• WK-21 ½” measurement microphone (manufacturer: SONOPAN).
Other accessories specified in chapter 1.1 are optional.
1.4. Measured quantities
Many quantities used to characterize acoustic noise can be measured simultaneously using
DSA-50 digital sound analyzer. The set of measured quantities depends on instrument operating
mode. Measurement duration and battery condition information is displayed at every operating
mode.
1.4.1. Sound level meter (SLM) mode
For two frequency weightings, selected among from three ones: A, C and Z available in the
instrument, following quantities can be measured simultaneously (in symbols listed below X stands
for selected frequency weighting):
• LXF current RMS, F-time-weighted sound level,
• LXFmn minimum RMS, F-time-weighted sound level,
• LXFmx maximum RMS, F-time-weighted sound level,
• LXS current RMS, S-time-weighted sound level,
• LXSmn minimum RMS, S-time-weighted sound level,
• LXSmx maximum RMS, S-time-weighted sound level,
• LXI current RMS, I-time-weighted sound level,
• LXImn minimum RMS, I-time-weighted sound level,
• LXImx maximum RMS, I-time-weighted sound level,
• LXPk current peak sound level,
• LXMPk maximum peak sound level,
• LXeq time-average (equivalent continuous) sound level,
• LXeq1s time-average sound level for last one second.
If A frequency weighting is one of two selected weightings, then following quantities are
measured additionally for this weighting:
• LAE sound exposure level,
• LAEX,T noise exposure level referred to specified exposure time T.
1.4.2. 1/1-octave analyzer mode
1/1-octave analyzer contains nine digital octave-band filters having exact centre frequencies
specified for binary octave ratio. Nominal centre frequencies are 31.5Hz, 63Hz, 125Hz, 250Hz,
500Hz, 1kHz, 2kHz, 4kHz and 8kHz. Specifications of filters conform to Class 1 requirements of
EN 61260:1995 and EN 61260:1995/A1:2001.
Following quantities can be measured simultaneously for each filter centre frequency (and also
for wide-band response), when one frequency weighting is selected among from three available
weightings (A, C and Z):
• LFcurrent RMS, F-time-weighted sound level,
• LS current RMS, S-time-weighted sound level,
• Leq time-average sound level.
If A frequency weighting is selected, then following quantities are measured additionally:
• LASmx maximum RMS, A-frequency-weighted, S-time-weighted sound level,
• LAeq time-average A-frequency-weighted sound level,
• LCMPk maximum peak C sound level.
1.4.3. 1/3-octave analyzer mode
1/3-octave analyzer contains twenty nine digital 1/3-octave band filters having exact centre
frequencies specified for binary octave ratio. Nominal centre frequencies are 25Hz, 31.5Hz, 40Hz,
50Hz, 63Hz, 80Hz, 100Hz, 125Hz, 160Hz, 200Hz, 250Hz, 315Hz, 400Hz, 500Hz, 630Hz, 800Hz,