Mantech Ap2000 User manual

MAN-MD-H-0507-01
Page 1of 24
Instruction Manual
MANTECH Ap2000 Turbidimeter

MAN-MD-H-0507-01
Page 2of 24
General information
In no event will the manufacturer be liable for direct, indirect, special, incidental or
consequential damages resulting from any defect or mission in this manual.
The manufacturer reserves the right to make changes in this manual and the products
it describes at any time, without notice or obligation.
N O T I C E
The manufacturer is not responsible for any damages due to misapplication or misuse
of this product including, without limitation, direct, incidental and consequential
damages, and disclaims such damages to the full extent permitted under applicable law.
The user is solely responsible to identify critical application risks and install appropriate
mechanisms to protect processes during a possible equipment malfunction.
Please read this entire manual before unpacking, setting up or operating this equipment.
Pay attention to all danger and caution statements.
Failure to do so could result in serious injury to the operator or damage to the equipment.
Make sure that the protection provided by this equipment is not impaired. Do not use or
install this equipment in any manner other than specified in this manual.

MAN-MD-H-0507-01
Page 3of 24
1.0 Specifications
Measurement method
Ratio determination using a primary nephelometric light scatter signal
(90°) to the transmitted light scatter signal.
Reading units
NTU/ EBC
Lamp Source
Tungsten Lamp
Detector
Silicon Photocell
Measuring Range
0 to1000 NTU/ 0 to 250 EBC
Resolution
0,01 on lowest range
Accuracy
± 2% of reading: 0 to 500 NTU / ± 3% 500 to 1000 NTU
Repeatability
± 1% of reading or 0,01 NTU
Automatic Reading
With user-defined intervals 0 to 250 seconds
Maximum uncertainty
± 2% of full scale
Display
LCD 2 lines / 16 characters
Response Time
Programmable 6 to 41 sec
Data Logger
Up to 1000 data
Auto ShutOff
Programmable from 1 to 60 min
Fast Cal function
Quick Calibration for single point
Software Functions
Signal averaging, "Fast Settling", results freezing, analyst and sample
identification, calibration status, verification and calibration reminder,
calibration history, password
Sample Required
15 ml vial with lid
Sample vials
Round borosilicate glass vial with screw and caps (Φ = 24,5 mm)
Power Supply
4 AA batteries or power supply cable
Indicator
Low battery indicator / battery exchange
Serial Output
USB
Enclosure rating
IP-67, waterproof
Storage Conditions
0 to 40°C ( instrument only)
Dimensions
114 x 198 x 83 mm
Parts and Accessories - Included.
3 vials of 24,5 mm
Batteries AA e power supply cable
Instruction Manual and Carrying Case
Parts and Accessories - Optional.
Backlight Display
Silicone oil
Cleaning Cloth

MAN-MD-H-0507-01
Page 4of 24
2.0 Installation
2.0 Install the battery
1.
With an Philips screw driver, remove the battery cover.
2.
Install 4 AA alkaline or 4 AA nickel metal hydride (NiMH) batteries. Make sure that the
batteries are installed in the correct orientation.
3.
Replace the battery cover. (for optimal seal we recommend using torque screw driver with
4N.m)

MAN-MD-H-0507-01
Page 5of 24
3.0 User interface and navigation
3.1 User interface
1.DISPLAY: Displays readings, diagnostics
and operational data.
2.MENU: selects options to configure the
instrument, selects analysis and moves the cursor
to the right.
3.SAVE: Store Selections and data, saves the
result to be USB transfered and selects the
parameters.
4.ESC: Powers off the unit, aborts operations,
return to the previous screen.
5.ON/Off-Read: switches the instrument on,
confirms options, sample reading, move the
cursor to the left.
6. ▲: scroll through menus, enter numbers and
letters
7. ▼:scroll through menus, enter numbers and
letters
4.0 Operation
4.1 Startup
Push the ON/OFF key to turn the meter on of off.
If the meter does not turn on, make sure that the batteries, or that the AC power supply is properly

