SICK LMS 200 Installation guide

Quick Manual
MST DEMO setup
- 1-
© SICK AG · Auto Ident · Germany · All rights reserved
Quick Manual for
MST Demo setup
Software setup
and configuration
Version 1.0 August 2002 (HW/MV)

Quick Manual
MST DEMO setup
- 2-
© SICK AG · Auto Ident · Germany · All rights reserved
Contents
A. INTRODUCTION 3
1. Basics of the LMS System 3
B. HARDWARE SETUP 4
1. Before you start ... : Required components 4
2. Power supply 5
3. Serial interface for data exchange 7
C. COMMUNICATION VIA MST DEMO 9
1. Principle of MST-Demo 9
2. Installation of MST 9
2. Communication LMS to PC with MST 10
3. Record LMS scan data 11
4. Re-play the stored data 12
5. Editing the Configuration file of the recorder 13
6.Transferring the file LMSData.dat to a LMSData.txt file 14
D. APPENDICES: 18
1. LMS in High-Speed Mode 18
2. Further information and documentation 20
! This is a quick start manual. For extended information please refer to the documents
listed in section D.4. !

Quick Manual
MST DEMO setup
- 3-
© SICK AG · Auto Ident · Germany · All rights reserved
A. Introduction
1. Basics of the LMS System
The Laser Measurement System LMS 200, LMS
220, LMS 211, LMS 221, LMS 291 is based on a
time-of-flight measurement principle (Laser Radar).
A single Laser pulse is sent out and reflected by an
object surface within the range of the sensor. The
elapsed time between emission and reception of the
Laser pulse serves to calculate the distance between
object and LMS.
Via a integrated rotating mirror the Laser pulses
sweep a radial range in front of the LMS unit. A 2-
dimensional measurement field / detection area is
defined.
(For further details please refer to the Technical Description
Laser Measurement Systems SICK order no.: 8008970)
Main benefits of this measurement principle are:
•Detection of objects independent of object color
and surface structure
•Reliable detection of object presence
The Laser Measurement System LMS offers a
solution for a large range of applications:
•Determining the volumes of objects (measuring
packages, pallets, containers)
•Determining the position of objects (pallets,
containers, transport vessels)
•Collision prevention for vehicles or cranes
•Controlling docking processes (positioning)
•Classification of objects (vehicle detection, camera trigger)
•Process automation (e.g. calander rollers)
•Checking overhang/area monitoring in automated multi-storey car parks
•Determining the volumes or contours of bulk materials
•Monitoring open spaces for building security (facades, grounds, shafts)
•... and many more
Fig.1: Measurement principle LMS
Fig.2: Angular range LMS

Quick Manual
MST DEMO setup
- 4-
© SICK AG · Auto Ident · Germany · All rights reserved
B. Hardware setup
1. Before you start ... : Required components
As a minimum requirement to set up an LMS sensor with a data connection to a PC, the following
components are required:
For LMS200/LMS291:
Item/Specification Description SICK order number
LMS200-30106 10158501 x LMS unit with cable
connectors LMS291-S05 1018028
1 x power supply 24 V DC +-
15%, min. 2.5 A Power supply DC 24V / 2.5 A 6010361
Cable set 1: 5 m length 20189641 x cable for power supply
and output signals (open
cable ends
1 x cable for data interface Cable set 2: 10 m length 2018965
For LMS211/LMS220/LMS221:
Item/Specification Description SICK order number
LMS211-30206 1018023
LMS220-30106 1015945
1 x LMS unit with cable
connectors LMS221-30206 1018022
1 x power supply 24 V DC +-
15%, min. 2.5 A for LMS
electronics
Power supply DC 24V / 2.5 A 6010361
1 x power supply 24 V DC +-
15%, min. 5 A for LMS
heating
Power supply DC 24V / 10 A 6011156
1 x cable for power supply
and output signals (open
cable ends
1 x cable for data interface
Cable set: 5 m length 2019561
Note: In case of longer cables, please source your own cable. Please make sure to use correct
cable diameters to avoid voltage drops along large cable lengths.
•Suitable brackets, mounting.
A laptop or desktop computer serves as user interface with the LMS unit. The computer is
connected via a serial interface:
•RS232 interface (up to 10 m cable length) (standard COM1 or COM2 in most computers)
•RS422 interface (up to 1200 m cable length) (special interface card required)
In case a high-speed data transmission is required, SICK offers special interface boards for 500
kBaud transmission rate. For details and information please refer to section D.2.

Quick Manual
MST DEMO setup
- 5-
© SICK AG · Auto Ident · Germany · All rights reserved
2. Power supply
The LMS unit is supplied with 24 V DC +/- 15%. Depending on the length of the power supply
cable, the power supply needs to provide sufficient current.
Due to the LMS models, connection diagrams are different for LMS200/LMS291 and
LMS220/211/221.
a. For LMS 200 / LMS 291
LMS 200
LMS 291
Control Cabinet
Power Supply
24 V DC
Pin assignment:
Pin
No. Signal designation Input/
Output Wire
color
1GND_EXT (ground) Supply brown
2Restart Input blue
3VCC_EXT (24V DC +- 15%) Supply red
4Not connected - -
5OUT C (for field detection)/
weak signal Output grey
6Not connected - -
7Not connected - -
8OUT B (for field detection) Output turquoise
9OUT A (for field detection) Output orange
Note: Minimum connections marked in bold
PIN 5 / 8 / 9 cannot be used for data processing purposes with the LMS sensor

