EMC Precautions
The Avant Stealth Audiometer needs special precautions regarding EMC and needs to be
installed and put into service according to the following EMC information.
List of all cables and maximum lengths of cables from transducers and accessories:
Warnings!
•The Avant Stealth Audiometer generates high frequency for its own use.
•The Avant Stealth Audiometer is intended to create a medical system.
•The use of accessories, transducers and cables other than those specified, with the exception of
transducers and cables sold by the manufacturer of the Avant Stealth Audiometer as replacement
parts for internal components, may result in increased emissions or decreased immunity of the
Avant Stealth Audiometer.
•The Avant Stealth Audiometer should not be used adjacent to or stacked with other equipment
and if adjacent or stacked use is necessary, the Avant Stealth Audiometer should be observed to
verify normal operation in the configuration in which it will be used.
•The Avant Stealth Audiometer may be interfered with by other equipment, even if that other
equipment complies with CISPR emission requirements.
•The Avant Stealth Audiometer does not have life supporting function.
•Portable and mobile RF communications equipment can affect the Avant Stealth Audiometer.
Guidance and manufacturer’s declaration – electromagnetic emissions
The Avant Stealth Audiometer is intended for use in electromagnetic environment specific below. The customer or the
user of the Avant Stealth Audiometer should assure that it is used in such an environment.
Electromagnetic environment - guidance
CISPR 11
The Avant Stealth Audiometer uses RF energy
only for its internal function. Therefore, its RF
emissions are very low and are not likely to
cause any interference in nearby electronic
equipment.
CISPR 11
The Avant Stealth Audiometer is suitable for use
in all establishments, including domestic
establishments and those directly connected to
the public low - voltage power supply network
that supplies buildings used for domestic
purposes.
ADP-I-MADPI-1 MedRx AVANT Stealth Installation Manual
Effective Date 06/24/2013 11