SGM LEKTRA RPmag Guide

RPmag
electromagnetic induction fl ow measurement with data logger
technical documentation EN Rev. C

Page 2 of 64 www.sgm-lektra.com
RPmag - contents
CONTENTS
1-WARRANTY
2-CALIBRATION CERTIFICATE
3-PRODUCT
4-PERFORMANCE SPECIFICATIONS
5-FLOW RANGE
6-DIMENSION
7-INSTALLATION
8-ELECTRICAL CONNECTIONS
9-LOCAL OPERATOR INTERFACE (LOI)
10-PROGRAMMING
11-TROUBLESHOOTING
12-FACTORY TEST AND QUALITY CERTIFICATE
pag. 3
pag. 3
pag. 4
pag. 5
pag. 6
pag. 8
pag. 13
pag. 21
pag. 31
pag. 35
pag. 61
pag. 64

Page 3 of 64
www.sgm-lektra.eu
2-CALIBRATION CERTIFICATE
All the electromagnetic fl owmeter are tested by 3 point rigs calibration.
The producer releases a document on letterhead certifying the average error of the 3-point calibration.
The calibration certifi cate is supplied with the unit.
The company archives the test data of each electromagnetic fl owmeter.
The calibration rig is certifi cated by N.I.M. (National Institute of Metrology), which is internationally recognizedby B.I.P.M.
(Bureau International des Poids et Metrologie) and complies with NTC ISO IEC 17025 standard.
All calibrations are made in accordance to EN 45001 standards and with an accuracy better than 99.97%
RPmag - warranty / calibration certifi cate
Products supplied by SGM LEKTRA are guaranteed for a period of 12 (twelve) months from delivery date according to
the conditions specifi ed in our sale conditions document.
SGM LEKTRA can choose to repair or replace the Product.
If the Product is repaired it will maintain the original warranty terms, whereas if the Product is replaced it will have 12
(twelve) months of warranty.
The warranty will be null if the Client modifi es, repair or uses the Products for other purposes than the normal conditions
foreseen by instructions or Contract.
In no circumstances shall SGM LEKTRA be liable for direct, indirect or consequential or other loss or damage whether
caused by negligence on the part of the company or its employees or otherwise howsoever arising out of defective goods
1-WARRANTY

Page 4 of 64 www.sgm-lektra.com
RPmag - features
3- PRODUCT
COMPACT VERSION
1. Sensor
2. Converter
REMOTE VERSION
1. Sensor
2. Connection housing
3. Connection cables
4. Converter, wall mouting
3.1 IDENTIFICATION
Each meter has an adhesive identifi cation plate on which are the meter main data. The following picture describes the
information and data on the identifi cation plate.
1. Product code
2. Serial number
3. Production batch
4. Power supply
9. Protection
10.QR code, connecting a
product web page
5. Process connection
6. Lining material
7. Electrodes material
8. Sensor factor
Model
Serial n°
Meter n°
P.Supply
Connection
Lining
Electrode
Sens. Factor
RPMAGN0032B2B1A0A1
FM005141720
032377
85
÷
265Vac 50
÷
60Hz
DN32 PN40
RUBBER
SS316
0.2293
IP67

Page 5 of 64
www.sgm-lektra.eu
4-FEATURES
RPmag - features
Flow rate range
RPmag is able to process signals from fl uids with fl ow rates of up to 10m / s in both directions (bidirectional meter).
Range dimension / lining material
PTFE DN10 ÷ DN500 / RUBBER DN65 ÷ DN2000
Sensor material
SS321
Housing material
epoxy painting aluminium
Electrodes material
SS316L - Hastelloy C - Titanium - Tantalum - Platinum
Measure range
<0,1m3/h ÷ >110000m3/h
Accuracy
±0,5% standard; ±0,2% optional
Repeatability
±0,1%
Fluid conductivity
>5μS/cm.
Power supply
85÷265Vac, 24Vac/dc, 12Vdc.
Consumption
6W, max. 8W.
Ambient Temperature Limits
Remote version operating temperature: RUBBER -10 ÷ +80°C; PTFE -40 ÷ +150°C
Compact version operating temperature: RUBBER -10 ÷ +80°C; PTFE -40 ÷ +100°C
Storage temperature: -40÷85°C
Communication protocol
Modbus RTU or Bluetooth App Android (opt.) or Hart (opt.)
Data Logger
Internal data logger to USB pen drive for fl ow measurements and analog inputs storing;
the measurement storage interval can be set from 15 to 3600 seconds
Output
4÷20mA: 0÷500Ω
Frequency output: 0,1÷10000 Hz
Pulse output: 24Vdc galvanically isolated or open collector galvanically
isolated 24V 20mA (opt)
Alarm output: 2 relays, 3A 230Vac N.O.
Input signals
RPmag has 2 active analog inputs at 24Vdc for 2-wire transmitters connection (eg. Temperature or
pressure) and 1 digital input for an external contact connection for the integrated batch function restart and for
partial totalizer management.
Reverse Flow
Allows measure and totalization of reverse fl ow.
Output Testing
Relays output: Transmitter can switch relays at testing value.
Current Source: Transmitter can be commanded to supply a specifi ed test current between 4.0 and 20.0 mA.
Frequency Source: Transmitter can be commanded to supply a specifi ed test frequency between 1 and 10000 Hz.
Low Flow Cutoff
Adjustable. Below selected value, instantaneous fl ow and outputs are driven to the zero fl ow rate signal level.
Humidity Limits
0-100% RH to 150 °F (65 °C), not condensing.
Damping
Adjustable between 1 and 99 seconds.
Compact version IP rating
IP67
Remote version IP rating
sensor IP67 / IP68 (by request) - converter IP67
Anti-condensation fi lter
Anti-condensation fi lter installed on converter

