SUTO S551 Manual

English
Instruction and operation manual
S 551
Compressed air analyser

Dear Customer,
thank you for choosing our product.
The operating instructions must be read in full and carefully observed
before starting up the device. The manufacturer cannot be held liable
for any damage which occurs as a result of non-observance or non-
compliance with this manual.
Should the device be tampered with in any manner other than a
procedure which is described and specified in the manual, the warranty
is cancelled and the manufacturer is e empt from liability.
The device is destined e clusively for the described application.
SUTO offers no guarantee for the suitability for any other purpose.
SUTO is also not liable for consequential damage resulting from the
delivery, capability or use of this device.
2 S 551

Table of contents
1. Safety instructions.......................................................................5
2. Application.................................................................................7
3. Features.....................................................................................7
4. Technical Data............................................................................8
4.1 General.................................................................................8
4.2 Electrical Data........................................................................8
4.3 Input-Signals.........................................................................8
4.4 Output-Signals.......................................................................8
5. Installation ................................................................................9
5.1 Installation Requirements........................................................9
5.2 Installation Procedure ............................................................9
5.3 Electrical connection ............................................................10
6. Configuration ...........................................................................13
7. Operation ................................................................................13
7.1 Value screen........................................................................14
7.2 The main menu ...................................................................15
7.3 Description of display icons in status bar .................................16
7.4 Graphic screen.....................................................................16
7.5 Sensor setting .....................................................................17
7.5.1 Dew point setting ...........................................................17
7.5.2 Flow sensor setting ........................................................19
7.5.3 Oil vapor sensor setting ..................................................21
7.5.4 Analog input setting .......................................................22
7.5.5 Power meter S 110-P setting............................................23
7.5.6 Analog e tension module.................................................23
7.5.7 Laser particle counter S 130.............................................23
7.5.8 Ultrasound liquid flow meter S 460....................................23
7.6 Logger ................................................................................24
7.7 Files....................................................................................25
7.8 Service info..........................................................................25
7.9 System setting.....................................................................26
7.10 Communication ..................................................................26
8. Signal inputs ............................................................................28
8.1 Digital inputs........................................................................28
8.2 Analog input ........................................................................28
9. Signal outputs...........................................................................28
9.1 Interface ............................................................................28
10. Optional e tra accessories.........................................................28
11. Calibration..............................................................................29
12. Maintenance............................................................................29
13. Disposal or waste.....................................................................29
S 551 3

1. Safety instructions
1. Safety instructions
Please check if this instruction manual accords to the
product type.
Please observe all notes and instructions indicated in this
manual. It contains essential information which have to be
observed before and during installation, operation and
maintenance. Therefore this instruction manual has to be read carefully
by the technician as well as by the responsible user / qualified
personnel.
This instruction manual has to be available at the operation site of the
compressed air analyser at any time. In case of any obscurities or
questions, regarding this manual or the product, please contact the
manufacturer.
WAR I G!
Compressed air!
Any contact with quickly escaping air or bursting
parts of the compressed air system can lead to
serious injuries or even death!
•Do not e ceed the ma imum permitted pressure range (see
sensors label).
•Only use pressure tight installation material.
•Avoid that persons get hit escaping air or bursting parts of the
instrument.
•The system must be pressureless during maintenance work.
WAR I G!
Voltage used for supply!
Any contact with energized parts of the product, may
lead to a electrical shock which can lead to serious
injuries or even death!
•Consider all regulations for electrical installations.
•The system must be disconnected from any power supply during
maintenance work.
•Any electrical work on the system is only allowed by authorized
qualified personal.
S 551 5

