Introduction
Congratulaons on purchasing a Turtle Tough 6000 Series Digital Smart Analyser. This analyser has been
specically developed to achieve opmum performance from your Turtle Tough Sensor.
This instrucon manual provides informaon for the correct installaon and use of a Turtle Tough 6000 Series
Digital Smart Analyser, to ensure you get the maximum life and performance from your sensor. Any use other
than the one described here compromises the safety of persons and the enre measuring system and is, there-
fore, not permied. The manufacturer is not liable for damage caused by improper or non-designated use.

TURTLE
TOUGH
®
3
Installation
As with all instrumentaon, the installaon and commissioning of this instrument are crucial to its
safe and eecve operaon. This instrument must only be used for its purpose as outlined in this
manual. It must be installed and commissioned by this manual and by trained, qualied personnel.
Site Selecon
Please choose a suitable locaon for the installaon of the electronics. The choice of installaon
point on any site is a compromise and is best undertaken by an experienced installaon engineer.
The following is a list of the factors that need to be taken into consideraon.
• Ensure that the mounng allows access to all serviceable parts.
• Try to mount the electronics in a posion where they are not habitually hosed down in a
cleaning process.
• The electronics enclosure should be mounted away from sources of heat or direct sunlight.
• Consider the length of wiring runs when mounng the instrument.
• Try to keep the electronics away from substaons, motors or other large EMI emiers.
• Consider whether the sample will be representave and well mixed.
• Consider sample line run mes.
• Consider sample return points.
• In a plasc run, with a low conducvity liquid sample, consider earthing the sample.
• If the instrument is controlling a dosing pump, size the pump appropriately.
Note: This list is not intended as a checklist neither is it implied that the list is complete.
Unpacking
• Please have a copy of your order with you when you unpack your instrument.
All orders are checked when they leave the factory.
• Please check that you have all the parts that were ordered as soon as you open the box.
• If anything is missing or damaged, please contact your sales outlet immediately.
• If the instrument needs to be returned for any reason please follow the return instrucons given
in this manual.
• Please dispose of the packing in an environmentally responsible manner and in compliance with
local regulaons.
Danger
Electric
Shock
Hazard
Important
Important

TURTLE
TOUGH
®
4
Specifications
Product Name 6000 Series Samrt Digital Analyser
Conducvity 0~500mS/cm
Resoluon 0.1us/cm;0.01ms/cm
Intrinsic Error ±0.5%F.S
Resisvity 0~18.25MΩ/cm
Resoluon 0.01KΩ/cm;0.01MΩ/cm
TDS 0~250g/L
Resoluon 0.01mg/L;0.01g/L
Salinity 0~700ppt
Resoluon 0.01ppm;0.01ppt
Temperature -10 to 150 °C
Resoluon ±0.3 °C
Temperature Compensaon Automac or manual
Power Supply 2 Rd 4~20mA
Oponal Power Supply 85~265VAC,9~36VDC, Power: ≤3W
Communicaon Output Modbus RTU RS485
Other Funcon Data recording, curve display, data uploading
Relay Control Contact 3 Groups: 5A 240VAC,5A 28VDC or 120VAC
Environmental Temperature -10~60 °C
Relave Humidity No more than 90%
Protecon Grade IP65
Weight 1.5kg
Dimensions 144*144*118 mm
Mounng Hole Dimensions 138*138mm
Installaon Embedded, wall - mounted, pipe type

TURTLE
TOUGH
®
5
Back fixed hole size Embedded mounting Cut-out size
138 mm +0.5 mm
138 mm +0.5 mm
M4x4
45x45 mm
144 mm
136 mm
144 mm
26 mm
118 mm
Installation
Dimensions

TURTLE
TOUGH
®
6
Installation completion diagram
Installation completion diagram
Install fixing bracket for instrument
Insert into the cut-out hole
a. Install fixing bracket for instrument
b. Wall screw fixing
D+ D-
B2 LN
RELAY A
RELAY B
RELAY C
Install fixed bracket top view,
pay attention to
6x15 mm
150.3 mm
58.1 mm
Installation
Embedded Installaon
Wall Mounted Installaon

