IKALOGIC SP259 Series User manual

SP259 series datasheet and user manual
SP259 and SP259i 9 channels, 250 MSPS logic analyzer with industrial
inputs option
www.ikalogic.com | support@ikalogic.com
August 31, 2023

SP259 Series User Manual SP259 / SP259i
Contents
SP259 Series overview 4
Embeddedmemory ....................................... 4
Evolution from older SP209 series . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Typicalapplications ....................................... 5
Producthighlights ........................................ 6
Warning 7
SP259 system architecture 8
Main characteristics 10
Operatingconditions....................................... 10
Timingandmeasurements.................................... 10
Logicinputsspecifications.................................... 10
Powerrequirements ....................................... 10
SP259(i) Interfaces 11
Industrial ports (SP259i version only) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Principle of operation 12
Embeddedmemoryvsstreaming ................................ 12
Versatiletriggersystem...................................... 12
Flexitrig®timingperformances.................................. 13
External trigger OUT specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
External trigger IN specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Industrial port (SP259i only) 14
RS485receiverspecifications .................................. 15
RS232receiverspecifications .................................. 15
CANreceiverspecifications.................................... 15
LINreceiverspecifications .................................... 16
What’s in the box 17
Unpacking and first usage 17
StatusLEDsbehavior....................................... 17
SowareQuickStartguide.................................... 17
Capturingyourfirstsignal .................................... 18
(c) Ikalogic SAS 2023 2/26

SP259 Series User Manual SP259 / SP259i
Mechanical data 18
ModelSP259i ........................................... 19
ModelSP259 ........................................... 20
Probes............................................... 21
Soware technical requirements 22
Ordering information and customer support 22
Accessories and maintenance 22
Certifications and regulations 22
Safety information 24
Symbolsdefinitions ....................................... 24
Importantsafetynotes...................................... 24
Limited warranty & limitation of liability 25
Document Revisions 26
(c) Ikalogic SAS 2023 3/26

SP259 Series User Manual SP259 / SP259i
SP259 Series overview
SP259 series logic analyzers and protocol decoders oer in depth analysis of logic signals and protocols
with 250MHz (4ns) timing resolution. 9-channel operation allows 8-bit parallel data to be captured
along with a clock or strobe signal. SP259 series logic analyzers connect to a computer via a USB3
interface to configure the device and to visualize captured data.
SP259 Series is composed of two devices, SP259 and SP259i. Both oer 9 logic inputs, but
SP259i integrates industrial receivers for CAN, LIN, RS232 and RS485 buses.1
Figure 1: SP259 Logic Analyzer
Figure 2: SP259i logic analyzer (industrial version)
Embedded memory
SP259 series logic analyzers feature a 1Gb HyperRAM memory running at 4Gbps and is used to store
compressed samples before streaming to host computer. This embedded memory ensures uninter-
rupted captures of high frequency signals for long period of time without needing any reliance on host
computer performances.
1Industrial receivers in the SP259i are multiplexed with one or more of the existing 9 logic channels
(c) Ikalogic SAS 2023 4/26

SP259 Series User Manual SP259 / SP259i
Evolution from older SP209 series
SP259 are meant to replace and superceed older SP209 series of logic analyzers. SP259 devices have
all the features found in older SP209 devices, and have the following enhancements:
New SP259 Series Older SP209 series
250MHz maximum sampling frequency 200MHz maximum sampling frequency
Adjustable sampling frequency Sampling frequency fixed at 200MHz
USB3.0 interface USB2.0 interface
Fine adjustment of Threholds for input frontend Only 3 fixed thresholds
Typical applications
SP259 series logic analyzers are perfectly adapted for demanding applications, where it is needed to
capture logic signals with maximum time resolution on all 9 channels for long periodes of time.
•Embedded systems debugging
•Research and education
•Serial protocols analysis, like I2C, SPI, UART or 1-Wire (non exhaustive list)
•Industrial bus analysis: RS232, RS485, CAN, LIN
•ADC and DAC system diagnostic
(c) Ikalogic SAS 2023 5/26

SP259 Series User Manual SP259 / SP259i
Figure 3: Example of logic signals captured and analyzed by SP259 logic analyzer
Product highlights
•Schmitt trigger input stages with adjustable thresholds (fine tunable)
•250 MHz maximum sampling rate, with all 9 channels used.
•External clock option (state mode), up to 62.5MHz
•Precise trigger-In and trigger-Out signals on dedicated SMA ports
•USB3 interface for fast data visualization on host computer
(c) Ikalogic SAS 2023 6/26

SP259 Series User Manual SP259 / SP259i
SP259 system architecture
A functional SP259 measurement system consists of several main components that are presented in
the image below.
3
4
1
2
GND
9
8
7
6
5
4
3
2
1
Figure 4: SP259 system architecture
The main labeled components are:
1.
SP259 main host unit that contains the main memory buer and the processing power to capture
and compress signals before sending them over USB.
2. Type A to Micro USB3.0 Cable, 1 meter.
(c) Ikalogic SAS 2023 8/26

SP259 Series User Manual SP259 / SP259i
Main characteristics
Operating conditions
Temperature 10°C to 40°C
Relative humidity < 80% non condensing
Altitude < 2000m
Timing and measurements2
Sampling rate (MAX.) 250 MSPS
External clock Max rate (State mode) 62.5 MHz
Logic Inputs digital bandwidth 75 MHz
Embedded memory 1 Gb
Trigger output Yes, via SMA connector
External trigger input Yes, via SMA connector
Logic inputs specifications
Number of channels 9
Input impedance 200 KΩ|| 8 pF
Threshold circuits 3
Adjustable logic level (Fine tunable with 100mV
steps)
1.1 to 5V
Absolute max voltage on digital inputs
(Continuous or Transient)
±25 V
Power requirements
2HyperRAM sampling memory is used to buer samples before streaming to host computer.
(c) Ikalogic SAS 2023 10/26
This manual suits for next models