Protimeter BaleMaster GRN6165 Page 3
3 Moisture Measurement inBales
Measure moisture in bales as follows:
Note: Before using the stainless steel probe, be sure it is thoroughly clean
and dry. The probe must not be wet when measuring moisture.
a.Connect the Bale Probe to the instrument and switch ON by pressing and
holding button.
b.Push the Bale Probe in to the bale to the required depth. Note the reading
on thedisplay.
Note: When assessing the moisture content (mc) of whole bales, it is
recommended that the average value is obtained from numerous readings
taken from different sides and at different depths.
c.To obtain the correct average value, use the following moisture
measurement ranges as aguideline:
- Less than 15% mc: bales are in a safe condition for storage.
- Greater than 15% mc: bales are not in a safe condition for storage.
Additional drying is recommended.
d.After use, switch the instrument OFF by pressing and holding the
button, or leave the instrument to switch OFFautomatically.
4 CalibrationCheck
Calibration can be checked using the Protimeter Calcheck device (Part #
BLD5086) and a 2-pin moisture probe (Part # BLD5079), connected via the
Bale probe socket. Correctly calibrated BaleMaster instruments register a
value in the range 14.0 – 14.4 at 20°C across the Calcheck.
5 Care andMaintenance
When the BaleMaster is not in use, keep it in its pouch together with its
accessories. Clean and dry the probe, and wrap it in the bubble wrap. Then
store the probe for later use. Replace the batteries when the low battery power
symbol appears on the display, or remove the batteries if the instrument is not
going to be used for extended periods.