Model C-212
TRMS AC/DC Clamp Meter
Instruction
Manual

reedinstruments
www com
2
Table of Contents
Safety ............................................................................................ 3
Features......................................................................................... 4
Specications.............................................................................4-6
Instrument Description ...............................................................7-8
Measurement Procedures ........................................................9-12
Battery Replacement................................................................... 13
Product Maintenance .................................................................. 13
For service on this or any other REED product or information on other
REED products, contact REED Instruments at info@reedinstruments.com.

reedinstruments
www com
3
Safety
•
Read this manual before using the meter and follow all safety guidelines.
• Use the meter only as specied in this instruction manual. Otherwise
the meters protection may be impaired.
•
Never use this meter on a circuit with voltages greater than 750Vrms @ 50/60 Hz.
• Do not operate the meter if the body or test leads look damaged.
• Check the function selector range switch and make sure it is at the cor-
rect position before each measurement is taken.
• Do not perform resistance and continuity tests on a live circuit.
• Exercise extreme caution when measuring live systems with voltages
greater than 60VCC or 30VCA.
• Use extreme care when working around bus bars and bare conductors.
• Do not use the meter in over range/overload conditions (OL).
•
Change the battery when the symbol appears to avoid incorrect readings.
Environmental Conditions
Altitude up to 2000 meters.
Operating Temperature: 0°C ~ 40°C, <80% RH, non-condensing
Storage Temperature: -10°C ~ 60°C, <70% RH, battery removed
Pollution Degree: 2
Installation Categories III (600V), II (1000V)
Symbols
Attention! Refer to operation Instructions.
Dangerous voltage may be present at terminals.
This instrument has double insulation.
Approvals: EN61010 600V CAT III 1000V CAT II
When servicing this meter, use only specied replacement parts. For service
(repairs or calibration) on this or any other REED product or information on other
REED products, contact REED Instruments at info@reedinstruments.com.

reedinstruments
www com
4
Features
• True RMS measurements
• 3-3/4 digit LCD with high-speed, 42-segment bargraph
• Max/Min and Data Hold functions
• Peak Hold function
• Frequency measurement
• Diode and continuity test with beeper
• Polarity and overload indications
• Auto or manual ranging
• Auto power-off
• Low battery indication
Specifications
General Specification:
Digital Display: 3 3/4 digits LCD display with maximum reading 3999
Analog Display: Fast 40 segment analog bar graph display
Symbol & Scale Range:
Automatic display according symbols
and range input signal
Polarity:
is displayed when negative signal applied to the input
Over Load: is displayed when input signal exceeds range
Sample Rate: 2 times/sec for digital data; 20 times/sec for analog bar
Low Battery Indication: is
displayed when the battery
is below the required voltage
Auto Power Off: By default, the meter is in auto power off mode. The
meter will power itself off after 30 minutes without
operation. To disable the auto power off, press and
hold the “MAX/MIN” button and turn the meter on.
Jaw Opening: 40mm
Power Supply: 9V, NEDA1604 or 6F22 or 006P
Battery Life: 40 hours typically (Alkaline)
Dimensions (L x W x H):
9.53 x 2.60 x1.4” (242 x 66 x 36 mm)
Weight: 400g (14.10 oz) with battery
Includes: Carrying case, test leads, 9V battery
and instruction manual
continued ...

reedinstruments
www com
5
Electrical Specifications
The accuracy specication is dened as ± (…% of reading+ …count) at
23±5°C,
=
<
80%RH
TRMS for ACV and ACA accuracy are specied from 5% to 100% of the
range. For the accuracy add ±(1% of reading) on Crest Factor 1.4<CF<3
at full scale & CF<6 at half scale.
ACA
Range Resolution Accuracy Overload Protection
40A 0.01A
50 ~ 60Hz
± (1.9% rdg +
5 digits)
60 ~ 500Hz
±(2.5% rdg +
5 digits )
1200Arms400A 0.1A
1000A 1A
DCA
Range Resolution Accuracy Overload Protection
40A 0.01A
± (2.5% reading + 10 digits) 1200Arms400A 0.1A
1000A 1A
ACV (Autorange)
Range Resolution Accuracy Overload Protection
400V 0.1V ± (1.5% reading + 5 digits)
50 ~ 500Hz 1000Vrms
750V 1V
DCV (Autorange)
Range Resolution Accuracy Overload Protection
400V 0.1V ± (1% reading + 2 digits) 1000Vrms
1000V 1V
Ohm (Resistance) (Autorange)
Range Resolution Accuracy Overload Protection
400Ω 0.1Ω ± (1% reading + 2 digits) 600Vrms
4000Ω 1Ω
continued ...

