fetatrack 360 User manual

fetatrack®360
fetal monitoring system
User Manua
l

fetatrack 360is issue 1 1
CONTENTS
INTRODUCTION 2
EQUIPMENT 3
SAFETY 4
SPECIFICATIONS 5
CONTROLS AND INDICATORS 7
OPERATING THE FETATRACK 360 9
PAPER LOADING 10
THE PRINTER 11
TRANSDUCER BELTS 12
FETAL HEART RATE MONITORING 13
DIRECT FETAL ECG 15
FETALSAT PULSE OXIMETRY 17
ALARMS 18
UTERINE ACTIVITY MONITORING 19
INTRAUTERINE PRESSURE MONITORING 20
FETAL ACTIVITY MONITORING 22
MATERNAL HEART RATE MONITORING 22
PREVENTATIVE MAINTENANCE 23
MONITOR CHECK-OUT PROCEDURE 24
SETTING UP THE FETATRACK 360 25
WARRANTY 26
GUIDELINES 28
NOTES 29

fetatrack 360is issue 1 2
INTRODUCTION
The FETATRACK 360 Cardiotocograph (available as an Antepartum or
Intrapartum monitor) is designed to provide cost effective care
throughout pregnancy. Internal or external transducers may be connected
to the FETATRACK 360. The functions described below are summarized
in the section on EQUIPMENT.
All units include twins ultrasound fetal heart rate measurement using dual
Doppler ultrasound transducers and external uterine pressure UA can be
monitored using the external (tocodynamometer).
The intrapartum monitor adds fetal ECG (FECG) and has an option of
including maternal ECG (MECG) measurement and an intrauterine
pressure transducer (IUP) for more accurate uterine pressure
measurement.
Both units can add an option for the monitoring of Maternal pulse
oximetry.
The FETATRACK 360is has the addition of the Nonin FetalSat fetal
pulse oximetry system.
All parameters are displayed on a large graphic displays which also
displays messages as appropriate. Ultrasound signal quality is shown by a
change in the flashing heart associated with the fetal rate display, so you
can be sure the recording is accurate and reliable.
The information is presented on a high resolution thermal printer running
at 1, 2 or 3 cm/min, and printed on standard format full width z-fold
paper. Event marks, date, time and mode are printed on the paper to
complete the record.
All FETATRACK 360s have bradycardia and tachycardia alarm
functionality, memory backup and an output socket for remote display of
screen information on a local computer, for example in the consultant's
office

fetatrack 360is issue 1 3
EQUIPMENT
The FETATRACK 360 is supplied with the following accessories for each version:
360is 360i 360a
Wide Angle Ultrasound
Transducer 2 2 2
FECG Leg Lead 1 1 n/a
Fetal Scalp Electrodes 2 2 n/a
External Uterine Pressure
Transducer 1 1 1
Internal Uterine Pressure
Transducer Option Option n/a
IUP Cable Option Option n/a
Maternal Pulse Oximetry Option Option Option
Maternal ECG Option Option n/a
FetalSat Pulse oximetry 1 N/a N/a
FetalSat Dual Sense ecg /
FSpO2 1 N/a N/a
Event Marker 1 1 1
Elastic Belt 3 3 3
Belt Buckles 3 3 3
Chart Paper 1 1 1
Ultrasound Coupling Gel
0.25ltr 1 1 1
AC line cord 1 1 1
User Instructions 1 1 1
REORDER INFORMATION
Part No. Item
0215-0003 Wide Angle Ultrasound Transducer (Blue)
M/1/0059 Wide Angle Ultrasound Transducer (Pink)
0211-0003 External Pressure Transducer
9800-0000 Internal Pressure Transducer
0261-0005 FECG Leg Lead LifeTrace Safelink
9021-0002 Transducer Belt Set
9260-0005 Chart Paper 20 BPM/cm
9260-0006 Chart Paper 30 BPM/cm
9000-0005 Coupling Gel 0.25L x 12 bottles
9088-0003 Fetal Scalp Electrode “Safelink”
FetalSat leg connector
FetalSat dual sense ecg / FspO2 sensor

