Nidec PG-200 User manual

DIGITALPRESSUREGAUGE
INSTRUCTION MANUALVer.7.1
PG-200
1
Nishi-ShinjukuPrimeSquare.,7-5-25
Nishi-ShinjukuShinjuku-kuTokyo160-0023,Japan
https://www.nidec-copal-electronics.com
Formoredetailedinformationpleaseaskforthe
nearestdistributororthefollowingsalescenter.
YouaregreatlyappreciatedforpurchasingCopal'sModel
PG−200 digital Pressure Gauge. You should read this
instructionmanualthoroughlytooperatethePG−200in
optimumstate.
CAUTION Thiscautionmarkdescribeswhenthereisapossibilitythatusermaysufferfromdamageorphysical
damagemayoccuriftheproductisusedimproperly.
CAUTION
⑴Theproduct,whichisaprecisioninstrument,mustbetakenspecialcarenottobedamagedbyimpactnorbybeingdroppedwhenitistrasportedandinstorage.
⑵Theproductmustbeavoidedinstoragewhereisdusty,drippingandvibrated.
⑴Mountingtheattachmentforpaneltype
Iffightenigtheattachmenttothepressureport,thetighteningtorqueshouldbearound1N・m.
⑵Pipingforstand-alonetype
Forpiping,usethehexagonalportionofthepressureportfordriving.
⑶Mediatobeused
ThePressureGaugeshouldbeusedforfluidsthatcannotcorrodethediaphragmofSUS316L,pressureportofSUS316andO-ringoffluororubber.
⑷Excessivepressure
The pressure to be measured is within the specified range. Measurement of a higher pressure than the maximum allowable one may cause performance
degradationordestruction.
⑸Internalbattery
①ReplacebatterywithToshibaBatteryCo.,Ltd./ERVM3.6V/lithiumbatteryonly.Useofanotherbatterymaypresentariskoffireorexplosion.Ifyouneed
replacebattery,pleaseplaceanorderwithus.
②When“LOBAT”signisdisplayed,thebatteryshouldbereplaced.
③Thebatteryusedinthisdevicemaypresentriskoffireofchemicalburnifmistreated.Donotrecharge,disassemble,heatabove100℃(212°F),orincinerate.
④Disposeofusedbatterypromptly.Keepawayfromchildren.Donotdisassembleanddonotdisposeofinfire.
⑹Maintenance
IfthePressureGaugegetsdirty,wipeitoffwithhardlysqueezedclothcontaininganeutraldetergent.Donotuseanythinnerandbenzine.
⑺Protectionagainstnoises
①NoiseshavingmixedintheswitchoutputorpowerlinemaycausethePressureGaugetochangethepressureindication,malfunctionorbebroken.
TakesomemeasuretoputthePressureGaugeawayfrompowerlineoruseshieldedwire.ItiseffectivetogroundthePressureGauge.
②Ifthe switchoutputhasaninductionloadsuch asrelayorsolenoidconnectedtoit, itshouldhaveasurgeabsorbercircuit putinit.Therelaycontacts
shouldhaveacontactpreventivecircuitputinittopreventnoisesfrombeinggenerated.
⑻Prohibitionofshort-circuitingtheswitchterminals
Donotshort-circuittheswitchoutputterminaltoanypowerterminal.Theinternalcircuitmaybebroken.
§
1. HandlingNote
§
2. TRANSPORTATION&STORAGE

