Trotec T210 User manual

TRT-BA-T210-TC2014-04-002-EN
T210
EN
OPERATING MANUAL
THERMOHYGROMETER

2 EN
thermohygrometer T210
Table of contents
Notes regarding the operating manual.................................2
Safety .....................................................................................2
Information about the device................................................3
Transport and storage...........................................................5
Operation ...............................................................................5
PC software............................................................................8
Maintenance and repair ........................................................9
Errors and faults....................................................................9
Disposal .................................................................................9
Notes regarding the operating manual
Symbols
Warning of electrical voltage
This symbol indicates dangers to the life and health of
persons due to electrical voltage.
Warning
This signal word indicates a hazard with an average
risk level which, if not avoided, can result in serious
injury or death.
Caution
This signal word indicates a hazard with a low risk
level which, if not avoided, can result in minor or
moderate injury.
Note
This signal word indicates important information (e.g.
material damage), but does not indicate hazards.
Info
Information marked with this symbol helps you to carry
out your tasks quickly and safely.
Follow the manual
Information marked with this symbol indicates that the
operating manual must be observed.
You can download the current version of the operating manual
and the EU declaration of conformity via the following link:
T210
https://hub.trotec.com/?id=44142
Safety
Read this manual carefully before starting or using the
device. Always store the manual in the immediate vicinity
of the device or its site of use!
Warning
Read all safety warnings and all instructions.
Failure to follow the warnings and instructions may
result in electric shock, fire and/ or serious injury.
Save all warnings and instructions for future
reference.
This appliance can be used by children aged from
8years and above and persons with reduced physical,
sensory or mental capabilities or lack of experience
and knowledge if they have been given supervision or
instruction concerning use of the appliance in a safe
way and understand the hazards involved.
Children shall not play with the appliance. Cleaning and
user maintenance shall not be made by children
without supervision.
• Do not use the device in potentially explosive rooms.
• Do not use the device in aggressive atmosphere.
• Do not immerse the device in water. Do not allow liquids to
penetrate into the device.
• The device may only be used in dry surroundings and must
not be used in the rain or at a relative humidity exceeding
the operating conditions.
• Protect the device from permanent direct sunlight.
• Do not expose the device to strong vibrations.
• Do not remove any safety signs, stickers or labels from the
device. Keep all safety signs, stickers and labels in legible
condition.
• Do not open the device.
• Observe the storage and operating conditions as given in
the Technical data chapter.

EN 3
thermohygrometer T210
Intended use
Only use the device for indoor measurements of temperature
and humidity. Observe and comply with the technical data.
To use the device for its intended use, only use accessories and
spare parts which have been approved by Trotec.
Improper use
Do not use the device in potentially explosive atmospheres, or
for measurements in liquids.
Any unauthorised changes, modifications or alterations to the
device are forbidden.
Personnel qualifications
People who use this device must:
• have read and understood the operating manual, especially
the Safety chapter.
Residual risks
Warning of electrical voltage
There is a risk of a short-circuit due to liquids
penetrating the housing!
Do not immerse the device and the accessories in
water. Make sure that no water or other liquids can
enter the housing.
Warning of electrical voltage
Work on the electrical components must only be
carried out by an authorised specialist company!
Warning
Risk of suffocation!
Do not leave the packaging lying around. Children may
use it as a dangerous toy.
Warning
The device is not a toy and does not belong in the
hands of children.
Warning
Dangers can occur at the device when it is used by
untrained people in an unprofessional or improper way!
Observe the personnel qualifications!
Caution
Keep a sufficient distance from heat sources.
Note
To prevent damages to the device, do not expose it to
extreme temperatures, extreme humidity or moisture.
Note
Do not use abrasive cleaners or solvents to clean the
device.
Information about the device
Device description
The device enables a quick and precise determination of the air
temperature and relative humidity indoors. Both measured
values are indicated in real time on a pseudo colour display.
In addition to the relative humidity, further climate parameters
such as the dew point or absolute moisture content in the air
can be calculated with this device.
The temperature indication can be switched from °C to °F and
back. A minimum, maximum and average function is available
for the direct analysis of the measured data. Besides, the
currently measured value can be recorded via the hold function.
The sensor of this thermohygrometer located in the device's
measuring head is protected from dust and dirt by means of a
metal grid filter,
The device can be operated via a capacitive touchscreen control
panel.
When not in use, an automatic switch-off saves the battery.
Device depiction
1
4
3
2
5
6
No. Designation
1 Display
2 Cross control
3 Battery compartment with cover
4 Power button
5 USB interface
6 Measuring sensor

