Series BG
10
GAUGE CONTROL LANGUAGE
The instrument can be controlled by an external device through the RS-232 channel.
The following is a list of supported commands and their interpretations. All commands
must be terminated with a Carriage Return character (hex 0D) or with a Carriage
Return/Line Feed combination (hex 0D+0A). The gauge responses are always termi-
nated with a Carriage Return/Line Feed.
A Enable Average mode
AD Disable Average mode
AM Select Average mode (if enabled)
ATn Average time. n=0.1-300.0 seconds
DELn Initial delay. n=0.1-300.0 seconds
TRFn Trigger force. n=value (+ for compression, - for tension)
SP EnableSetPoint mode
SPD Disable SetPoint mode
SPHn High set point. n=value (+ for compression, - for tension)
SPLn Low set point. n=value (+ for compression, - for tension)
AOFFn Auto-shutoff. n=0,1,5,10,20,30 minutes. 0=always on
AOUTn Auto-transmit every nth reading. n=0,1,2,4,8,16,32,64,128. 0=disabled
LB Switch units to pounds
KG Switch units to kilograms
N Switch units to Newtons
G Switch units grams
ET Enable External trigger mode
ETD Disable External trigger mode
ETE Edge triggered External trigger mode
ETL Level triggered External trigger mode
HL Reading captured on a high to low transition
LH Reading captured on a low to high transition
CUR Current mode (real time display)
PT PeakTensionmode
PC PeakCompressionmode
CLR Clear peaks, start a new average, or external trigger test
Z Zero display and perform the CLR function
? Request the displayed reading
?C Request the current reading
?PT Request the peak tension reading
?PC Request the peak compression reading
?ET Request the reading obtained during the External trigger mode
?A Request the average reading obtained during the Average mode
FLTCn Digital filter for current (displayed readings). n=1,2,4,8
FLTPn Digital filter for peak readings. n=1,2,4,8
FLTAn Analog filter (2.5 Hz). n=1,2. 1=on, 0=off
Section 8