2
⑴Model
⑵Measurement………………Gaugepressure.(102AP/102AH:absolute)
⑶Indication…………………3-1/2digits,000〜1999Max,digitalLCDdisplay.
⑷Indication…………………Around3times/sec.
⑸Accuracy …………………±0.5%F.S.±2digit(at25℃±5℃)
⑹Temperaturecharacteristics
atzeropoint;±0.05%F.S./℃±2digit(102G/102GP/102VP/102VH/103G/103GP/103GMP)
±0.1%F.S./℃ ±2digit(101G/101GP/102AP/102AH)
inSPAN ;±0.05%Reading/℃ ±2digit
⑺Switchingfeatures
Numberofsetpoints ……Two,HIandLO,activatidatupperlimit.
Setrangedivisions………000〜1000,HIandLO.
Settingmethod……………WithtrimmereachforHIandLO.Setvalueindicationcanbemadetotimewithsettingswitch.
Operatingaccuracy ………Within±3divisionsfromindicationvalue.
Difference…………………Within4divisions,fixed.
Outputsystem……………NPNopencollectorof30Vdcand40mAatmax.
Operatingindication ……“H”inHImodeor“L”inLOmodeatONoutput.
⑻Operatingtemperature……0〜50℃
humidity…35〜85%RH,nosweatingallowed.
⑼Storagetemperature ……−20〜70℃,withlowerhumiditythan65%RH.
⑽Netweight…………………Netweight,paneltype:Around130g.
stand-alonetype:Around330g.
⑾Pressureport
……………Pressureport,paneltype:M50.8pitch.
stand-alonetype:R1/4(PT1/4).
⑿Attachments………………Halfunionfor6-mmODtube.
Valuejointfor6-mmOD,4-mmIDtube,paneltype.
⒀Options(seeTable2)……
a.Internalbattery
Type…Lithiumbattery,ER6VM.
Serviceabletime…Oneyear(9000hours),continuouslyoperable.
HasabatterychangeindicatorBATT.
b.Externalpowersource
Inputvoltage…5〜10or10〜24Vdc(seeTable2).
Currentconsumption…Lessthan20mA.
c.Analogvoltageoutput
Outputvoltage…0〜1V,notisolated.
Accuracy…±5%F.S.,whichisaddedtomainaccuracy.
Loadresistance…Higherthan1kohm.
Seriesname Series(Table1)
Type(Table2)
PG−200− −
§
3. Specification
*Measurementcanbemadeupto“−760mmHg”for102VHandupto“760mmHg”for102AH
(Accuracynotguaranteedbeyondtheratedpressure.).
101G
101GP
102G
102GP
102VP
102VH
102AP
102AH
103G
103GP
103GMP
mmH2O
kPa
kgf/cm2
kPa
kPa
mmHg
kPaabs
mmHgabs
kgf/cm2
kPa
MPa
1
0.01
0.001
0.1
0.1
1
0.1
1
0.01
1
0.001
5000
50
5
500
500
3678
500
3678
20
2000
2.0
5000
20
2
200
200
1471
200
1471
15
1500
1.5
0〜1000
0〜10.00
0〜1.000
0〜100.0
0〜−100.0
0〜−735
0〜100.0
0〜735
0〜10.00
0〜1000
0〜1.000
*
*
*See§5TerminalWiring
PRESSURERANGE
MAX
PRESSURE
Break-down
Pressure
RESOLUTION
UNITSERIES
Table1
*
Panel
Stand
−P
−1−P
−3−P
−S
−1−S
−3−S
HI、LO
HI、LO
HI
HI、LO
HI、LO
HI、LO
−
−
Voltage
−
−
Voltage
Internallithiumbattery
External5〜24Vdc
Sameasabove
Internallithiumbattery
External5〜24Vdc
Sameasabove
Table2
TYPE POWERSOURCE
SWITCH
OUTPUT
OPTIONAL
OUTPUT
*

3
Switchoutputofnpnopencollectortypewith30Vdcand40mAatmaximum.
§
6. OutputType
[Stand-alonetype]
[Paneltype] [Stand-alonetype]
ProceedonthePCboadinsidethemainbody.
The external power models have
powervoltageof10-24Vdcsetat
default.TooperateyourPressure
Gauge with 5-10 Vdc, change the
internalterminalblockwiring.
Select one of the terminal blocks for connection depending on the power
voltage.Donotapplyanyvoltageexceedingthevoltage range.Foroutput
wiring,usethinnerwirethanAWG26.
Theeachcommonofexternalpowerterminalandvoltageoutputterminal
areconnectedtotheinternalcircuit.
[Paneltype]
⑴Mountingontopanel
Cutaholeoutasshowninthepanelcutdrawingabove.Forciblyputthe
mainbodyintothecuthole.Thehookwillfixthemainbodytothepanel.
⑵Removingoutofpanel
Drawthemainbodyoutofthepanelwhilenippingthehookbetweenyour
thumbandmiddlefingers.
§
4. ExterminalDimensions tolerance(±0.5mm)
§
5. TerminalWiring
−
72
40
69.5 +0.5
0
68.5 +0.5
0
Unitofdisplay
Pressureport
M5,depth5
Panelcut
Panelthickness
(1〜2.5mm)
Panelmontnghook
13
72
72
40
13.5
R1/4
(PT1/4)
M5(P=0.8)depth6
Unitofdisplay
11.5
7222
13.5
28.5
Terminalblockof6pins
Atdefault
Black Red
Battery
(−P)
Exterminalpower
(−3−P)
1
SwitchoutputHI
Voltageoutput
SwitchoutputLO
SwitchoutputHI
Powersource
23456
Exterminalpower
(−1−P)
SwitchoutputHI SwitchoutputLO
Powersource
・COM.
・SW.OUTHI
・COM.
・SW.OUTLO
・5〜10V
・COM.
・10〜24V
・COM.
・Outputvoltage
Externalpowermodels
}
+
−
−
−
−
+
+
+−+−
+−+−+−
+−+−+−
Powersource(+)
Power(−)source
Transistorprotectivediode
BC
E−
+
Exampleofrelay
connection
5〜10V 10〜24V
−
COM.

