Tack life CM02A User manual

DIGITAL AC CLAMP METER
User Manual

Contents
English 01 - 11
Italiano 45 - 55
Deutsch 12 - 22
Français 23 - 33
Español 34 - 44

1. Safety information
Warning
•Please operate this instrument with great care. Improper operation may result in an electric shot
or damage to the instrument. Throughout the operation, you should follow the generally accepted
safety procedures and take the safety measures as required by the Operation.
•Please read carefully this Manual and take the operational methods as specified herein so as to
make full use of the instrument's functionalities and ensure safe operation.
•This instrument is in strict compliance with the safetyrequirements as specified inIEC-61010-1,
IEC-61010-2-030 and IEC-61010-2-032 for electrical measuring instruments. Its pollution reachesthe
level of Class II and over-voltage standard is CAT III 600V.
•Please strictly follow the guideline for safe operation so as to ensure safety while operating this
instrument.
1.1 Preparation
1.1.1 The user must observe the standard safety rules when operating this instrument:
•General protection against electrical shock
•Prevention of unintended use
1.1.2 Upon the arrival of the instrument, check any damage that arises during transportation.
1.1.3 Upon the arrival of the instrument that has been stored and shipped in rough conditions, check
and identify any damage.
1.1.4 The instrument must be kept in agoodcondition. Prior to its use, check the possible damage to
insulation part and potential exposed metal wire of the lead.
1.2 Symbols
CAT III
Note (For important safety information, see Operation Manual)
This symbol indicates that it can be used on a hazardous live conductor.
Double insulation protection(Category II)
refers to over-voltage level III (installation) as specified in IEC-61010-1standard.
Pollution level 2 refers to the level of impulse withstand voltage protection.
It is compliant with appropriate EU standard.
Grounding
- 01 -

1.3 Maintenance
•Do not attempt to open the bottom case to adjust or repair instruments. Such operation can only
be executed by an electrician who is fully familiar with the instrument and electric shock risks.
•Remove pen-shaped meter from the line to be measured, before opening the instrument's
bottom case or battery cover
•To avoid an electric shock that results from any false readings, replace existing battery when the
symbol is displayed.
•Don’t use any abrasive agentsor solvents when a wet cloth and mild detergent are being employed
to clean the instrument.
•Power off and keep the range switch to the position "OFF" when the instrument is not in use.
•Remove batteryto avoid any damage to the instrument when the instrument is not in use for a
long period.
2. Description
2.1 Components
1. Non-contact voltage detection
2. Data hold/Backlight key
3. Rotary switch
4. Display screen
5. Input socket
6. Function Selecting/non-contact voltage detection key
7. Trigger
8. Current clamp head: for current measurement
- 02 -
①
②
③
④
⑤
⑥
⑦
⑧

Shutdown position
AC current measurement
AC/DC voltage measurement
Low resistance voltage measurement
Electric resistance measurement
ON/OFF measurement
Capacitance measurement
Temperature measurement
OFF
2.2 The description of rotary switch
AC &DC
On/Off indication
Automatic range measurement mode
Auto shutdown indication
Low battery
Hold status
Volt(Voltage);Ampere(Current)
Ohm, kilohm and megohm (Resistance)
Hz, kHz
Non-contact voltage detection
uF
℃℉
Low resistance measurement mode
V;A
Ω; kΩ; MΩ
Hz, kHz,
NCV
uF
℃℉
LowZ
- 03 -

3. Specification
The instrument shall be re-calibrated at an interval of one year under the conditions of 18 ℃-28 ℃
and relative humidity of less than 75%.
3.1 Overview
• The instrument is designedto automatically select measurement functions and measuring ranges.
• Full range overload protection.
• Allowable max voltage between terminal to be measured and ground:600V DC or 600V AC
• Working weight: max2000m
• Display unit: LCD
• Max display value:6000 numbers
• Polarity indication: automatic indication.‘-’indicates negative polarity.
• Over range indication:‘0L’ or ‘-0L’
• Sampling rate: approximately 3 times per second
• Unit display: to display functions and electric quantity
• Automatic shut-down time:10 minutes
• Power supply: 1.5V AAA battery ×2
• Battery under-voltage indication: LCD display the symbol .
• Temperature coefficient: less than0.1x accuracy/℃
• Working temperature:18℃~28℃
• Storage temperature:-10℃~50℃
3.2 Technical parameters
3.2.1 AC current
- Minimum input current: 0.01A AC current
- Max input current: 600A AC current
- Frequency range:45~65Hz ;
- At the moment of AC current measurement, the instrument will automatically enable the internal
low pass filter and filter out high frequency current,
Measuring range
6A
60A
400A
400A~600A
Resolution
0.001A
0.01A
0.1A
0.1A
Accuracy
±(2.5% Reading + 8digits)
±(3% reading + 10 digits)
- 04 -

