
Trace2o UV Digester HMUV1825P
Instruction Manual v3.0
5. PROGRAMMING THE DIGESTION TIME
The digester is fitted with a programmable countdown timer. A digestion time
up to 10 hours can be programmed although 30 mins to 1 hour is sufficient
for most samples.
To set the timer press the ‘programme’ button. The seconds will flash. To
adjust the time press the + and - buttons to adjust the time up or down
respectively. Subsequent presses of the ‘programme’ button will step through
from seconds to minutes to hours. Pressing the ‘programme’ button again will
set the time. (If the digits are flashing the timer will not operate).
Once the time is set press the ‘start’ button. The pump will turn on and the
lamp will illuminate. This can be confirmed by looking through the viewing
window. The timer will start counting down and an alarm will sound when the
cycle has completed. If the logo does not illuminate the bulb will need
replacing. (See maintenance section).
After the sample is complete, press reset to set the time back to the pre-
programmed time.
Lift the tubes clear of the sample water and start the timer again. Run until
the tubes are empty of liquid.
Note : Running the pump dry will not harm the system.