Quick Manual
MST DEMO setup
- 6-
© SICK AG · Auto Ident · Germany · All rights reserved
b. For LMS 211 / LMS 220 / LMS 221
LMS 211
LMS 220
LMS 221
Control Cabinet
Power Supply
24 V DC
Heating
Electronics
Power Supply
24 V DC
Pin assignment:
Pin
No. Signal designation Input/
Output Remark
1RxD- Interface
2RxD+ Interface
3RS232/RS422 Jumper 1 With jumper àRS422
Without jumper àRS232 (default)
4GND_EXT for heating Supply
5Flap control Input For flap control details please refer to
technical description
6Not connected -
7OUT C (for field detection)/
weak signal Output
8GND_EXT for electronics Supply
9TxD- Interface
10 TxD+ Interface
11 RS232/RS422 Jumper 2
12 VCC_EXT for heating Supply
13 OUT A (for field detection) Output
14 OUT A (for field detection) Output
15 Restart
16 VCC_EXT for electronics Supply
Note: Minimum connections marked in bold
Use separate cables for connecting the scanner electronics and the heating.
Fig.2: Connections on backplane of LMS211/LMS220/LMS221

Quick Manual
MST DEMO setup
- 7-
© SICK AG · Auto Ident · Germany · All rights reserved
3. Serial interface for data exchange
For data exchange in measuring tasks, the LMS is equipped with a serial RS 232 / RS 422
interface.
By setting a jumper, RS422 or RS232 can be enabled. Default setting is RS232.
a. Connection of the RS 232 interface.
1. For LMS 200 / LMS 291:
Pin assignment:
LMS PC
Signal designation PIN No. PIN No. Signal designation
Not connected 1 1 NC
RxD 2 2 RxD
TXD 3 3 TXD
Not connected 4 4 Not connected
GND 5 5 GND
Not connected 6 6 Not connected
Not connected 7 7 Not connected
Not connected 8 8 Not connected
Not connected 9 9 Not connected
Note: PINs 2 and 3 are crossed in the cable.
The connector at the PC side is a 9 pin DSub
2. For LMS 211 / LMS 220 / LMS 221
Note: The connector at the PC side is a 9 pin DSub
Fig.3: Wiring of the RS232 for LMS211/LMS220/LMS221

Quick Manual
MST DEMO setup
- 8-
© SICK AG · Auto Ident · Germany · All rights reserved
b. Connection of the RS 422 interface
1. For LMS 200 / LMS 291
Pin assignment:
LMS PC
Signal designation PIN No. PIN No. Signal designation
RxD- 1 1 RxD-
RxD+ 2 2 RxD+
TxD- 3 3 TxD-
TxD+ 4 4 TxD+
GND 5 5 GND
Not connected 6 6 Not connected
Jumper 1 7 7 Not connected
Jumper 2 8 8 Not connected
Not connected 9 9 Not connected
Note: As there is no standard for RS422 connector pinning, please check the signal
assignments of your RS422 interface card to the respective PINs. Refer to the
documentation of your RS422 interface board manufacturer.
In case of deviations, change the cable assignments accordingly.
For selection of the RS422, there needs to be a jumper connection between between
PINs 7 and 8 on the LMS side inside the cable connector.
PINs 2 and 3 are crossed in the cable, PINs 1 and 4 are crossed in the cable. It is
recommended to use twisted pair cables for RS422 connections.
2. For LMS 211 / LMS 220 / LMS 221
Note: For selection of the RS422, there needs to be a jumper connection between between
PINs 3 and 11 on the LMS side in the cable connector.
PINs 2 and 3 are crossed in the cable, PINs 1 and 4 are crossed in the cable. It is
recommended to use twisted pair cables for RS422 connections.
Fig.3: Wiring of the RS232 for LMS211/LMS220/LMS221

Quick Manual
MST DEMO setup
- 9-
© SICK AG · Auto Ident · Germany · All rights reserved
C. Communication via MST Demo
1. Principle of MST-Demo
The MST Demo Tool is made for two main purposes. Primarily it demonstrates what can be done
with the software package MST 200 and what functions are available for custom software
development. On the other hand it is a tool for quick and easy recording of scans for a
preliminary data evaluation. The feasibility of the LMS sensor as a data source for a specific
application can be evaluated with this tool.
2. Installation of MST
MST is delivered as a ZIP file. Please create a dedicated MST home directory on your harddrive.
After unpacking the ZIP file into the home directory, you get the following file structure.
By starting MSTDemo.exe you will get the MST start screen with an active menu (Start | Options).

Quick Manual
MST DEMO setup
- 10 -
© SICK AG · Auto Ident · Germany · All rights reserved
2. Communication LMS to PC with MST
MST Provides a simple data connection. The Hardware connection between PC and LMS is
shown in section B3.
Please select Options in the start window and set your connection parameters.
Note: Always select 9600Bd . MST will automatically use the highest Baud rate possible
when started.
The Software is now ready.
Within the Start button you will have several applications which can be started.
For every application MST_Demo will create a configuration file in the MST home directory. This
configuration file is a text file and can be edited with every editor.
In this quick start manual the application Record LMS scan data and Play back LMS scan data
are discussed. The other applications follow the same idea by principle.
Other manuals for LMS 200
1
This manual suits for next models
4
Table of contents
Other SICK Measuring Instrument manuals

SICK
SICK RFH620 User manual

SICK
SICK FLOWSIC100 Flare User manual

SICK
SICK GM901-02 User manual

SICK
SICK NAV-LOC User manual

SICK
SICK Ruler3000 User manual

SICK
SICK WSE16 Series User manual

SICK
SICK WTF9-3 User manual

SICK
SICK MARSIC300 Manual

SICK
SICK Mounting kit 2 User manual

SICK
SICK FLOWSIC30 Technical Document