SAFETY
The FETATRACK 360 is designed to comply with EN60601-1, Class 1
with protective earth via the AC line cord. Connection must be made to
ground and supply complying with local standards. For correct operation
check that the voltage setting and fuse ratings are correct for the local
supply.
The FETATRACK 360 is not explosion proof and must not be used in the
presence of flammable anaesthetics.
If in doubt about the correct operation of the FETATRACK 360, fetal
condition must be checked immediately by an alternative method.
We would recommend that, to maintain the standard of performance of
the FETATRACK 360, whenever possible the monitor is included in a
scheduled maintenance scheme.
The following symbols have been used on the front and rear panels of the
FETATRACK 360 and are here defined according to EN60101-1.
Alternating Current
Power On (Connection to the
mains)
Power Off (Disconnection
from the mains)
Type B Equipment
Type CF Equipment
Attention: consult
accompanying documents
fetatrack 360is issue 1 4

fetatrack 360is issue 1 5
SPECIFICATIONS
Size: 350 x 370 x 140 mm
Weight: 8 kg
Voltage: 100-240 Volts AC 50/60 Hz
Safety: Complies with EN60101-1
ULTRASOUND
Type: Continuous Doppler
Operating Frequency 2MHz±2% and 1.8MHz ±2%
Power Output 5 mW/sq cm max. SATA
Signal Processing Software AUTOCORRELATOR
DIRECT FECG (Intra models only)
Signal Range 40µV-5mV
Input Noise: <1,4µV p-p
CMRR: >125dB
ALARMS
Low Level: 60 - 120 BPM, Adjustable in steps of 5 BPM
High Level: 160 - 200 BPM, Adjustable in steps of 5 BPM
EXTERNAL UTERINE ACTIVITY
Type: Flat faced Tocodynamometer
Signal Range: 0 - 110 relative contractions strength
INTERNAL UTERINE ACTIVITY (Intra models only)
Intrauterine Pressure: Type optional
Signal Range 0 - 100 mmHg
EVENT MARKER
Hand held, person operated
PRINTER
Type: Thermal Array 8 dots/mm
Print Speed: 1,2 and 3 cm/min
Paper Size: 145 mm x 15 mtr
FHR Sensitivity: 20 or 30 BPM/cm
UA Sensitivity: 30/cm
IUP Sensitivity: 30mmHg/cm

fetatrack 360is issue 1 6
DISPLAY/PRINTER RANGES
FHR Range: 30 - 240 BPM
FHR Accuracy: +/- 1 BPM
UA Range: 0 - 110 units
FspO2 Range: 0 – 100%
OXIMETER
Oxygen Saturation 0% to 100%
(%FSpO2)
Pulse Rate Range 30 to 300 bpm
Accuracy
Blood Oxygen Saturation For specification, see footnote “a”
Pulse Rate 30 to 300 bpm; ±3 bpm
Displays
Graphical Displays 3-digit
Pulse Strength Graphic Bargraph
Measurement Wavelengths and Output Power
Red 661 nanometers @ 3 mw nominal
Infrared 905 nanometers @ 3 mw nominal
“a”. Calibration of the FetalSAT System was accomplished through carefully
controlled hypoxia studies in a laboratory animal model to establish the calibration
coefficients. For specifications, see FetalSAT Specification Supplement (available
upon request).

CONTROLS AND INDICATORS
FRONT PANEL
Set Uterine Pressure
to Zero
Audio Selector for
Ultrasound 1 & 2
Printer On/Off &
Speed Control
Ultrasound Volume
Decrease
Ultrasound Volume
Increase
Maternal & Fetal
Electrocardio-graph &
Ultrasound 2 Transducers
Input
Internal & External Uterine
Pressure Transducers Input
Ultrasound 1
Transducer Input
Socket
Power On
Indicator
Printer Door
Release Lever
fetatrack 360is issue 1 7

REAR PANEL
Mains On/Off
Power Inlet
Fuse
Compartment
Network or Local
Computer Port
Pulse Oximeter
Input
Event
Marker
fetatrack 360is issue 1 8

fetatrack 360is issue 1 9
OPERATING THE FETATRACK 360
The FETATRACK 360 comes to you with a multi-range power supply which can be
switched externally to suite the local AC line voltage. It will operate with an input
voltage between 100 and 240 volts AC, 50 or 60 Hz.
Connect the AC supply cord and switch the FETATRACK 360 on. The AC input
on/off switch is located on the rear of the unit adjacent to the AC line input socket.
Observe that the front panel power on indicator (1/0) is illuminated.
The FETATRACK 360 is now operational and the display will indicate the
operating mode.