8
300sec. 4.0mA 550hrs. 1,500hrs. 450hrs.
670sec. 2.2mA 1,005 hrs. 2,740hrs. 1,005hrs.
1,800sec. 1.3mA 1,754 hrs. 4,782 hrs. 2,700hrs.
3,600sec. <1mA 2,251 hrs. 6,139 hrs. 5,400hrs.
*ContactusifyouneedtocalculateD,EBL,orTTFforotherloggingintervals(T).
Asshowninthetableabove,inordertocompletelyfilltheinternalmemoryatloggingintervals
equaltoapproximately670secondsorlonger,ahighcapacity(6,000mAh)Li‐Ionbatterywillbe
requiredifmeterisnotconnectedtopoweradapter,assumingmeterisfairlynewandbatterywas
fullychargedinitially.
IMPORTANT:DOWNLOADTHEDATALOGSREGULARLYTOPREVENTPOSSIBLEDATALOSS.ONCETHE
MEMORYISFULL,PREVIOUSLYLOGGEDDATAWILLBEOVERWRITTEN.
NOTE:TheAAbattery‐poweredMeter(ESCM‐18-CC)willhaveanestimatedbatterylifesimilarto
thatofits2,200mAhLi‐Ionbattery‐poweredcounterpart(ESCM‐18).
ChargingtheMeter
MeterspoweredbyrechargeableLi‐Ionbatteriescanbechargedusingthesupplieduniversal
poweradapteroraUSBportonapoweredcomputer.Li‐Ionbatterieswillbefullychargein5‐8
hoursandwillloseabout8%oftheirchargingcapacitypereveryyearofuseat25°C.
ProductOverview
Thisdataloggerisdesignedtomeasurecarbondioxide(CO2),percentofrelativehumidity(%RH),
andtemperaturethroughdiffusionsampling.Itstorescollecteddatatoitsinternalmemorytobe
retrievedatalatertimeforanalysis.WhencombinedwithourGasLab®software,youcanalsosee
real‐timedataonyourcomputer’sscreen.
Scientificdevicessuchasthisdataloggerrequiretheusertohaveanintimateknowledgeofthe
meter,itsoperation,therequiredsoftware,andthemeterspecificationspriortouse.Wehighly
recommendsreadingthisuser’smanualbeforeoperatingthedevice,especiallytheImportant
Safeguardssectiononpage4.
TheoryofOperation
TheCO2sensorinsidethismeterusesNDIR(non‐dispersiveinfrared)technologytosense,asa
functionoftransmittedlight,theconcentrationofCO2intheair.Ithasbeenfactorycalibratedto
operatewithinthespecifiedrangeandprecision.