Greyline OCF 5.0 User manual

www.greyline.com
USER'S GUIDE
Installation & Operation
Instructions
Open Channel Flow Monitor
Model OCF 5.0
Manual Series A.1.4

Note: This page has been left blank intentionally.

Page 3
OCF 5.0 Open Channel Flow Monitor
INDEX
CONNECTIONS: ...............................................................................................4
KEYPAD SYSTEM............................................................................................6
CALIBRATION MENU.....................................................................................7
RUN....................................................................................................................8
MESSAGE..........................................................................................................8
STATUS .............................................................................................................8
PASSWORD.......................................................................................................9
UNITS/MODE..................................................................................................10
CALIBRATION................................................................................................11
CHANNEL SETUP ..........................................................................................13
RELAY PARAMETERS..................................................................................14
DATA LOGGING ............................................................................................15
SPECIAL FUNCTIONS...................................................................................17
SENSOR MOUNTING METHODS................................................................19
APPLICATIONS HOTLINE............................................................................26
PRODUCT RETURN PROCEDURE..............................................................27
APPENDIX A - OPTIONS...............................................................................29
CONVERSION GUIDE ...................................................................................33
SPECIFICATIONS...........................................................................................34
IMPORTANT NOTE: This instrument is manufactured and calibrated to meet product specifications.
Please read this manual carefully before installation and operation. Any unauthorized repairs or
modifications may result in a suspension of the warranty.
Available in Adobe Acrobat pdf format

Page 4
OCF 5.0 Open Channel Flow Monitor
CONNECTIONS:
POWER INPUT: The standard model requires AC power input between 100 to 240 VAC 50/60Hz. No
adjustments are necessary for voltages within this range.
Optional DC: 9-32 VDC. Connect to + and - terminals.
Optional Thermostat and Heater modules are available rated for 115 VAC or 230 VAC.
IMPORTANT NOTE: To comply with CSA/UL electrical safety standards, AC power input and relay
connection wires must have conduit entry to the instrument enclosure. Installation requires a switch,
overcurrent fuse or circuit breaker in the building (in close proximity to the equipment) that is marked
as the disconnect switch.
Risk of electric shock. Loosen cover screw to access connections. Only qualified personnel
should access connections.
Note: Use of instrumentation over 40°C ambient requires special field wiring.
Note: User replaceable fuse is 2 Amp 250V (T2AL250V).
!

Page 5
OCF 5.0 Open Channel Flow Monitor
CONNECTIONS
HEATER OPTION
AC
LN
NO
C
NC
NO
C
NC
–
+
4-20mA/0-5V
RLY2
RLY1
AC
GROUND
NC
C
NO
NC
C
NO
NC
C
NO
NC
C
NO
RLY3 RLY4 RLY5 RLY6
EXTRA RELAYS OPTION
POWER
INPUT
GND
SENSOR
DATA LOGGER
USB A
Plug

Page 6
OCF 5.0 Open Channel Flow Monitor
KEYPAD SYSTEM
The OCF 5.0 uses a menu system. Arrows show the four directions to leave a menu box. Press a key to
move to the next item in the direction shown. Move the cursor (underline) under numerals and increase
or decrease numerals with the and keys.
To store calibration values permanently (even through power interruptions), press .

Page 7
OCF 5.0 Open Channel Flow Monitor
CALIBRATION MENU
--Password----------
Password 0000
--Data Logging-------
Log Site ID 0
Mode Flow
Set Date Sep 22/2009
Set Time 14:24:40
Interval 30sec
Log NO
}
--Units/Mode--------
Mode Flow
Linear ft
Volume USG
Time min
Temperature C
}
--Calibration-------
Min Range 1.000ft
Max Range 12.000ft
Damping 10%
LOE Time 30sec
}
--Relay Parameters--
Relay 1
Function Flow
Mode Pump
On 10000 USG/m
Off 8000 USG/m
LOE mode Off
Alternate No
}
USG/m
Tot 20130.8 USG
Relays 1 2 3 4 5 6
0.000
--Message-----------
Data log Stopped
Log Used 0%
Maximum 199.45ft /s
Sensor Good
Temperature 21C
³
--Status------------
Range 1.00ft
Tot 0.000ft
EC 100%
Relays 1 2 3 4 5 6
}
³--Special Functions-
Language English
Analog Out 4-20mA
Backlight High
Reset Totalizer NO
Restore Defaults NO
New Password 0000
}
--Simulation--------
}Test Actual
Flow 250USG/m
4-20mA Flow 4.00
Relays 1 2 3 4 5 6
--Configuration-----
1.18.13
Sonar 1.05
Logger 1.16T
Relays 2
Analog Out 1
}
Utility
--Channel Setup------
Type V-notch
Size 22.5
}
°
--Menu Selections----
Units / Mode
Calibration
Channel Setup
Relay Parameters
Data Logging
Special Functions
Simulation
Configuration
}
--24 hr log----------
Date Feb. 12/2010
Total 50138 USG
Average 34.82 USG/m
Maximum 52.20 USG/m
Max Time 11:08:00
Minimum 0.000 USG/m
Min Time 9:15:00
}

