The item numbers refer to the spare parts list RN 01.256.
11.1 Dismantling from the line system
a. Shut off line pressure in the product line and discharge line if
possible.
b. Disconnect electric and pneumatic connecting lines.
c. Remove hex. screws (3). Screw two screws into the threaded holes
of the housing cover (2), thus, pressing the measuring device and
the housing cover off.
11.2 Dismantling of wear parts
- If the measuring device has additional wear parts, the
corresponding operating instruction must be observed.
a. Pull off housing seal (4).
11.3 Installation of seals and measuring devices
a. Pull the slightly greased seal on the flange of the measuring
device or of the housing cover (2).
See to a correct fit of the seal.
b. Place the measuring device or housing cover straightly on the
housing flange and tighten it by the hex. screws (3).
c. Connect electric and pneumatic lines.
12. Trouble Shooting
- leakage between housing flange
and flange of measuring device or
housing cover : replace housing seal.
13. Spare Parts Lists
(see annex)
11. Assembly Instructions
7Inline Measuring Techniques
Operating Manual
Rev.0
UK