10 PTV_Series_V2.2 01.2019
GB
Specifications
Technical Specifications
The Lovibond®PTV Series Turbidimeter is available in four versions, one ISO and three
US EPA compliant versions. The PTV 1000 IR displays turbidity values in FNU units and
is compliant to ISO method 7027. The PTV 1000 WL, PTV 2000 and PTV 6000 are EPA
approved methods and display turbidity values in NTU units. Approval information for the
PTV 1000 WL, PTV 2000 and PTV 6000 are referenced in Federal Register / Vol. 82, No 143
/ Thursday, July 27, 2017 / Rules and Regulations, 34861 – 34868.
Specification Details
Measurement Method Nephelometry, scattered light collected at 90° to the incident
beam
Operating temperature 0 to 50°C (32 to 122°F)
Measurement Range PTV 1000 / PTV 2000: 0.0001 to 100 NTU / FNU
PTV 6000: 0.0001 to 20.00
Display units FNU, NTU, mNTU, TE/F, mg/l PSL, mg/l Kaolin, Degree, custom
Accuracy ± 2% of reading from 0 to 10 NTU
± 4% of reading between 10 to 100 NTU
Repeatability Less than 1% at 1 NTU/FNU, expressed as percent relative
standard deviation (%RSD)
Linearity error Less than 1% for 0 to 5 NTU and 2% for turbidity values
greater than 10 NTU (requires 2-point calibration)
Stray Light PTV 1000 IR: <0.005 / 5 mNTU
PTV 1000 WL: <0.015 / 15 mNTU
PTV 2000: <0.008 / 8 mNTU
PTV 6000: <0.005 / 5 mNTU
Limit of Detection PTV 1000: <0.0005 NTU
PTV 2000: <0.0001 NTU
PTV 6000: <0.0001 NTU
Limit of Quantitation PTV 1000: Better than 0.005 NTU
PTV 2000: Better than 0.001 NTU
PTV 6000: Better than 0.001 NTU
Resolution 0.0001 NTU or FNU (range dependent 5 digits displayed)
Response time 10% change within 40 seconds @ 200 mL/min
Response 10 to 90% Less than 240 seconds @ 200 mL/min at 1 NTU
Signal Averaging User Selectable: 1, 3, 6, 10, 30, 60, and 90 seconds.
Default set to 30 seconds
Sample temperature10 to 50°C (32 to 122°F) 1For best performance the sample temperature
should be at least 5°C less than the ambient temperature.
Sample Flow 30 to 500 2mL/min (0.476 to 7.925 gal/hr),
40 to 120 mL/min (0.634 to 1.902 gal/hr) recommended flow;
2For flow rates above 100 ml/min (1.585 gal/hr) use external flow monitoring (not
supplied).
Sample pressure 0.03 to 5.5 bar (0.435 to 80 psi)
Sample Volume 285-ml (Measurement Module installed in Flow Body)