4 EN
thermohygrometer T210
Cross control
7
10
9
8
11
No. Designation
7 Up button
8 Right/back key
9 Down button
10 OK button
11 Left/menu key
Display
12
18
17
16
15
14
13
No. Designation
12 CFG symbol (configuration mode display)
13 Measuring mode
14 Upper measurement value display (temperature)
15 Temperature unit
16 Lower measured value display (humidity/dew point)
17 Unit for humidity or dew point
18 Configuration mode
Technical data
Parameter Value
Model T210
Temperature
Sensor type NTC
Measuring range -20°C to +50°C or -4°F to 122°F
Accuracy +/-0.4 °C or +/-0.7 °F
Resolution 0.1 °C or 0.1 °F
Relative humidity
Sensor type Capacitive
Measuring range 0.0 to 100.0 %RH
Accuracy ±2 %RH
Resolution 0.1 % RH
General technical data
Display LCD
Measuring frequency 2 x per second
Interface USB
Operating conditions 0°C to 50°C with <95 %RH
(non-condensing)
Storage conditions -30 to 60°C at <95%RH
(non-condensing)
Power supply 4x 1.5 V, AAbatteries
Weight approx. 270 g
Dimensions
(lengthx widthx height) 202mmx 63mmx 35mm
Scope of delivery
• 1x Device T210
• 4x 1.5V batteries, typeAA
• 1 x Specific test report
• 1 x Quick guide

EN 5
thermohygrometer T210
Transport and storage
Note
If you store or transport the device improperly, the
device may be damaged.
Note the information regarding transport and storage of
the device.
Transport
When transporting the device, ensure dry conditions and and
protect the device from external influences e.g. by using a
suitable bag.
Storage
When the device is not being used, observe the following
storage conditions:
• dry and protected from frost and heat
• protected from dust and direct sunlight
• The storage temperature is the same as the range given in
the Technical data chapter.
• Remove the batteries from the device.
Operation
Inserting the batteries
Insert the supplied batteries before first use.
Caution
Make sure that the surface of the device is dry and the
device is switched off.
1. Remove the battery compartment cover(3).
2. Insert the batteries with correct polarity.
3. Insert the battery compartment cover(3).
ðThe device can now be switched on.
Switch-on and measurements
Note:
Note that moving from a cold area to a warm area can lead to
condensation forming on the device's circuit board. This
physical and unavoidable effect can falsify the measurement. In
this case, the display shows either no measured values or they
are incorrect. Wait a few minutes until the device has become
adjusted to the changed conditions before carrying out a
measurement.
Please observe the information regarding the measurement
principle.
The cross control is very sensitive. Therefore, avoid dirt on the
control panel, because it could be misinterpreted by the device
as keystroke.
Before use make sure that the touchscreen control panel is dirt-
free.
If required clean the touchscreen control panel according to
chapter Cleaning the device.
1. Press the On/Off key(4) until a beep is emitted.
ðThe device performs a short self-test.
ðThe device name and firmware version is shown on the
display.
ðThe battery charge is indicated on the display.
ðThe device is ready for operation.
ðThe displayed units are based on the settings of the last
utilization.
2. Select the desired measuring mode.
3. Point the device at the area to be measured.
ðThe measured values will be displayed.
Key lock
1. Briefly press the On/Off key() during operation.
ðThe device emits a short beep.
ðLoC on is indicated on the display.
ðKey lock is activated.
2. Press the On/Off key() again.
ðThe device emits a short beep.
ðLoC oFF is indicated on the display.
ðKey lock is no longer activated.

