will not be
achieved.
b.
To defeat the auto shutoff feature and enable continuous operation, press and hold the
FREQ button while turning the unit on. After the unit turns on, all three frequency LEDs will
briefly illuminate to indicate the shaker is in CONTINUOUS operating mode. The
ReferenceMate will now continue to operate until the ON/OFF button is depressed again.
This feature may be useful when the ReferenceMate is connected to an AC supply, such as
in a laboratory environment. See Section 6.1.
When the unit is turned off, it will revert to the AUTO-OFF mode.