Page 8
OCF 5.0 Open Channel Flow Monitor
RUN
The main display shows the units selected from the Units/Mode menu, Level,
Range, Flow, HRT, and Volume rate being measured, TOTALIZER and RELAY
states. The OCF 5.0 will start-up with this display and will return to this screen
after a timeout if keys are not pressed in other menus.
MESSAGE
Press from the RUN display to view status of the data logger and error/warning
messages provided by the instrument. The word Message will appear on the
RUN display if error messages are being generated by the instrument. Refer to the
manual section Error/Warning Messages for a description. Press to return to the
main display.
STATUS
Press from the RUN display to view instrument status. Range will be displayed
in linear units.
Tot Displays the current totalizer reading.
EC Displays echo confidence.
Relays Energized relays will display as a white character on a black
background.
24 HR LOG
Press from the RUN display to view a formatted flow report from instruments
with a built-in data logger. Press to scroll down one day or repeatedly to scroll
to a specific date. Up to 365 days can be stored. Newest date will overwrite the
oldest. Press to return to the main display.
USG/m
Tot 20130.8 USG
Relays 1 2 3 4 5 6
0.000
--Message-----------
Data log Stopped
Log Used 0%
Maximum 199.45ft /s
Sensor Good
Temperature 21C
³
--Status------------
Range 1.00ft
Tot 0.000ft
EC 100%
Relays123456
}
³
--24 hr log----------
Date Feb. 12/2010
Total 50138 USG
Average 34.82 USG/m
Maximum 52.20 USG/m
Max Time 11:08:00
Minimum 0.000 USG/m
Min Time 9:15:00
}

Page 9
OCF 5.0 Open Channel Flow Monitor
PASSWORD
The Password (a number from 0000 to 9999) prevents unauthorized access
to the Calibration menu.
From the Run display press to get to Password. Factory default password
is 0000 and if it has not been changed press the to proceed to the Menu
Selections screen.
If a password is required, press to place the cursor under the first digit and or
to set the number, then to the second digit, etc. Press or to proceed to
the Menu Selections screen.
A new password can be stored by going to Special Functions/New
Password.
--Password----------
Password 0000

Page 10
OCF 5.0 Open Channel Flow Monitor
UNITS/MODE
From Mode press the and then the or to select Level, Range,
Flow, HRT or Volume.
Range displays distance from the sensor to the target or liquid surface like a
tape measure. Range mode is useful to measure the exact distance
from the sensor to the zero level during calibration.
Level used to measure tank level in linear units, or "Head" in an open
channel for comparison with flow tables.
Volume displays tank inventory in engineering units.
HRT select Horizontal Round Tank mode to calculate and display volume
units in a horizontal round tank.
Flow calculates open channel flow through a flume or weir.
Volume,HRT or Flow gives the additional choice of volumetric units:
ft3 - cubic feet
USG - US gallons
USMG - US million gallons (FLOW only)
IG - Imperial gallons
IMG - Imperial million gallons (FLOW only)
m3 - cubic meters
L - liters
bbl - U.S. oil barrel
Press to store the selection, then to the next menu item and to enter.
Linear press and then or to select units of measurement.
Press to store the selection.
Time press and then or to select day, hr, min or
sec.
Temperature press and then or to select C or F (Centigrade or
Fahrenheit).
Press or to return to the Menu Selections screen.
--Units/Mode--------
Mode Flow
Linear ft
Volume USG
Time min
Temperature C
}
--Units/Mode--------
Mode Flow
Linear i
ft
m
mm
}n
--Units/Mode--------
Volume US
ft3
bbl
L
m3
IMG
IG
USMG
}G
--Units/Mode--------
Temperature C
F
}
Table of contents
Other Greyline Measuring Instrument manuals