4
⑴MakewiringbyfollowingSection5,“TerminalWiring.”
⑵TurnthePOWERswitchon.Fortheexternalpowertype,supplytoitthepowerindicatedonthenameplate.
⑶Checktoinsurethatthepressureindicationiszerowithnopressureapplied.Thezeroindicationmaybechangedslightlywithpostureofthemainbody.Ifthe
indicationisover5counts,makezeroadjustmentwiththeZEROadjustingtrimmer.
⑷MakeHIandLOsettingswiththeSETswitchandHIandLOsettingtrimmers(seeSection7,step5).
⑸ApplyapressuretothePressureGauge.Thenitwillindicatethemeasuredvalue.
⑹Iftheswitchoutputison,thePressureGaugewillshow“L”or“H”.
CopalElectronicswarrantstheproductfortheperiodofoneyearafterthedateofthecustomerʼsreceipt.Wewillrepairthetroubledproductscausedbyour
improperdesigningand/orproductioncontrolatourcost.Ourwarrantyislimitedtotheproductsonly,notonanotherdamagethatiscausedbytheproduct
malfunction.
Thefollowingdamage/replacementarerepaired/takenatthecustomerʼscharge;
⑴Failureordamagecausedbyunsuitableorwrongusage,modificationorrepairbythecustomer.
⑵FailureordamagecausedbydisasteroranActofGod.
⑶Replacementofattachments(HalfUnionandBarbJoint).
§
8. Howtouse
§
9. WARRANTY
⑴Frontpanelcontrols
⑵Descriptionofthefeatures
1.
2.
3.
4.
5.
6.
7.
8.
9.
Pressureindicator
Polarityindicator
Pressureport
POWERswitch
SETswitch
HIsettingtrimmer
LOsettingtrimmer
SPANadjustingtrimmer
ZEROadjustingtrimmer
Showsapressurevalueenteredtothepressureport.Alsoshowsasetvaluegivenwiththesetswitch.
Showsthenegative(−)uponnegativepressure.
Hasapressureapplied.
TurnsonoroffthepowertothePressureGauge.
SelectsamodetocheckaHIorLOsetvalue.IfyoupresstheSETswitchonce,theindicationwillbeautomaticallychangedinthe
sequenceshownbelow.TochecktheLOsetvalue,waitforaround20secondsafterpressingtheSETswitch.
Ifsettingendsortointerruptsettinginthecourse,presstheSETswitchagain.Thenthepressurewillbeindicated.
UsedtoadjustpressuretotheHIsetvaluethatisindicatedwiththeSETswitch.
UsedtoadjustpressuretotheLOsetvaluethatisindicatedaftertheHIsettingtimehaselapsed.
Usedtoadjustthesensitivitysothatthepressurecanindicatethefullscalewiththeratedpressureappliedtothepressureport.
Asthetrimmeriscalibrated,itissealedup.
Usedtoadjustthesensortozero.Thepressureindicationshouldbeadjustedto“000”withnopressureapplied.
§
7. FrontPanelControlsandFeatures
Frontcontrols
Pressureindicator
Batterychange
indicator
Polarityindicator
Switchingindicator
Decimalpoint
SETswitch LOsetting
trimmer
HIsetting
trimmer
POWERswitch SPANadjusting
trimmer
ZEROadjusting
trimmer
Indicationmode
Settingmode
Around20sec Around20sec
NO
YES
SET
Pressure
indicaton HIsetting LOsetting
Other Nidec Measuring Instrument manuals

Nidec
Nidec Kato Engineering KRM2000 User manual

Nidec
Nidec DT-200LR-S12 User manual

Nidec
Nidec TNP Series User manual

Nidec
Nidec FG-7000T User manual

Nidec
Nidec FG-7000L User manual

Nidec
Nidec SHIMPO DT-5TXR User manual

Nidec
Nidec FG-3000R-S User manual

Nidec
Nidec ST-1100 User manual

Nidec
Nidec FG-3000R Series User manual

Nidec
Nidec DT-361 User manual

Nidec
Nidec Shimpo MT-200 User manual

Nidec
Nidec ST-1100 User manual

Nidec
Nidec TNX Series User manual

Nidec
Nidec FGE-HXY Installation manual

Nidec
Nidec DT-361 User manual

Nidec
Nidec FG -3000 Series User manual

Nidec
Nidec MFD-02 User manual

Nidec
Nidec ST-1100 User manual

Nidec
Nidec PG-35 User manual

Nidec
Nidec DT-501X User manual