6 EN
thermohygrometer T210
Configuration mode
1. Press the Left/menu key(11) for approx. 2seconds.
ðThe device emits a short beep.
ðThe CFG symbol(12) is displayed in the upper right
corner.
2. Use the buttons on the cross control(2) to select the
desired option.
12
18
By use of CAL a single-point calibration can be carried out for
the selected sensor indications. All sensors are already factory-
calibrated and have a corresponding characteristic calibration
curve. By stating a calibration value (offset) a global shift of the
calibration curve, which has an effect on the entire measuring
range, is performed for the single-point calibration! The offset
value to be entered is that value by which the calibration curve
will be shifted.
Example:
The displayed value is always "5" too high => change the offset
value for this measurement channel to "-5".
The offset value's default setting is 0.0.
Configurati
on mode
Description
Lamp Setting brightness.
Available are values between 20 and 100 and
AL.on.
Switch-off after 30 min unless with setting AL.on.
Pabs[hPa] Only available when the unit g/kg is active.
Specifying the absolute pressure (value range:
600 and 1200hPa).
1. Enter the value using the cross control(2).
2. Confirm with the OK button(10).
Alt.[ft.] Only available when the unit gr/lb is active.
Specifying the local altitude in feet (value range:
-999 und 9999 ft).
1. Enter the value using the cross control(2).
2. Confirm with the OK button(10).
CAL1 Specifying the temperature offset.
A negative offset allows no measured values of
less than 0.00.
1. Enter the value using the cross control(2).
2. Confirm with the OK button(12).
CAL2 Only available when %rH is active.
Specifying the offset for relative humidity (value
range: +/- 30 % rH).
A negative offset allows no measured values of
less than 0.00.
1. Enter the value using the cross control(2).
2. Confirm with the OK button(10).

EN 7
thermohygrometer T210
Temperature display configuration
The temperature can be displayed in either Celsius (°C) or
Fahrenheit (°F).
15
1. Press the Up key(7) to switch between °C and °F.
ðThe temperature will be displayed in the selected unit.
ðThe unit(15) will be indicated on the display(1).
Humidity display configuration
The humidity can be displayed in % RH, g/m3, g/kg and gr/lb,
and the dew point temperature in dp°C and dp°F.
17
1. Press the down key(9) until the desired unit is displayed.
ðThe humidity or dew point temperature will be displayed
in the selected unit in the lower measured value
display(16).
ðThe unit(17) will be indicated on the display(1).
Unit Description
g/m3Absolute humidity in g/m3
g/kg g/kg dry air
gr/lb gr/lb dry air
dp °C Dew point temperature in °C
dp °F Dew point temperature in °F
% RH Relative humidity in per cent
Measuring mode
1. Press the keys Right/back(8) or Left/menu(11) until the
desired measuring mode is displayed.
2. The selected measuring mode(13) will be indicated on the
display(1)
13
The device comes with the following measuring modes:
Measuring
mode
Description
ACT Measured value in real time
AVG Average value of measurements since switch-on
Min Minimum measured value
Max Maximum measured value
HOLD Measured value will be held
Holding the measured value
1. Set the measuring mode to HOLD.
ðThe current measured value will be held and displayed.
ðThe device will hold this value until the measured values
are set back or the device is switched off.
Resetting the measured values
1. Press the OK button(10) for approx. 2 seconds.
ðAll previously stored measured values of the measuring
modes AVG, MIN, MAX and HOLD will be set back.
ðAll previously stored measured values of the measuring
modes AVG, MIN, MAX and HOLD will be set back.
Measured value storage
Please note that measured values cannot be saved on the
device itself. In order to save measured values, the device has
to be connected to a PC via a USB cable using the MultiMeasure
Studio software.
1. Briefly press the OK button(10).
ðThe displayed measured value will be saved in the
software.
Further information can be gathered from the help text of the
MultiMeasure Studio software.
USB interface
The device can be connected to a PC via the USB interface(5).
See chapter PC software.
Switch-off
1. Press and hold the On/Off key(4) for approx. 3 seconds
until a beep is emitted.
ðThe device is switched off.

