3 pHElectrode use and maintenance
3 Use and maintenance of pH electrode
(1) There is an appropriate amount in the protective cover of the front section of the electrode3.3mol/LKCLsolution, in
which the electrode tip is immersed to maintain activation of the sensitive membrane and liquid junction.
(2) When using the electrode, the front transparent protective cover needs to be removed, and the glass bulb and the
liquid junction should be immersed in the solution for use.
(3) before installation, be sure to use a raw tape (3/4Threads) do a good job of
waterproof sealing to prevent water from enteringpHelectrode, causingpHElectrode
cable is shorted.
(4) measurement, should be washed in distilled water (or deionized water) first, and use filter paper to absorb moisture
to prevent impurities from entering the measured liquid, and the electrode sensitive membrane and liquid junction should be
completely immersed in the measured liquid.
(5) Check whether the terminal is dry. If it is stained, please wipe it with
anhydrous alcohol and use it after drying.
(6) It is recommended that the user regularly clean the sensitive membrane and liquid junction in front of the
electrode, and regularly calibrate with the instrument.
(7) The electrode should be washed when not in use, and inserted into the protective cover with saturated potassium
chloride solution.