SCIENTIFIC INSTRUMENTS SCM10 User manual

Scientic Instruments, Inc. Model SCM10
1
090-267 Rev A
Copyright 2012
Scientic Instruments, Inc.
All Rights Reserved.
Printing History
Rev A
Certication
Scientic Instruments, Inc. (SII) certies that
this product met its published specications
at the time of shipment. SII further certies
that its calibration measurements are
traceable to the United States National
Institute of Standards and Technology (NIST).
Warranty
This product is warranted against defects in
materials and workmanship for a period of
one year from date of shipment. During this
period SII will, at its option, either repair or
replace products which prove to be defective.
For products returned to SII for warranty
service, the Buyer shall prepay shipping
charges and SII shall pay shipping charges
to return the product to the Buyer.
However, the Buyer shall pay all shipping
charges, duties, and taxes for products
returned to SII from another country.
Warranty Service
For warranty service or repair, this product
must be returned to a service facility
designated by SII.
Limitation of Warranty
The foregoing warranty shall not apply to
defects resulting from improper or inadequate
maintenance by the Buyer, Buyer supplied
products or interfacing, unauthorized
modication or misuse, operation outside
of the environmental specications for the
product, or improper site preparation or
maintenance.
The design and implementation of any circuit
on this product is the sole responsibility of
the Buyer. SII does not warrant the Buyer’s
circuitry or malfunctions of this product that
result from the Buyer’s circuitry.
In addition SII does not warrant any damage
that occurs as a result of the Buyer’s circuit
or any defects that result from Buyer-supplied
products.
Notice
The information contained in this document
is subject to change without notice.
SII makes no warranty of any kind with regard
to this material, including, but not limited to,
the implied warranties of merchantability and
tness for a particular purpose.
SII shall not be liable for errors contained
herein or for incidental or consequential
damages in connection with the furnishing,
performance, or use of this material. No
part of this document may be photocopied,
reproduced, electronically transferred, or
translated to another language without
prior written consent.
Trademark Acknowledgement
All product and company names are
trademarks or trade names of their respective
companies.
Safety
The Model SCM10 does not contain any user
serviceable parts. Do not open the enclosure.
Do not install substitute parts
or perform any unauthorized modication
to the product. For service or repair,
return the product to SII or an authorized
service center.

Scientic Instruments, Inc. Model SCM10
2090-267 Rev A
TABLE OF CONTENTS
Introduction ...................................................................................................5
Product Description .................................................................................5
Product Specications .............................................................................6
Input Specications ..........................................................................6
Analog Output Specications ...........................................................7
Front Panel.......................................................................................7
Relays and Alarms ...........................................................................7
Interface ...........................................................................................8
Input Power Specications ...............................................................8
General ............................................................................................8
Power Supply Specications ............................................................8
Safety Summary ......................................................................................8
Related Safety Symbols ..........................................................................9
Absolute Maximum Ratings.....................................................................9
Instrument Summary..................................................................................10
General..................................................................................................10
Inspection and Unpacking .....................................................................10
Rear Panel Denition ............................................................................10
Input Power Connector.......................................................................... 11
I/O Connector ........................................................................................ 11
Serial Connector....................................................................................12
Ethernet Connector ...............................................................................12
Sensor Input ..........................................................................................13
Analog Output .......................................................................................13
Alarms and Relays ................................................................................14
Front Panel Operation ................................................................................15
Instrument Power up .............................................................................15
Display Denition...................................................................................15
Display Units & Alarm Annunciators ......................................................15
Keypad Operation .................................................................................16
Instrument Reset ...................................................................................16
Menu Structure ......................................................................................16
Input Setup ............................................................................................18
Alarm Setup...........................................................................................19
Relay Setup ...........................................................................................20
Analog Output Setup .............................................................................20
System Settings ....................................................................................21
Network Settings ...................................................................................22
Remote Interface ........................................................................................23
General..................................................................................................23
Ethernet Connection..............................................................................23
Ethernet Conguration ..........................................................................23
Serial RS232 Physical Connection........................................................25
Serial RS232 Connector – Hardware Support & Data Format ..............26
ASCII Message String Format...............................................................26
ASCII Command Summary ...................................................................27

Scientic Instruments, Inc. Model SCM10
3
090-267 Rev A
Graphical User Interface (GUI) ..................................................................33
Installation .............................................................................................33
Operation...............................................................................................33
Starting GUI for the rst time..........................................................33
Status Page....................................................................................34
Alarm and Relay LED status Indicators ..................................34
Data Logging...........................................................................34
Conguration Page ........................................................................35
Selecting Sensor Type ............................................................36
Setting the Alarms ...................................................................36
Settings the Relays .................................................................37
Adjusting the Analog Output ...................................................37
Changing Display Filter ...........................................................37
Switching Communication Modes ...........................................37
Curve Handler: Downloading a calibration curve ...........................38
Curve Header and Data Format..............................................38
Preparing a curve for download ..............................................39
Downloading the curve ...........................................................39
Hardware Calibration .....................................................................40
Service .........................................................................................................41
Troubleshooting .....................................................................................42
Instrument Calibration ...........................................................................42
Sensor input Calibration .................................................................44
Current Source Calibration .....................................................44
ADC Offset & Gain Calibration ................................................45
Analog Output Calibration ..............................................................47
Analog Output Current Mode Calibration ................................47
Analog Output Voltage Mode Calibration ................................48

