MSA Ultima XI User manual

Ultima®XI Infrared
Gas Monitor
Instruction Manual
THIS MANUAL MUST BE CAREFULLY READ BY ALL INDIVIDUALS WHO HAVE OR WILL
HAVE THE RESPONSIBILITY FOR USING OR SERVICING THE PRODUCT. Like any piece
of complex equipment, this device will perform as designed only if it is used and serviced
in accordance with the manufacturer’s instructions. OTHERWISE, IT COULD FAIL TO
PERFORM AS DESIGNED AND PERSONS WHO RELY ON THIS PRODUCT FOR THEIR
SAFETY COULD SUSTAIN SEVERE PERSONAL INJURY OR LOSS OF LIFE.
The warranties made by Mine Safety Appliances Company with respect to the product are
voided if the product is not used and serviced in accordance with the instructions in this
manual. Please protect yourself and others by following them. We encourage our
customers to write or call regarding this equipment prior to use or for any additional
information relative to use or service.
In the U.S., to contact your nearest stocking location, dial toll-free 1-800-MSA-INST
To contact MSA International, dial 1-412-967-3354.
© MINE SAFETY APPLIANCES COMPANY 2009 - All Rights Reserved
This manual is available on the internet at www.msanet.com.
Manufactured by
MSA NORTH AMERICA
P.O. Box 427, Pittsburgh, Pennsylvania 15230
(L)-Y Rev 3 10052972
"! WARNING

i
MSA Permanent Instrument Warranty
1. Warranty- Seller warrants that this product will be free from
mechanical defect or faulty workmanship for the following period:
• IR Sensor source: ten (10) years from date of shipment
• All other IR components: two (2) years from date of shipment.
This warranty is applicable provided it is maintained and used in
accordance with Seller's instructions and/ or recommendations.
This warranty does not apply to expendable or consumable parts
whose normal life expectancy is less than one (1) year. The Seller
shall be released from all obligations under this warranty in the
event repairs or modifications are made by persons other than its
own or authorized service personnel or if the warranty claim results
from physical abuse or misuse of the product. No agent, employee
or representative of the Seller has any authority to bind the Seller
to any affirmation, representation or warranty concerning the goods
sold under this contract. Seller makes no warranty concerning
components or accessories not manufactured by the Seller, but will
pass onto the Purchaser all warranties of manufacturers of such
components. THIS WARRANTY IS IN LIEU OF ALL OTHER
WARRANTIES, EXPRESSED, IMPLIED OR STATUTORY, AND IS
STRICTLY LIMITED TO THE TERMS HEREOF. SELLER
SPECIFICALLY DISCLAIMS ANY WARRANTY OF
MERCHANTABILITY OR OF FITNESS FOR A PARTICULAR
PURPOSE.
2. Exclusive Remedy- It is expressly agreed that Purchaser's sole
and exclusive remedy for breach of the above warranty, for any
tortious conduct of Seller, or for any other cause of action, shall be
the repair and/ or replacement at Seller's option, of any equipment
or parts thereof, which after examination by Seller is proven to be
defective. Replacement equipment and/ or parts will be provided at
no cost to Purchaser, F.O.B. Seller's Plant. Failure of Seller to
successfully repair any nonconforming product shall not cause the
remedy established hereby to fail of its essential purpose.
3. Exclusion of Consequential Damage- Purchaser specifically
understands and agrees that under no circumstances will seller be
liable to purchaser for economic, special, incidental or
consequential damages or losses of any kind whatsoever, including
but not limited to, loss of anticipated profits and any other loss
caused by reason of non-operation of the goods. This exclusion is
applicable to claims for breach of warranty, tortious conduct or any
other cause of action against seller.

General Warnings and Cautions
1. The Ultima XI Infrared Gas Monitors described in this manual
must be installed, operated and maintained in strict accordance
with their labels, cautions, warnings,instructions, and within the
limitations stated. Verify that the class, group, and temperature
ratings of the equipment agree with the actual classification of
the location.
2. The Ultima XI Infrared Gas Monitor is designed to detect gases
or vapors in air. It can also measure the concentration of gases or
vapors in steam or inert or oxygen-deficient atmospheres. .
3. Use only genuine MSA replacement parts when performing any
maintenance procedures provided in this manual. Failure to do so
may seriously impair instrument performance. Repair or alteration
of the Ultima XI Infrared Gas Monitor, beyond the scope of these
maintenance instructions or by anyone other than authorized MSA
service personnel, could cause the product to fail to perform as
designed.
4. The Ultima XI Infrared Gas Monitor detects the presence of most
combustible gases by identifying the difference in the amount of
infrared light energy absorbed during the presence of these gases.
This monitor, however, does NOT detect the presence of hydrogen
gas and must never be used to monitor for hydrogen gas.
5. The standard Ultima XI Infrared Gas Monitor does not detect the
presence of acetylene gas. The presence of acetylene gas will
desensitize sensor performance. Custom-built acetylene sensors
are available through your MSA representative.
6. CSA performance Certification to standard C22.2 No. 152 is valid
only when the instrument is calibrated on methane per the
instruction manual.
"WARNING
ii

