The following are some common messages that may be displayed, including error messages. If an
error message other than those listed below is displayed, please contact technical support in the USA
at (803) 329-0162 (ext. 0).
LCD Message
HI
LO
LO
ER
Description
In READ mode: test sample concentration is above
the measurement range (test specific).
In READ mode: test sample concentration is below
the measurement range (test specific).
In ZERO mode: sample absorbance (due to a
cloudy or colored sample or a dirty cell) is too high
to zero, the meter will read “LO”.
Excessive stray light detected. Normally this
does not occur, even when testing in sunlight.
Low battery indication.
Corrective Action
Dilute and retest. Dilution Kit available
(Part Number 487200)
Sample value is below measurement
range.
Dilute sample, filter sample, or clean cell.
Testing cannot proceed until a valid ZERO is
acheived.
Place the LIGHT BLOCKING CAP over the CELL
for zeroing and for reading result. Moving
to a shaded area can also fix this problem.
Replace the batteries.
About The Built-In Cell
The built-in CELL is transparent plastic and, when filled to the top, contains 4ml. The sturdy CELL
design will last for over 20,000 readings. Scratches on the CELL will not interfere or compromise
the accuracy of the readings because of its fixed position. For best accuracy, rinse cell with clean
water immediately after a test is completed. Do not use solvents, such as acetone, to clean the
cell. When the CELL becomes stained or cloudy from repeated testing, or when the meter does
not blank when you press the ZERO/ON button, the cell needs to be cleaned. Clean as follows: Fill
cell with clean water and add two drops of Acid-1 or HCl-1. Leave meter undisturbed for 5
minutes. Afterwards, rinse the cell and the meter is ready for use again. Cleaning the cell
regularly will not be necessary if you rinse the CELL immediately after the test.
eXact®LEADQuick Meter Messages
eXact®Photometer 2-Year Limited Warranty
Registration of your eXact
®
photometer must be received within 30 days from date of purchase to
activate the warranty. The eXact
®
photometer is warranted to be free from defects in materials and
workmanship for a period of two (2) years from the date of purchase by the customer. ITS will repair
or replace any part of the product which is deemed to be faulty or otherwise defective. The
non-transferable warranty does not cover product damage caused by abuse (such as crushing a tablet
in the cell) or improper use. If the meter is faulty or otherwise defective contact ITS by phone
(+1-803-329-9712
Ext.
0)
or
email
(
[email protected])
to
describe
the
problem
and
obtain
a
return
authorization form before returning the photometer to ITS. Damage caused by improper packing of the
photometer for return shipment to ITS will not be covered by the warranty. Customer is responsible for
shipping charges to ITS. ITS pays postage when photometer is returned to customer. A maximum
processing fee of $120 will be charged for repair or replacement of non-registered photometers and
damages not covered by this warranty. Registration is available over the phone (+1-803-329-9712
Ext. 0) or online at http://www.sensafe.com/micro/warranty/ (Personal data is kept confidential).
1. Unscrew the O-ring sealed battery cover counter-clockwise. Use proper sized pliers if
necessary. Do not disturb the sealing O-ring. Batteries are not included.
2. Remove the used batteries and install 4 new AAA batteries following the diagram for
correct polarity (see diagram). We recommend high quality AAA alkaline batteries be used.
4. Replace the battery cover. Be sure to tighten the cover securely. This is necessary for meter
to be waterproof.
5. Dispose of the used batteries in accordance with
your local regulations.
6. Press ZERO/ON button to confirm the meter
turns on. The meter is now ready for operation.
7. Meter will not work if battery orientation is incorrect.
To Install/Replace “AAA” Batteries:
TOP VIEW
9