Scientic Instruments, Inc. Model SCM10
4090-267 Rev A

Scientic Instruments, Inc. Model SCM10
5
090-267 Rev A
1. INTRODUCTION
PRODUCT DESCRIPTION
The SCM10 single channel monitor provides the accuracy, resolution and
interface features of a bench top temperature monitor in an easy to use,
compact and easily integrated instrument. With appropriate sensors, the SCM10
measures temperature from 1.4K to 800K and in difcult sensing conditions,
including high vacuum and magnetic elds. SCM10 has two programmable
alarms, two relays and a user congurable analog voltage or current output.
Serial interface and Ethernet interface are provided for remote operation.
A PC-compatible utility software is provided with the SCM10. This graphical user
interface provides an easy way to set all instrument parameters and evaluate
the performance of the instrument. The SCM10 also has an embedded web
server that can be used to connect to the instrument from anywhere within the
LAN using an internet browser. The instrument web page shows the current
temperature reading and shows the status of all alarms and relays.
The SCM10 Temperature Monitor supports diode and RTD sensors. The
instrument can be congured for the appropriate sensor type using the
front panel or the utility software. Two convenient sensor excitation currents
are provided to choose from. The SCM10 provides a 4 –leaded differential
measurement for accurate temperature measurements. Measurements are
available in temperature units K, ˚C or ˚F , or sensor units V or Ω.
The SCM10 has 9 standard sensor curves that are programmed into the
instrument from the factory. It also provides room for one 200 point user curve,
which may be loaded into the instrument using the utility software.
With a serial, Ethernet, and convenient graphical user interface, as well as an
ASCII command set and LabVIEW™ drivers, the SCM10 can be used as a
stand-alone temperature monitor or can be easily integrated into other systems.

Scientic Instruments, Inc. Model SCM10
6090-267 Rev A
PRODUCT SPECIFICATIONS
Input Specications
• Number of Inputs: 1
• Measurement Type: 4-leaded differential
• Excitation Current: Constant Current (10µA or 1mA)
• Isolation: Measurement not isolated from chassis ground
• A/D Resolution: 24 bit
• Input Accuracy: Sensor Dependent
• Measurement Resolution: Sensor Dependent
• Measurement Rate: 5 rdgs/s
• Supported Sensors:
Diodes Silicon, Gallium Arsenide (25K & above)
RTDs RuO 1kΩ, RuO 100kΩ, Pt 100Ω, Pt 1000Ω, Cernox™, Carbon-Glass
(4K & above)
• Standard Curves: Si410, Si430, DT470, DT670, CTI, Pt100, Pt1000,
RO600, RO105
• User Curve: One user curve in non-volatile memory (up to 200 points)
• Settings: Sensor Type, Sensor Curve
• Input Connector: DB15
Sensor Performance Specications
Sensor Type Silicon Diode Ruthenium Oxide Platinum
Model Number Si410 RO105 Pt100
Temperature
Range
1 to 450K 2.0 to 273K 30 to 800K
Sensor Units Volts Ohms Ohms
Input Range 0 to 5V DC 0 to 500 kΩ 0 to 500 Ω
Sensor
Excitation
10µA 10µA 1mA
Electronic
Resolution
10µV 1 Ohm 1 mOhm
Sensor
Sensitivity
-36.36mV/K at 4K
-1.701mV/K at 77K
-2.293mV/K at 300K
12.652kΩ/K at 4K
174.55Ω/K at 77K
19.52 Ω/K at 270K
0.231Ω/K at 36K
0.422Ω/K at 77K
0.386Ω/K at 300K
Electronic
Accuracy
Sensor Units:
±150uV ±60Ω ±0.05Ω
Temperature: ±3.7mK at 4K
±87mK at 77K
±66mK at 273K
±4.5mK at 4K
±0.31K at 77K
±2.8K at 273K
±0.21K at 36K
±0.12K at 77K
±0.13K at 273K

