Jain Technology XONIC-100P User manual

1
www.jain.co.kr 82-2-856-4114
USER MANUAL
XONIC-100P PORTABLE FLOWMETER

2
Table of Contents
Preview....................................................................................................................................................................4
Introduction
Safety Consideration
Installation Steps
Specifications
Key Functions
Understanding Display
Best Transducer Installation Place
Section 1. GENERAL
1.1SITE NAME .......................................................................................................................................7
1.2 PASSWORD......................................................................................................................................7
1.3 METER TYPE ...................................................................................................................................8
1.4 APPLICATION ..................................................................................................................................9
1.5 DISPLAY ...........................................................................................................................................9
Section 2. CH SELECT ......................................................................................................................................10
Section 3. PIPE data..........................................................................................................................................11
Section 4. LIQUID type......................................................................................................................................13
Section 5. INSTALLATION.................................................................................................................................14
Section 6. OPERATION......................................................................................................................................18
Section 7. FLOW ................................................................................................................................................21
Section 8. IN / OUTPUT .....................................................................................................................................24
Section 9. DATALOGGER..................................................................................................................................26
Section 10. Diagnostics ....................................................................................................................................27
DRAWINGS.........................................................................................................................................................28

3
Delivery Items
Flow Computer
Transducer
Mounting Track
Cable
Easy Track Magnetic Track

Preview
Introduction
Xonic-100P Clamp-On Ultrasonic Flowmeter is fully digital, state-of-art flowmeters using DSP (digital signal
processing) technology to measure time difference of ultrasonic signals. Please read this manual carefully
and make installation. Xonic 100 will keep best accuracy.
Safety Consideration
Installation Steps
Select best point for transducer installation
1. input GENERAL DATA
2. select CHANNEL
3. input PIPE DATA
4. select LIQUID TYPE
5. do INSTALLATION
6. select FLOW UNIT & TOTAL UNIT
7. select DATALOGGER
Specifications
Type
Clamp-On Ultrasonic Flowmeter
Priciple
Transit-Time
Measuring
Pipe Size
15~ 6000 mm
Upstream 10D, Down stream 5D (single path)
Upstream 7D, Down stream 3D (dual path)
Transducer -40 ~ +120 ℃
(Flow, Total, Velocity, Delta T, Oscilloscope Shape)
Temperature Flow Computer -20 ~ +60 ℃
4
Accuracy ±1.0% (single path),±0.5% (dual path)
Flow Velocity ±0.02 ~ 20.0 m/s
Turn-Down Ratio 1000:1
Repeatability ±0.25%
Required Straight
Run
Xonic-100 use inner batteries and AC 85~264V. 50/60Hz. power adapter. So please consider about safety
and any damage to flow computer during installation. Especially, many application is near moisture place,
so please be careful any electric shock.
Data Input Two 4~20mADC for Temperature , Pressure .etc.
Data OUTPUT
Communication
Two 4~20mADC for flow (load max 750 ohm)
Two Ralay for Total
RS-232C
Power Battery with 85 ~ 264 VAC. 50/60 Hz. power adaptor
Enclosure NEMA 4 (IP67, Rain Proof)
Transducer NEMA 7 (IP68, Water Proof)
Cable Standard 10 m. (Option max 200 m.)
Data Logger 32 Mbytes (above 1,000,000 data)
Display Graphic Color LCD

5
Key Functions
Notice : Touch keys do not have alphabet table, so user must select alphabet by pressing several times
numeric keys.
Understanding Display
Site Name : user must input site name for communication
Type : single ch means single path/channel
Flow Unit : user selected flow unit
Flow : measured flow
Total : total flow
Total Unit : total flow unit
Keys
Functions
Press when start MENU
Accept Key Data
Clear
0~9 numeric
Use when input numbers
.
Use when input “.”
Function Key
Direction keys
Negative input

6
Best Transducer Installation Place
Choose best installation place which has enough straight pipe runs, 10D for upstream and 5D for down stream.
For example, if pipe dameter is 1000mm, then please find 10m straight run for up stream and 5 straight run for
down stream
Find 10D for Upstream and 5D for Downstream

7
Section 1. input GENERAL data
1.1 SITE NAME
User must input SITE MENU for initial setup. SITE NAME can be used as ID of flowmeters when
communicate remotely many flowmeters. Press Cursor move as below.
Move cursor as above and input site name [JAIN________]
Move level from A to C or D and input data
User can input alphabet character by pressing keys and to complete.
Press UP/DN key until cursor show “J”.
Press UP/DN key until cursor show “A”.
Press UP/DN key until cursor show “I”.
Press UP/DN key until cursor show “N”.
The cursor will move to next menu. If user want to finish menu, press .
※User can use alphabet and space characters.
1.2 PASSWORD
User must input password to secure Xonic 100 from any unauthorized man’s operation. Just input
number and alphabet by keypad.

8
1.3 METER TYPE
User can select meter type by selecting as belows,
FLOWMETER : Clamp-On flowmeters for single path, dual path/channel.
VEL NORMAL MODE : Velocity Meter mode for m/s base
VEL TIME MODE : Velocity Meter mode for time base
Single Path ‘V”Method Installation Single Path ‘Z”Method Installation
Dual Path ‘V”Method Installation

9
1.4 APPLICATION
Xonic 100 has single path and dual path/channel mode. User can buy only single path type and
can add one more channel if need.
1.5 DISPLAY
Xonic 100 display only channel 1 or channel 2 or two channel simutanously. User can select by
cursor.
Single Path Display Dual Path Display

10
Section 2. CH SELECT
User can select single channel or dual channel. Single channel means user will use only 1 pair
transducers for one pipe. Dual channel means user will use 2 pair transducers for two pipes. In
case of dual channel, flowmeter measure two pipes simultaneously. Also, user can select single
path or dual path. Single Path means user will use 1 pair transducer for one pipe. Dual Path
means user will use 2 pair transducers for one pipe to increase accuracy.