Operating Instructions:
1. Measuring %mc
a) Inspect the cell to ensure it is clean and empty
b) Turn on the analyzer by pressing ON/TEST button. The >> MEASURE << menu will be
automatically displayed.
c) Accept mean by pressing the ON/TEST button.
d) By using the or arrows select the crop or commodity for measurement.
e) Measure the volume of crop/commodity as indicated by the displayed calibration.
The volume will be either 9 ml or 11 ml. dependant on the crop or commodity. For
convenience these are provided in the cap of the grinder/compressor. Fill the appropriate
cup to be level with the surface of the cap. Pour the measured volume into the measuring
cell and screw down the cap clockwise until the stainless steel pin in the cap and
measuring cell prevent further rotation. see figures 1a - 1 d.
f) Press the ON/TEST button to start the test. The progress of the test is indicated by a ∗on
the display every second. After 15 seconds the moisture content will be displayed on the
left-hand side of the LCD e.g. 14.9 % and the right-hand will show the average of all
measurements, e.g. 14.7%.
g) The analyzer will retain the chosen crop displayed and a new measurement can
be made by emptying the test cell and re-filling with a new sample.
Note: THOROUGHLY CLEAN THE CELL WITH THE WIRE BRUSH PROVIDED
BEFORE PERFORMING A NEW TEST
2. To select crop or function
a) When pressing the ON/TEST the analyzer will show software version and the country
code, followed by the level of battery and lastly the >> MEASURE<< menu. If you wish
to accept this choice of menu press the ON/TEST button and the desired crop can be
selected by use of the or arrows. If you wish to return to the >> MEASURE<<menu
simply press the ESC/OFF button and use the or arrow keys to cycle through the
menus.
3. How to adjust a calibration
Each calibration can, if required be adjusted individually, up or down by 3.9%mc in
increments of 0.1%mc.
a) Turn the analyzer by pressing ON/TEST. The menu >>MEASURE<< will be displayed.
b) Using the or arrow keys the menu >>CALIBRATION<< can be selected.
c) Accept the menu by pressing ON/TEST.
d) By using the or arrow keys select the desired crop.
e) Accept this choice by pressing ON/TEST. The numeric value +0.0 is shown on the
display.
f) By using the or arrow keys the desired calibration can be adjusted up to a
maximum of -3.9 to +3.9 %mc.
g) Accept the adjusted value by pressing the ON/TEST button.
Note: Where an adjustment has been made to a factory installed calibration a * will
appear at the end of the crop name.
Some calibrations are factory biased on delivery and should only be changed with
extreme caution.