1. Safety instructions
WAR I G!
Permitted operating parameters!
Observe the permitted operating parameters, any
operation exceeding this parameters can lead to
malfunctions and may lead to damage on the
instrument or the system.
•Do not e ceed the permitted operating parameters.
•Make sure the product is operated in its permitted limitations.
•Do not e ceed or undercut the permitted storage and operation
temperature and pressure.
•The product should be maintained and calibrated frequently, at
least annually.
General safety instructions
•It is not allowed to use the product in e plosive areas.
•Please observe the national regulations before/during installation
and operation.
Remarks
•It is not allowed to disassemble the product.
ATTE TIO !
Measurement values can be affected by malfunction!
The product must be installed properly and
frequently maintained, otherwise it may lead to
wrong measurement values, which can lead to
wrong results.
Storage and transportation
•Make sure that the transportation temperature of the device is
between -20°C... 50°C.
•For transportation it is recommended to use the packaging which
comes with the device.
•Please make sure that the storage temperature of the device is
between -20°C... 50°C.
6 S 551

1. Safety instructions
•Avoid direct UV and solar radiation during storage.
•For the storage the humidity has to be <90%, no condensation.
2. Application
The S 551 is a compressed air analyser which is designed to display and
record all relevant parameters of compressed air and gases within the
permissible operating parameters. These parameters can be found in
the technical data section.
The S 551 can analyse for e ample the flow, dew point, pressure,
temperature, power consumption and many more.
The S 551compressed air analyser is not developed to be used in
e plosive areas. For the use in e plosive areas please contact the
manufacturer.
The S 551 compressed air analyser is mainly used in compressed air
systems in industrial environment.
3. Features
•Connectable sensors for all required measurement tasks (air flow,
air consumption, power consumption, pressure, temperature and
many more).
•Up to 24 inputs through e tension bo es and modbus.
•Several loggers can be combined: no need to have long cables
from the sensor to the logger.
•Third party sensors can be easily connected.
•IP65 casing provides robust protection in the industrial
environment.
•High resolution 5” colour touch screen interface.
•Because of the battery backup power, power glitches and cuts
wont effect the performance.
•Full software package is included:
◦S4M-S for basic analyses
◦CAA for compressed air audit analysis
S 551 7

4. Technical Data
4. Technical Data
4.1 General
Data logger 4 GB, up to 100 million values
Operating temperature 0°C... 50°C
Housing material PC + ABS
Protection class IP 65
Dimensions 365 mm 270 mm 169 mm
Display 5” high resolution graphic display, 800 480
pi els with touch interface
Weight 4 kg
4.2 Electrical Data
Power supply 100... 240 VAC / 50 VA, 47-60 Hz
Internal battery Rechargeable for up to 8 hours operation.
Charging time ca. 3 hours
4.3 Input-Signals
Analog input 0... 1 V, 0... 10 V, 0...20 mA, 4... 20 mA
Digital input 2 SDI Sensors
16 RS 485 Modbus RTU Sensors
4.4 Output-Signals
Communication Interface Ethernet, USB
8 S 551

5. Installation
5. Installation
Please make sure that all components listed below are included in your
package.
Qty Description Item No.
1 S 551-P4 / S 551-P6 P560 5100/5101
1 USB cable A553 0130
1 Instruction manual No P/N
1 Calibration certificate
5.1 Installation Requirements
ATTE TIO !
Wrong measurement is possible, if the device is not
installed correctly.
•The device is for indoor use only! At an outdoor installation, the
device must be protected from solar radiation and rain.
•It is strongly recommend not to install S 551 permanently in wet
environment as it e ists usually right after a compressor outlet.
5.2 Installation Procedure
Because of the casing the S 551 no need a costly installation. Just
connect the sensors like described in the chapter electrical connection.
S 551 9

5. Installation
5.3 Electrical connection
Please connect all sensors before switch on the S 551. Flow / dew point
sensors are detected automatically when connected. The same
e tension modules ( power meter, US flow meter, analog e tension)
which are connected to terminal M.
Process signals can not be detected automatically. The user has to
select the appropriate sensor type from the sensor selection menu. For
details please refer to chapter 7.5.4 analog input setting.
Connection on the casing
Channel Sensor type / Description
1 + 2 Flow and dew point sensors
3 + 4 Process signals
M +M Modbus sensor modules
5 Ethernet port
6 Mains supply
10 S 551