1. As with all gas monitors of this type, high levels of, or long
exposure to, certain compounds in the tested atmosphere could
contaminate the sensors. In atmospheres where an Ultima XI
Infrared Gas Monitor may be exposed to such materials, calibration
must be performed frequently to ensure that operation is
dependable and display indications are accurate.
2. The Ultima XI Infrared Gas Monitor must not be painted. If painting
is done in an area where a Monitor is located, care must be
exercised to ensure that paint is not deposited on the
environmental shield of the Ultima XI Infrared Gas Monitor, if so
equipped. Such paint deposits would interfere with the diffusion
process, whereby a sample of the atmosphere being monitored
diffuses into the Monitor.
3. The only absolute method to ensure proper operation of an Ultima
XI Infrared Monitor is to check it with a known concentration of the
gas for which it has been calibrated. Consequently, calibration
checks must be included as part of the routine inspection of the
system.
4. Protect the Ultima XI Infrared Gas Monitor from extreme vibration.
FAILURE TO FOLLOW THE ABOVE CAN RESULT IN PRODUCT
DAMAGE AND/OR AN UNSAFE CONDITION.
"CAUTION
iii

Table of Contents
Chapter 1,
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-1
General Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-1
Identifying Your Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-1
Installing Your Gas Monitor . . . . . . . . . . . . . . . . . . . . . . . . . . .1-2
Ultima XI Electrical Connections . . . . . . . . . . . . . . . . . . . . . . .1-4
For Milliamp Output . . . . . . . . . . . . . . . . . . . . . . . . . . .1-5
Chapter 2,
Start-up and Calibration . . . . . . . . . . . . . . . . . . .2-1
Calibration Basics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-1
Ultima XI Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-2
Calibration Kit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-5
Chapter 3,
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-1
Chapter 4,
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-1
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-1
Ultima XI Cleaning Procedure . . . . . . . . . . . . . . . . . . . . . . . . .4-1
Appendix A,
Calibration Guide
for Additional XI Gases . . . . . . . . . . . . . . . . . . . .A-1
iv

List of Figures
Figure 1-1. Explosion-Proof Ultima XI Infrared Gas Monitor .1-1
Figure 1-2. Explosion-Proof Ultima XI Infrared Gas Monitor
with Junction Box . . . . . . . . . . . . . . . . . . . . . . . . .1-2
Figure 1-3. Ultima XI Infrared Gas Monitor . . . . . . . . . . . . . . .1-3
Figure 1-4. Ultima XI Installation Requirements . . . . . . . . . . .1-4
Figure 1-5. Ultima XI Infrared Gas Monitor
with Junction Box . . . . . . . . . . . . . . . . . . . . . . . . .1-5
Figure 2-1. Intrinsically Safe Ultima XI Calibration Cap . . . . .2-3
Figure 2-2. Calibration Sequence Diagram . . . . . . . . . . . . . . .2-3
Figure 2-3. Environmental Guard . . . . . . . . . . . . . . . . . . . . . .2-4
List of Tables
Table 1-1. Ultima XI Maximum Cable Length
& 4-20 mA Signal Load . . . . . . . . . . . . . . . . . . . .1-5
Table 2-1. Instrument Operation . . . . . . . . . . . . . . . . . . . . . . .2-1
Table 2-2. Calibration Kit Table . . . . . . . . . . . . . . . . . . . . . . .2-5
Table 3-1 . Performance Specifications . . . . . . . . . . . . . . . . .3-1
Table 3-2. Certification and Approval . . . . . . . . . . . . . . . . . . .3-2
Table 3-3. Installation Drawings . . . . . . . . . . . . . . . . . . . . . . .3-2
Table 4-1. Parts List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-3
v

1-1
Chapter 1,
Installation
General Description
The Ultima XI Infrared Gas Monitor is designed to sample the
environment where mounted and alert you to potentially dangerous
levels of combustible gas. The Ultima XI Infrared Gas Monitor is
shipped factory-calibrated and is labeled with target gas, calibration gas
and span setting information. The target gas is not field-configurable
and cannot be changed. The Ultima XI Gas Monitor is a stand-alone
unit with a 4 to 20 mA output.
Identifying Your Unit
• The Ultima XI Infrared Gas Monitor is housed in a 316 stainless
steel explosion-proof enclosure (FIGURE 1-1).
Do not mix units with different area classifications. All units
Figure 1-1. Explosion-Proof Ultima XI Infrared Gas Monitor