3.2.2 DC voltage
- Minimal input current 0.001V DC
- Maximal input current:600V DC
3.2.3 AC voltage
- Maximal input current:600V AC (valid value)
- Frequency range:45 ~ 65Hz
3.2.4 Frequency
3.2.4.1 Frequency measurement for the current through clamp head:
- Measurement range:40Hz ~ 1000Hz
- Input signal range:≥ 0.2A AC current(valid value)
3.2.4.2 Frequency measurement for voltage through Input Jack:
- Measurement range:40Hz ~ 1000Hz
- Input signal range:≥ 0.8V AC voltage (valid value)
Measuring range
6V
60V
600V
Resolution
0.001V
0.01V
0.1V
Accuracy
±(0.5% Reading + +3 digits)
Measuring range
6V
60V
600V
Resolution
0.001V
0.01V
0.1V
Accuracy
±(0.8% Reading + +5 digits)
Measuring range
60.0Hz
1000Hz
Resolution
0.1Hz
1Hz
Accuracy
±(1.0% reading + 5digits)
Measuring range
60.0Hz
1000Hz
Resolution
0.1Hz
1Hz
Accuracy
±(1.0% reading + 5digits)
- 05 -

3.2.5 Electric resistance
- Overload protection:250V DC or AC(valid value)
3.2.6 Line On/Off test
- Overload protection:250V DC or AC(valid value)
3.2.7 Capacitance
- Overload protection:250V DC or AC(valid value)
3.2.8 Temperature
- Overload protection:250V DC or AC(valid value)
Measuring range
6k
60k
600k
6M
60M
Resolution
0.001k
0.01k
0.1k
0.001M
0.01M
Accuracy
± 0.8% Reading + +3 digits
Measuring range Resolution
1Ω
Functions
If the electric resistance of the line measured is less
than 50, the buzzer inside the instrument may sound.
Measuring range
600uF
6000uF
Resolution
0.1uF
1uF
Accuracy
± 4% Reading + +3 digits
Measuring range
-20~1000℃(-4~1832℉)
Resolution
1℃/2℉
Accuracy
± 1% Reading + +2 digits
- 06 -

4. Operational guidelines
4.1 Reading hold
During the process of measurement, gently touch the key if you want to hold readings, and
monitor's display value will be locked. Touch again the key , the readings hold will be removed.
4.2 Backlight
1)In the process of measurement, if the measurement environment is too dark, press the key
for more than 2 seconds to enable backlight function. Then about 1 minute later, the backlight
function will be automatically disabled.
2)During this process, press the key for 2 seconds to disable backlight.
4.3 Auto shutdown
1)If no operations occur within 10 minutes after the initialization, the instrument will be in the state
of dormancy. Auto shutdown at this moment can save power consumption. 2 minutes before
shutdown, the buzzer will sound at an interval of 1 minute.
2)Press any key after Auto shutdown to wake the instrument into operation.
3)The function of Auto shutdown will be disabled if SEL/NCV key is pressed while the instrument is
initialized.
4.4 AC current measurement
1)Turn the rotary switch to switch to the AC measuring range, hold the trigger, open clamp head,
put a lead wire of the line to be measured into the clamp. Then the primary display panel will display
the measured current value while the secondary display panel will show the measured frequency
value.
- 07 -

4.5 DC voltage measurement
Turn the selector switch to AC/DC voltage position, press SEL key to switch to DC voltage
measurement mode, and connect the pen-shaped meter to the signal to be measured. The red
meter is connected to the positive pole of the signal to be measured and the black meter to its
negative pole.
4.6 AC voltage measurement
1) Turn the selector switch to AC/DC voltage position and press SEL key to switch to AC voltage
measurement mode, connect the pen-shaped meter to the signal to be measured. Then the
primary display panel shows the measured voltage value and the secondary display panel the
measured frequency value of voltage.
- 08 -
Table of contents
Languages:
Other Tack life Measuring Instrument manuals