8 EN
thermohygrometer T210
PC software
Use the MultiMeasure Studio Standard PC software (free
standard version) or MultiMeasure Studio Professional (paid
professional version, dongle required) to carry out a detailed
analysis and visualisation of your measured results. You can
only use all configuration, visualisation and functional options of
the device when using this PC software and a TROTEC® USB
dongle (professional).
Installation requirements
Ensure that the following minimum requirements for installing
the MultiMeasure Studio Standard or MultiMeasure Studio
Professional PC software are fulfilled:
• Supported operating systems (32 or 64bit version):
– Windows XP from service pack 3
– Windows Vista
– Windows 7
– Windows 8
– Windows 10
• Software requirements:
– Microsoft Excel (to display stored Excel files)
– Microsoft .NET Framework 3.5SP1 (is automatically
installed during software installation, where applicable)
• Hardware requirements:
– Processor speed: 1.0GHz, minimum
– USB connection
– Internet connection
– 512MB RAM, minimum
1GB hard disk space, minimum
optional: TROTEC® USB dongle (Professional) for using the
professional version of the PC software
Installing the PC software
1. Download the current PC software from the Internet. To do
so, visit the website www.trotec.de. First click on Service,
followed by Downloads and then Software. Select the
MultiMeasure Studio Standard software from the list. If you
want to use the optionally available professional version of
the PC software "MultiMeasure Studio
Professional" (dongle), then get in touch with your
TROTEC® customer service.
2. Double-click on the downloaded file to start the
installation.
3. Follow the instructions of the installation wizard.
Starting the PC software
1. Connect the device to your PC via the USB connection
cable provided in the scope of delivery.
Note
Step 2 only needs to be performed, when using the
Professional software functions.
If you only use the Standard software functions, please
proceed to step 3.
2. In order to enable the Professional functions, connect the
TROTEC® USB dongle to a free USB port on your PC.
The TROTEC® USB dongle (Professional) is automatically
detected by the operating system.
If you only connect the TROTEC® USB dongle (Professional)
to your PC after starting the PC software, click the
"Parameters" menu item in the PC software. Afterwards,
click the USB symbol (dongle check) to read the connected
TROTEC® USB dongle (Professional).
3. Switch the device on (see chapter Switch-on and
measurements).
4. Start the MultiMeasure Studio software. Depending on the
activation process you will be asked to insert the access
code that has been previously assigned to you. Only then
the dongle for releasing the according Professional tools of
the software will be activated.
Note
Information regarding the use of the MultiMeasure
Studio software is provided in the help text of the
software.

EN 9
thermohygrometer T210
Maintenance and repair
Battery change
Change the batteries when the message Batt lo is displayed
upon switch-on or the device can no longer be switched on.
See chapter Operation Inserting the batteries.
Cleaning
Clean the device with a soft, damp and lint-free cloth. Make
sure that no moisture enters the housing. Do not use any
sprays, solvents, alcohol-based cleaning agents or abrasive
cleaners, but only clean water to moisten the cloth.
Repair
Do not modify the device or install any spare parts. For repairs
or device testing, contact the manufacturer.
Errors and faults
The device has been checked for proper functioning several
times during production. If malfunctions occur nonetheless,
check the device according to the following list.
The device does not switch on:
• Check the charging status of the batteries. Change the
batteries when the message Batt lo is displayed upon
switch-on.
• Check that the batteries are properly positioned. Check the
polarity is correct.
• Never carry out an electrical check yourself; instead,
contact your TROTEC® customer service.
Disposal
The icon with the crossed-out waste bin on waste
electrical or electronic equipment stipulates that this equipment
must not be disposed of with the household waste at the end of
its life. You will find collection points for free return of waste
electrical and electronic equipment in your vicinity. The
addresses can be obtained from your municipality or local
administration. For further return options provided by us please
refer to our website https://de.trotec.com/shop/.
The separate collection of waste electrical and electronic
equipment aims to enable the re-use, recycling and other forms
of recovery of waste equipment as well as to prevent negative
effects for the environment and human health caused by the
disposal of hazardous substances potentially contained in the
equipment.
In the European Union, batteries and accumulators must not be
treated as domestic waste, but must be disposed of
professionally in accordance with Directive 2006/66/EC of the
European Parliament and of the Council of 6September 2006
on batteries and accumulators. Please dispose of batteries and
accumulators according to the relevant legal requirements.

Trotec GmbH
Grebbener Str. 7
D-52525 Heinsberg
+49 2452 962-400
+49 2452 962-200
www.trotec.com
Table of contents
Other Trotec Measuring Instrument manuals

Trotec
Trotec BE40 User manual

Trotec
Trotec BE44 User manual

Trotec
Trotec BE40 User manual

Trotec
Trotec SL800-SET User manual

Trotec
Trotec SL3000 User manual

Trotec
Trotec BP17 User manual

Trotec
Trotec PDR 908 HP User manual

Trotec
Trotec BM12 User manual

Trotec
Trotec VSC206 User manual

Trotec
Trotec T3000 User manual