TURTLE
TOUGH
®
7
Wiring Connection
V+,V-,A1,B1 Digital Input Channel 1 REF1 PH/Ion reference 1
V+,V-,A2,B2 Digital Input Channel 2 INPUT1 PH/Ion measure 1
I 1,G,I 2 Output current TEMP1 Temperature 1
A3,B3 RS485 communicaon output SEN-,SEN+ Membrane dissoluon of oxygen/FCL
G,TX,RX RS232 communicaon output
P+,P- VDC power supply REF2 PH/Ion reference2
INPUT 2 PH/Ion measure 2
T2+,T2- Temperature wiring TEMP2 Temperature 2
EC1,EC2,EC3,EC4 Conducvity/Resisvity wiring GND Grounding (for tesng)
RLY3, RLY2, RLY1 Group 3 relays CE,RE,WE Constant voltage for FCL/CLO2/O3
L,N, L re wire, N neutral wire, geodesic
Connecon Between Equipment And Sensor
The power supply of relay, output signal, alarm contact and connecon between sensor and instrument are all
inside the instrument. Wiring is done according to Figure 3. The lead length of the cable xed by the electrode
is generally 5-10 meters. Insert the line with corresponding label or color on the sensor into the corresponding
wiring terminal inside the instrument and ghten it.
5

TURTLE
TOUGH
®
8
Navigation
Keypad Operaon
Short Press: Release the key immediately aer pressing. ((Default to short presses if not indicated below)
Long Press: Press the buon for 3 seconds and then release it.
Press & Hold: Press the buon and accelerate aer a certain me unl the data is adjusted to the required
value before releasing the buon.
Enter menu setting mode
Return measurements
Two modes switching
Enter standard solution calibration mode
Calibrated: check the calibration status
Re-calibration: press “ENT” again
Return to
previous menu
Confirmation
options
Menu setting mode:
press this key to loop
down the menu options
Menu setting mode:
press this key to
rotate menu options
Press the key in
measurement mode to
display the trend chart

TURTLE
TOUGH
®
9
Display
Before using please check all pipes and electrical connecons. Aer power supply, the analyser is shown as:
Measurement Mode
Calibraon Mode
Seng Mode
Trend Chart Display

TURTLE
TOUGH
®
10
Menu Structure
Seng
Electrode Set
Electrode Type Seng
EC
TDS
Salinity
Resisvity
Unit Setup
us/cm, ms/cm
KΩ/cm, MΩ/cm
mg/L, g/L
ppm, ppt
%
Electrode Constant 1.0 (Default, can be modied)
Temperature Coecient 2.0 (Default, can be modied)
Temperature
Temperature Sensor
NTC2.252 kΩ
NTC10 kΩ
Pt100
Pt1000
Temperature Oset 0.0000
Temperature Input Automac
Manual
Temperature Unit °C
°F
Calibraon
Standard Soluon
Calibraon
Calibraon Number 1 0.01us/cm (Default, can be modied)
Calibraon Number 2 84us/cm (Default, can be modied)
Calibraon Number 3 1413us/cm (Default, can be modied)
Calibraon Number 4 12.88ms/cm (Default, can be modied)
Calibraon Number 5 111ms/cm (Default, can be modied)
Calibraon Correcon
Voltage 1
Voltage 2
Voltage 3
Voltage 4
Voltage 5
Field Calibraon
Field Calibraon
Oset Adjustment
Slope Adjustment
Alarm Set
Relay 1
On-O state ON
OFF
Specify the type
High Alarm
Low Alarm
Clean
Relay 2
On-O State ON
OFF
Specify the type
High Alarm
Low Alarm
Clean
Limit seng
(open me - cleaning state) Connuous opening me
Lag
(o me- in cleaning state)
(The interval between the last opening and
closing and the next opening)
Relay 3
On-O State ON
OFF
Specify the type
High Alarm
Low Alarm
Clean
Limit seng
(open me - cleaning state) Connuous opening me
Lag
(o me- in cleaning state)
(The interval between the last opening and
closing and the next opening)