Page 6 of 64 www.sgm-lektra.com
5-FLOW RANGE
RPmag - fl ow range
5.1 FLOW RANGE GRAPHIC
Flow range from DN3 to DN500 (starting from DN10) Flow range from DN600 to DN2000

Page 7 of 64
www.sgm-lektra.eu
RPmag - fl ow range
5.2 FLOW RANGE TABLES
5.3 LOAD LOSS
Adaptation cones
DN10 ÷ 300
DN
(mm)
Range: Minimum (0,5 m/s) /
Maximum (10 m/s)
10 0.14 ÷ 2.9 m3/h
15 0.3 ÷ 6 m3/h
20 0.5 ÷ 12 m3/h
25 0.6 ÷ 18 m3/h
32 1 ÷ 30 m3/h
40 1.8 ÷ 42 m3/h
50 3 ÷ 66 m3/h
65 5.8 ÷ 120 m3/h
80 8.9 ÷ 180 m3/h
100 11 ÷ 282 m3/h
125 20 ÷ 450 m3/h
150 30 ÷ 600 m3/h
200 50 ÷ 1100 m3/h
250 85 ÷ 1700 m3/h
300 110 ÷ 2400 m3/h
DN350 ÷ 2000
DN
(mm)
Range: Minimum (0,5 m/s) /
Maximum (10 m/s)
350 180 ÷ 3300 m3/h
400 220 ÷ 4200 m3/h
450 270 ÷ 5400 m3/h
500 320 ÷ 6600 m3/h
600 490 ÷ 9600 m3/h
700 680 ÷ 13500 m3/h
800 900 ÷ 18000 m3/h
900 1200 ÷ 22500 m3/h
1000 1450 ÷ 28000 m3/h
1200 2500 ÷ 40000 m3/h
1400 3000 ÷ 55000 m3/h
1600 4000 ÷ 70000 m3/h
1800 5000 ÷ 90000 m3/h
2000 6000 ÷ 110000 m3/h

Page 8 of 64 www.sgm-lektra.com
RPmag - dimensions
6-DIMENSIONS
6.1 REMOTE VERSION CONVERTER
215,5 mm
177 mm 170 mm
6.2 WALL MOUNTING REMOTE VERSION CONVERTER
150 mm
110 mm

Page 9 of 64
www.sgm-lektra.eu
6.3 COMPACT VERSION DN10 ÷ DN80 PN16 - PN40
RPmag - dimensions
DN
(mm)
A
(mm)
PN 16 - PN 40
B
(mm)
ØD
(mm)
10
200
295 90
15 295 95
20 300 105
25 300 115
32 315 140
40 335 150
50 344 165
65 360 185
80 375 200
A
B
Ø D

Page 10 of 64 www.sgm-lektra.com
RPmag - dimensions
6.4 COMPACT VERSION DN100 ÷ DN1000 PN10 - PN16 - PN40
A
B
Ø D
DN
(mm)
A
(mm)
PN 10 PN 16 PN 40
B
(mm)
ØD
(mm)
B
(mm)
ØD
(mm)
B
(mm)
ØD
(mm)
100 250 - - 400 220 410 235
125 250 - - 420 250 435 270
150 300 - - 460 285 468 300
200 350 520 340 520 340 538 375
250 450 570 395 575 405 598 450
300 500 620 445 620 460 648 515
350 550 670 505 678 520 708 580
400 600 730 565 738 580 778 660
450 600 780 615 793 640 816 685
500 600 830 670 850 715 870 755
600 600 930 780 960 840 985 890
700 700 1050 895 1080 910 - -
800 800 1165 1015 1170 1025 - -
900 900 1270 1115 1275 1125 - -
1000 1000 1360 1230 1375 1255 - -
Other manuals for RPmag
1
Table of contents
Other SGM LEKTRA Measuring Instrument manuals

SGM LEKTRA
SGM LEKTRA Transit Time SGM-200H Guide

SGM LEKTRA
SGM LEKTRA SGM-100F User manual

SGM LEKTRA
SGM LEKTRA SGM-101H Guide

SGM LEKTRA
SGM LEKTRA MUXM User manual

SGM LEKTRA
SGM LEKTRA SGM-101F Guide

SGM LEKTRA
SGM LEKTRA RPmag Guide

SGM LEKTRA
SGM LEKTRA Pmag User manual

SGM LEKTRA
SGM LEKTRA FLOWMETER User manual