reedinstruments
www com
6
(Diode)
Range Test Current Open Circuit Voltage Overload Protection
1.7mA Max. 6V Max. 600Vrms
Continuity
Range Active Region Overload Protection
<100Ω 600Vrms
Frequency (Hz) (Autorange)
Function Range Resolution Sensitivity Accuracy
A-Hz 4KHz 1Hz 2Arms
±(0.1% reading +
1 digits) (20~10KHz)
10KHz 10Hz 5Arms
V-Hz 4KHz 1Hz 5Vrms
±(0.1% reading +
1 digits) (10~10KHz)
10KHz 10Hz 10Vrms

reedinstruments
www com
7
Instrument Description
1 — Current Sensing Clamp
2 — Clamp opening handle
3 — Function select dial
4 — Hold Button
5 — Peak Hold Button
6 — Maximum Minimum Button
7 — AC/DC Select Button
8 — Relative Button
9 — LCD Display
10 — “COM” Terminal
11 — “ + ” Test Terminal
Display
Auto Power Off indication
Low battery indication
Direct Source indication
Alternative
Positive Peak detect indication
Relative indication
Negative Peak detect indication
Manual range indication
Data Hold indication
Diode function indication
Continuity with beeper
Polarity indication
Analog Bar display
Resistance measure-
ment indication
Voltage & Current
measurement
indication
Frequency measure-
ment indication
continued ...

reedinstruments
www com
8
Data Hold Function
Press the “HOLD” button at any time to hold the current reading. To
release the data being held, simply press the “HOLD” button again.
Peak Hold Function
This meter is designed with 1ms peak hold function at ACA, ACV, DCA
and DCV ranges. Before performing a peak hold operation, a calibration
process must be completed rst.
To do this, press and hold the peak button for 2 seconds. “CAL” will dis-
play on the LCD and the offset will be calculated and kept in the meter.
After the calibration is complete, you can choose P+ or P- by pressing
the “PEAK + HOLD” button to keep the peak reading. To return the meter
back to normal operation, press the “PEAK” button for 2 seconds.
When the function range changes, the meter will require another calibra-
tion for peak measurement.
MAX/MIN Function
This meter is designed with a MAX/MIN function at ACA, ACV, DCA and
DCV ranges.
Press the “MAX/MIN” button once to set the meter to the MAX mode.
Press it once again to set the meter to MIN mode. Press it once again and
the meter will display the present reading and will continue tracking the
MAX and MIN changes. This mode will also be indicated with the blinking
“MAX MIN” symbol.
Set the meter back to normal operation by pressing the “MAX/MIN”
button for more than 2 seconds.
REL Function:
This meter is designed with REL function at ACA, ACV and DCV, DCA ranges.
Press the “REL” button to change the zero to the current reading and the
relative value will show on the LCD. Press it once again and the meter
will display the relative zero point. This mode will also be indicated with a
blinking “REL” symbol.
Set the meter back to normal operation by pressing the “REL” button for
more than 2 seconds.

reedinstruments
www com
9
Measurement Procedures
Current Measurement
Switch the main function selector to proper current range.
ACA
Open the clamp by pressing the jaw-opening handle and insert the cable
to be measured into the jaw. Choose AC current measurement by press-
ing the “AC/DC” button for ACA readings.
DCA
To measure DC current, choose DC measurement by pressing the
“AC/DC” button. Wait for the reading to stabilize then press the “REL”
button to reset the reading. Once completed, insert the cable to be
measured into the jaw and read the value.
Notes: Before taking this measurement, disconnect the test leads from the
meter for safety. If the reading is difcult to read, push the “HOLD” button
and read the result later.
CORRECT
INCORRECT
(2 wires inside)
continued ...

reedinstruments
www com
10
Voltage measurement
Switch the main function selector to Voltage Range. Connect the Red test
lead to the “+” terminal and the Black one to the “COM” terminal. Choose
AC or DC by pressing the AC/DC button. Measure the voltage by touching
the test lead tips to the test circuit where the value of voltage is needed.
Read the result on the LCD panel.
Resistance measurement
Switch the main function to the “Ω” range,
verify that the power line is disconnected from
the system. Connect the Red test lead to the
“+” terminal and the Black one to the “COM”
terminal. Connect the tips of the test leads to
the points where the value of the resistance is
needed. Read the result on the LCD panel.
Note: When measuring the resistance value of
a circuit system, make sure the power is off
and all capacitors required are discharged.
continued ...