9
Always wear protective gloves. Handle
blood with care, as it may be infectious.
Followlocal safety procedures for disposal
ofused microcuvettes.
1. To perform a test, the cuvette holder
should be in its loading position. The
display will show three flashing dashes
and ’READY’. Remove a HemoCue Plasma/
Low Hb Microcuvette from the vial.
2. Place a drop of well mixed sample onto
a hydrophobic surface using a suitable
transfer device.
3. Fill the microcuvette in one continuous
process. Do NOT refill.
4. Wipe off excess sample from the outside
of the microcuvette with a clean, lint-free
wipe. Do not touch the filling end of the
microcuvette. If a second sample is to be
taken, fill a new microcuvette from a new
drop of sample. This should not be done
until the measurement of the first sample
is completed.
5. Visually inspect the microcuvette. If the
microcuvette is not completely filled with
sample or there are air bubbles, discard
and fill a new microcuvette. Small bubbles
around the edge can be ignored.