• The Ultima XI Infrared Gas Monitor is also available with an
optional junction box equipped with a terminal strip for easy wiring
(FIGURE 1-2).
Do not mix units with different area classifications. All units
Installing Your Gas Monitor
Generally, the Ultima XI Gas Monitor should be mounted close to the
area where a leak is likely to occur or where the gas is expected.
Testing for ventilation patterns may be helpful in establishing locations
for the Ultima XI Gas Monitor in enclosed areas. Smoke tubes
(P/N 458481) are useful in measuring these patterns to help determine
where gases may flow or accumulate.
Mount the Ultima XI Gas Monitor on a support by using suitable
hardware (not supplied), and attach rigid conduit and a conduit seal
fitting (for XP installations). The optional condulet assembly can also be
used for wiring power and signal.
NOTE: The XI Gas sensor must be mounted with the sensor
extended horizontally as shown in FIGURE 1-2.
The Ultima XI Infrared Gas Monitor contains no user- or field-
serviceable parts and must be returned to the factory for
repair. Any attempt to open the monitor will damage the unit
and void the warranty.
Mount the Ultima XI Infrared Gas Monitor with the sensor
inlet fitting extended horizontally from the main enclosure
(FIGURES 1-1 and 1-2). Horizontal mounting will help pre-
vent the build-up of particulate or liquid matter on the moni-
tor's optical surfaces.
"CAUTION
"WARNING
Figure 1-2. Explosion-Proof Ultima XI Infrared Gas Monitor
with Junction Box
1-2

Do not paint the Ultima XI Infrared Gas Monitor. If painting is
done in an area where a sensor is located, exercise caution
to ensure paint is not deposited on the sensor inlet fitting.
Such paint deposits would interfere with the diffusion
process, whereby a sample of the monitored atmosphere
diffuses into the sensor. In addition, solvents in the paint
may cause an alarm condition to occur.
Protect the Ultima XI Infrared Gas Monitor from extreme
vibration. Do not mount sensing head in direct sunlight as
this may cause the sensor to overheat.
During sensor installation or removal, do not use any tools
or apply excessive force to the two legs that support the
unit's reflectors (FIGURE 1-3). Applying force to the legs can
permanently damage the monitor.
The monitor's environmental guard should be installed on
the unit at all times. If the monitor is to be operated without
the guard, frequent checks should be made to ensure par-
ticulate or liquid matter has not collected on the windows.
Use a wrench on the hex nut in the sensor neck to install or
remove the sensor.
Do not mix units with different area classifications. All units
Figure 1-3. Ultima XI Infrared Gas Monitor
1-3

Ultima XI Electrical Connections
The Ultima XI Gas Monitor is provided with input wires for power and
signal connections. An optional condulet assembly is also available.
An external power supply is required.
• For power requirements, see Chapter 3, "Specifications".
• All connections should be made by following appropriate wire code
procedures.
NOTE: Refer to TABLE 3-3 for appropriate installation drawing.
Before wiring the Ultima XI Infrared Gas Monitor, disconnect
or isolate all power connected to the monitor; otherwise,
electrical shock could occur.
Be sure to install your Ultima XI Infrared Gas Monitor in
accordance with National and local Wiring Regulations.
Failure to do so can result in an unsafe condition.
The Ultima XI Monitor is a three-wire transmitter that operates in the
current source mode and can be wired directly to the input wires
(FIGURE 1-4) or with the optional junction box (FIGURE 1-5). TABLE
1-1 provides typical cable lengths and wire sizes for installation.
"WARNING
Figure 1-4. Ultima XI Installation Requirements
1-4
Table of contents
Other MSA Measuring Instrument manuals

MSA
MSA MiniOx I User manual

MSA
MSA SUPREMA Calibration User manual

MSA
MSA ALTAIR 5X User manual

MSA
MSA Ultima Installation and operating instructions

MSA
MSA Chillgard RT User manual

MSA
MSA PrimaX User manual

MSA
MSA altair 4x User manual

MSA
MSA Chillgard 5000 User manual

MSA
MSA ICM 2000 Plus User manual

MSA
MSA Legend Series User manual

MSA
MSA altair 4 User manual

MSA
MSA altair 4 User manual

MSA
MSA altair 5 Manual

MSA
MSA Mini ID R-1234yf User manual

MSA
MSA ALTAIR 5X User manual

MSA
MSA Ultima X Series Installation and operating instructions

MSA
MSA Chillgard LS User manual

MSA
MSA 487995 User manual

MSA
MSA ULTIMA X5000 User manual

MSA
MSA ULTIMA XL Series User manual