MAN-MD-H-0507-01
Page 6of 24
connected to an electrical outlet.
The display will show, instrument version, data log and time/date.
Note: The Auto-Shutoff option can also be used to turn off the meter.
4.2 Navigation
The instrument Menu structure is divided in 3 main functions with submenus :
-ID –Access the user ID function
-Calibrate –Access the calibration functions
-Config –Access the confifuration functions
-Serviçe- Access the Service Functions (only for certified technicians)
Please see Menu structure in the last page of this manual.
Press and hold the MENU key for 3 seconds to enter the Menu Function.
You will see ID –Identification; Use the ▲ or ▼ keys select the desired function, then press READ to
enter it.
If the menu key is pressed and released quickly the Analysis type selection will appear, select the
type of measurement using ▲ or ▼.
-0-250 EBC (special measurement for food industry)
-0-1000 NTU.
-0.0-.99.9 NTU
-0.00 to 40.0: Rapid Cal Function.
-Auto NTU: Instrument choose automatcally best measuring range in NTU. RECOMMENDED.

MAN-MD-H-0507-01
Page 7of 24
-Sensor: Measurement of the light frequency. (to be used only by certified technician)
-User1: User calibrated courve (require full calibration of this courve).
-User2: User calibrated courve (require full calibration of this courve).
Press and release save to use the desired measuring range/function.
NOTE: The 0.00 to 40.0 function can only be selected for rapid single point calibration (Rapid
cal function).

MAN-MD-H-0507-01
Page 8of 24
4.3 Sample Reading.
Before you read a sample its possible to ID the user and the Sample that will be read, If you don´t
know what to please go to 5.4.2 Taking a Measurement.
4.3.1 ID
Select the Sample ID.
Use ▲ and ▼ to choose an sample name from the list, press and hold SAVE for 3 seconds to
store the data.
4.3.1
USer
Select the User for the sample.
Use ▲ and ▼ to choose an user from the list, press and hold SAVE for 3 seconds to store the
data.
4.3.2
Taking a Measurement.
For greater precision we recommend that you calibrate the 0.10 NTU point before you start
your measurements. ( Refer to CALIBRATION for more information)
Rinse the vial with the sample;
Place sample in the vial;
Dry the tube with a thin/absorbent paper;

MAN-MD-H-0507-01
Page 9of 24
Holding the vial by the cap, gently stir the sample by inversion 3 times.
PLACE THE VIAL IN THE READING CELL.
Align the "l" mark of the sample vial forward with the arrow on the sample cup of
the instrument and close the instrument lid;
-If using Auto mode just wait for the reading to stabilize.
-If using manual mode press READ;
To turn the instrument off press and hold ESC until the display reads "Off".

MAN-MD-H-0507-01
Page 10 of 24
4.4 Sample Vials.
Before being use the vials should always be washed with with detergent to remove any dirt
or fingerprints.
Dry the outside of the tubes with a clean, lint-free cloth. Let the tubes dry in the air, in an
inverted position to prevent dust from entering the vials.
The handling of the vials has great importance. Scratch, fingerprints and water droplets in the vial or
inside the light chamber can cause interference, leading to inaccurate results.
It is very important that vials and light chamber are clean and dry constantly.
The variability in the glass geometry is the predominant cause of the variability of the results. The
position variation of the vial in the light chamber will greatly affect the results. To obtain accurate results,
the vials must be positioned exactly,
The "I" mark in the vial must be aligned with the arrow (▲) in the instrument sample chamber.
Sample Vials calibration
The turbidimeter AP 2000 vials are factory selected for optimal precision, however in low turbidity
conditions, small changes in the glass can cause different readings on the same sample.
To measure samples in the lower range, below 1NTU, for greater accuracy, a vial calibration is
recommended.
1.
Fill each vials with dilution
water *.
2.
Complete a reading of all the vials, take
note of each reading.
3.
Mark the vials with the lowest
reading with an "R", Those will
be your reference vials.
This tag should be near the
top, close to the lid, and with
permanent marker.
4.
The difference between the value of the
reference vials "R" with the other vials is the
correction factor for each of the vials.
This value may be used when
comparing the readings with different vials.
* dilution water - prepared according to Standard Methods - Method 2130 (latest edition) with membrane
filtration for retention of particles ≥ 0.1 μm - generating a nominal value of 0.02 or below 0.1 NTU.