Scientific SM6020 User manual

User Manual
LCR Meter
S 6020M

Scientific Mes-Technik Pvt. Ltd.,
B-14 Industrial Estate, Pologround,
INDORE 452015(INDIA)
Tel : 0731-2422330/31/32/33
Fax : 0731-2422334,561641
Email : info@scientificindia.com
Copyright Scientific All rights reserved.
This instrument contains proprietary information, no part of this manual may
be photocopied , reproduced or translated without any prior written
consent.
Information in this manual supercede all corresponding previous released
material.
Scientific continues to improve products and reserves rights to amend part
or all of the specifications , procedures, equipment at any time without
notice.
Rev 1.02 / 0721
2 S 6020 M User Manual

LCR Meter 6020SM
Table of Contents
1. Introduction 4
2. Technical Specifications 5
3. Terms & Symbols 7
4. Front Panel Controls 10
5. Operating Instructions 13
General Information 13
Safety 13
Operating Conditions 13
First Time Operation 14
Understanding & Measuring Passive Components 14
Understanding Displayed Parameter 16
Measurement of Transformer Parameters 17
Component Sorting & Binning Function 19
Equivalent Circuits 19
Display Types 20
Menu 20
Zero Compensation 22
Frequency of Test 23
Impedance Range 24
Precision & Accuracy 25
Basic Accuracy Graph 27
Remote Programming 29
6. Maintenance 34
7. Despatch Procedure for Service 34
8. Warranty 35
3
S 6020 M User Manual

LCR Meter S 6020M
Comprehensive range of functions, L+Q, C+D, R+Q, |Z|+ , R+X,
G+B, L+AL etc.
Measurement accuracy 0.2 %
Test frequencies standard 100 Hz to 25 kHz in 11 steps, and an
additional User defined, from 500 possible frequencies between
100 Hz and 25 kHz
Simultaneously two measured parameters displayed
Absolute value, value, % value
Bin function for component sorting
Save & Recall of 25 measurement values
Compact and Low Cost
The LCR Meter is LC measuring bridge, an impedance measuring S 6020M R
instrument allowing automatic measurement of resistance, capacitance ,
inductance , measurement of some parameters of Transformer 1m to 100 M ,
The bridge has basic accuracy of 0.2% and has 11 standard Frequencies and one
definable / selectable by the user. The instrument has a powerful microcontroller
with logic circuits, allowing to control the display, keyboard and component
measurement.
S 6020M allow to store 25 measurements in internal memory and recall at any
given time, the memory is non volatile .
An internal bias voltage of 2 Volts can be selected for use when testing electrolytic
capacitors. This instrument is equipped with self test and self calibration menu , for
calibrating the instrument at regular interval without opening the case.
4S 6020 M User Manual

Technical Specifications
5
S 6020 M User Manual

6S 6020 M User Manual

Terms & Symbols
Parameter Measurement Symbol of Unit
Z Complex impedance ohm
Y Admittance, 1/Z Siemens, S
| Z | Module impedance Z, ohm,
Rs or ESR Resistance Series, ohm
Real part of the Impedance
X Reactance ohm
Imaginary part of the Impedance
G Conductance,
Real Part of admittance (Y) Siemens, S
B Susceptance, Siemens, S
Imaginary Part of admittance Siemens, S
Cs Capacity Series Farad, F
Cp Parallel capacity Farad, F
Ls Series Inductance Henry, H
Lp Parallel inductance Henry, H
Rp Parallel resistance ohm
Q quality Factor Without unit
D dissipation factor Without Unit
Phase angle Degree
M Mutual inductance Henry H
N Turns Ratio Without Unit
Vp Primary voltage Volts AC, V
(Transformer)
Vs Secondary voltage Volts AC, V
(Transformer)
7
S 6020 M User Manual