Scientic Instruments, Inc. Model SCM10
7
090-267 Rev A
Analog Output Specications
• Isolation: Output not isolated from chassis ground
• Update Rate: 2 rdg/s
Voltage Current
Range: 0-10 V 4-20mA
Resolution: 0.15 mV 0.3 μA
Accuracy: ±1.25 mV ±2.5 μA
Min. Load
Resistance
500 Ω
(short-circuit protected)
N/A
Compliance
Voltage
N/A 10 V
Load
Regulation
N/A ±0.02% rdg (0 to 500Ω)
• Settings: Scale Voltage or Current with Temperature, Sensor Units or
Natural Logarithm of Sensor Units
• Connector: DB15
Front Panel
• Display Type: 6 digit LED
• Display Units: K, °C, °F, V, Ω
• Display Update Rate: 2 rdg/s
• Temperature Display Resolution: 0.001° between 0-99.999°,
0.01° between 100-999.99°, 0.1° above 1000°
• Sensor Units Display Resolution: 100µV (for diode type), Sensor Dependent
(for RTD type)
• Keys: Menu , ▼(Down ),▲(Up), Enter
• Front Panel Features: Alarm LEDs, View & Edit Conguration,
Instrument Reset
Relays and Alarms
Alarms:
• Number: 2
• Settings: Enabled or Disabled, High or Low Setpoint, Deadband,
Latching or Non-Latching
• Actuators: Front Panel LEDs, Relays
Relays:
• Number: 2
• Contacts: Normally Open (NO), Normally Closed (NC) and Common (COM)
• Contact Rating: 24VDC @ 1A
• Settings: Manually ON, Manually OFF, Follow Alarms
• Connector: DB15

Scientic Instruments, Inc. Model SCM10
8090-267 Rev A
Interface
Serial Interface:
• Format: RS232
• Baud Rate: 9600, 19200, 57600, 115200
• Functionality: Graphical User Interface, LabVIEW™ Drivers (Version 8.0)
• Connector: DB9 (DCE)
Ethernet Interface:
• Format: 10 Base T
• Speed: 10 Mbps
• Functionality: Graphical User Interface, LabVIEW™ Drivers,
Embedded Web Server
• Connector: RJ-45
Input Power Specications
• Power Requirement: Regulated 9-24V DC @ 300mA
• Connector: Barrel Plug 5.5mm OD x 2.5 mm ID x 9.9 mm L, Center Positive
• Internal Fuse Specications: 1.5A, 250V, Fast Acting, 5 X 20mm
General
• Ambient Temperature Range: 15-35˚C at rated accuracy, 10-40˚C at
reduced accuracy
• Size: 96mm W X 48mm H X 166mm D (3.8 in x 1.9 in x 6.5 in)
• Mounting: Panel mount into 91mm W X 44mm H (3.6 in x 1.7 in) cutout
• Weight: 1 lb
• Approval: CE, RoHS compliant
Power Supply (#490-060)- supplied with the SCM10
• Input Power Requirements: 90-264VAC, 47-63Hz, 0.25A rms @ 120Vac,
0.15A rms@ 240Vac
• Output Voltage: +12V DC
• Output Power: 12W (1A)
SAFETY SUMMARY
Observe these general safety requirements with the instrument at all times.
Failure to comply with these precautions or with specic warnings elsewhere
in the manual violates safety standards of design, manufacture and intended
instrument use. Scientic Instruments assumes no liability for customer failure
to comply with these requirements.
The SCM10 has been designed and tested to protect the user from electric
shock, burn, mechanical hazards and spread of re from the instrument, under
normal working conditions. The SCM10 is designed for indoor use only. Improper
use of the instrument may pose a hazard to the operator and surrounding area.

Scientic Instruments, Inc. Model SCM10
9
090-267 Rev A
The power supply included with the SCM10 meets or exceeds the European
Union standard, EN-60950-1.
Scientic Instruments assumes no liability for damage caused by controlling
or switching any safety equipment or high integrity equipment with the SCM10’s
relays and alarms.
Do not operate the instrument in an explosive atmosphere. Operation of any
instrument in the presence of ammable gases and fumes that is not rated for
such use constitutes a safety hazard.
Do not install substitute parts in the SCM10 or perform any unauthorized
modication to the instrument. Return the instrument to Scientic Instruments
for service and repair to ensure that safety features are maintained.
RELATED SAFETY SYMBOLS
Direct Current (power line)
ABSOLUTE MAXIMUM RATINGS
The following table lists the absolute maximum ratings of the power input and
other analog input/output connections. This information is useful when selecting
the appropriate wiring conductor for these connections.
Input/Output Pin Maximum Voltage/Current Ratings
Sensor input (I+ to I-, E+ to E-) 6V, 1mA
Analog Output (with respect to COM) 11V, 20mA
Relay Pins (with respect to COM) 24VDC, 1A
Input Power 9-24VDC, 1A

Scientic Instruments, Inc. Model SCM10
10 090-267 Rev A
2. INSTRUMENT SUMMARY
GENERAL
This chapter describes the general installation instructions for the SCM10.
INSPECTION AND UNPACKING
Inspect shipping boxes for external damage before opening. Photograph any
box that has signicant damage before opening. If there is visible damage to
the contents of the container, please contact the shipping company and
Scientic Instruments immediately.
Items included with the SCM10 Temperature Monitor:
1. SCM10 Temperature Monitor
2. SCM10 Users Manual
3. SCM10 Utility Software CD
4. DB15 Mating Connector and Shell
5. DC Adapter
REAR PANEL DEFINITION
This section describes the connections on the rear panel of the SCM10. It
contains references to the sections that describe these connector pin outs in
detail. Figure 2.1 shows the rear panel of the SCM10.
Figure 2.1
Table of contents
Other SCIENTIFIC INSTRUMENTS Measuring Instrument manuals