
the screws that hold the sensor (the CO sensor contains sulfuric acid,
exercise caution). Attach a new sensor to the housing. Remove the short
between the sensor terminals and connect the wires to the sensor, observing
the correct color code. Carefully, close the case, making sure that no wires
or tubing are being pinched. Secure the case with the four screws.
7. INSTRUMENT CALIBRATION
The Model 60 Carbon Monoxide Analyzer has been calibrated at the factory by using
a calibrated gas that contains 200 ppm carbon monoxide in air (range switch in the
lower position). (However, any other calibration gas in the range of 50-2000 ppm CO
in air or nitrogen can be used, if available. Higher concentration may be used with the
range switch set in its upper position ). The sensor possesses good stability and does
not require frequent calibration.
A tentative recommended calibration schedule is as follows:
If the unit is used in engine exhausts, calibrate after 10- 20 hours of use. If the unit is
used in boilers calibrate every one or two weeks. If the unit is used for area
monitoring (safety) calibrate every 3-4 months.
If it is desired to check the calibration of the instrument, connect the probe to the
instrument and turn on the instrument. Turn on the pump. Wait two minutes and
carefully zero the instrument, making sure the air the instrument draws is clean and
free of any smoke or exhausts.
If you are using the ENERAC calibration kit follow the directions in the calibration
kit manual.
If you are using your own span gas, feed the calibrated gas to the stack probe at a rate
of approximately 1000-1500 cc/min. Use a flow meter to observe the flow into the
instrument. Listen to the sound of the pump. Adjust the flow rate between 1000-1500
cc/min. so that the sound of the pump appears normal as when it was aspirating air.
Wait till the CO reading of the display has stabilized. If you are using high
concentration span gas, make sure the area is well ventilated, or carry out the
calibration under a fume hood, since carbon monoxide gas is highly toxic. Remove the
small round black plastic cap that covers the span adjustment potentiometer. With a
small screwdriver, rotate the span adjustment potentiometer until the display indicates
the correct reading. Replace the black plastic cap. Shut off the gas, disconnect the