Unit of Measure
Multiplier Scientific Engineer Symbol
1000000 10 Mega M
6
1000 10 kilo k
3
0.001 10 milli m
-3
0.000001 10 Micro
-6
0.000000001 10 nano n
-9
0.000000000001 10 pico p
-12
Glossary
Coil : A coil is made of a multiple turns of wire on conductor / Core. It has the
property to oppose / resist changes in current. It is characterized by its Inductance.
Capacity : Characteristics of a Capacitor. The unit is the Farad (F).
Capacitor : Passive element consists of two conductors / plates separated by a
dielectric. The capacitor has the property of passing the AC while blocking the DC.
See also chapter capacity and reactance.
Test frequency : The frequency with which the measurement of parameters of
component is done. The value of a parameter sometime depend on frequency.
Range : Range of resistance that the instrument uses to perform measuring
function.
Impedance : Complex parameter defining a passive component. It include a real
part (resistance) in series with some imaginary part (reactance ). Standard or pure
resistance has no reactance, while perfect coils and capacitors have resistance.
Inductance : Property of a coil to oppose any change in current . The inductance of
a coil is in Henry (H).
8S 6020 M User Manual

Parameter : Electric property measured, The main or primary parameter is the
property of most commonly sought component (capacity, inductance, Resistance).
The second parameter of lesser importance characterized losses in the
component (quality factor, dissipation, phase angle).
Accuracy : The difference between the measured value and the true value of
component. The accuracy is expressed as a percentage for the principal
parameter. The accuracy depends on the value of the impedance and the
frequency. Typically the accuracy is a secondary parameter and is great
importance.
Accuracy Background : The accuracy of LCR bridge depends on a number of
parameters such as the frequency and impedance. The accuracy of the standard
programmed database is checked with precision instrument in optimum
conditions, i.e. at frequency of 1 kHz and for 100 and 10 k .
Source Resistance : Resistance in series with the testing signal generator. This
resistance relates to the range of impedance.
Resolution : The smallest value that can be shown on the display of the
instrument. Do not confuse this with the accuracy.
Test Voltage : rms This is the value of the alternating voltage that generates a
current in the test component. Because of resistance and impedance of
component under test, test voltage is always less than specified value.
9
S 6020 M User Manual

Power : When push button pressed , switches power ON to the
instrument.
Display : LCD displays the measured value & parameters selected, the
status of the instrument and different messages. In measuring mode the
normal primary value (L, C, R, | Z |) is displayed on the first line while the
associated secondary parameter (Q, D, X, ) is displayed in the second
line. The position of the decimal point is adjusted according to the function
automatically. The symbol front of the main parameter indicates that
measurement is a relative measure or an absolute deviation compared to a
nominal value. The presence of an arrow to the left of the first line indicates
that the instrument is in automatic mode. The presence of a single arrow to
the left of the second line indicates the instrument is placed in automatic
equivalent circuit / model mode. The range of impedance is shown right from
the second line. A letter ‘A’ form line number indicates that the instrument is in
the auto range. In manual mode or maintaining same range mode an “H”
becomes flashing. The frequency of testing is underway displayed on the
r i g h t
side of the second line. Test signal level is shown by HI and Lo on
first line on right side. Bias can be selected for biasing components and
shown by arrow on right side of display by arrow when signal is biased.
Front Panel Controls
1
2
AUTO LEVEL BIAS
RANGE
FREQ
MODEL/AUTO
AUTO : Push Long
Freq Set
HDHSLSLD
POWER
ON/OFF
LCR METER 6020 SM
Discharge test device before connecting
Cal
Disp
Menu
Eq Ckt
L/C/R N/Vs/M Freq Range
1
2
3 4
Fig 1
Save/Rcl
10 S 6020 M User Manual
Table of contents
Other Scientific Measuring Instrument manuals

Scientific
Scientific SM6025 User manual

Scientific
Scientific SM6026 User manual

Scientific
Scientific SME1402 User manual

Scientific
Scientific SME1202A User manual

Scientific
Scientific SMS3000X User manual

Scientific
Scientific SM6016 User manual

Scientific
Scientific SM6024 User manual

Scientific
Scientific SME1415 User manual

Scientific
Scientific SM6033 User manual

Scientific
Scientific SM6015A User manual