3
4. SPECIFICATIONS
Applicable standards: IEC61672-1: 2003 Class 2
IEC60651: 1979 Type 2
ANSI S1.4: 1983 Type 2
Measurement functions:
•Main processing functions
Sound level: Current time-weighted sound pressure level A or current time-
weighted sound pressure level C
Maximum time-weighted sound pressure level A or Maximum time-weighted
sound pressure level C
•Total range: 30 to 130dB
•Max. measurement level: 130dB
•Total linear operating range: In accordance with IEC 61672-1, A-weighted,
1000Hz: 30dB to 130dB.
•Level range selection: 5 ranges in 10dB steps 30 to 90dB, 40 to 100dB,
50 to 110dB, 60 to 120dB, 70 to 130dB
Frequency range: Overall characteristics including microphone: 31.5 to 8000Hz
Frequency weighting: A,meets the requirement of IEC 61672-1 for class 2
“A” weighting.
C,meets the requirement of IEC 61672-1 for class 2
“C” weighting.
Time weighting (RMS detection): Fast, according to IEC 61672-1 class 2.
Slow, according to IEC 61672-1 class 2.
•Reference conditions:
Type of the acoustic field: Free
Reference sound pressure level: 94.0dB (related to 20µPa)
Reference level range: 60 to 120dB
Reference frequency: 1000Hz
Reference temperature: +23℃
Reference relative humidity: 50%RH
Reference static pressure: 101.325 kPa
Reference incidence direction: Perpendicular to the front of the microphone
diaphragm.
•Calibration: Acoustic using sound calibrator.
Calibration check frequency is 1000Hz.
Nominal calibration level for the free field: 94.1dB
Nominal calibration level for the diffuse field: 94.0dB