Detecting Moisture in Insulation:
Use the BD-10 withthe #21-E Electrode and#608insulated pins to detect moisture in insulation.These pinsare insulated,
except at the tips.Only the uninsulatedtips makecontact withthe material,providinginformationas tothedepth at whichmoisture
is present.
Procedure:
Drill two 4.8mm (3∕16”) diameterholes, one inchapart through the siding.
Using the #21-E, insertthe#608 contact pins into the holes so thatyou make contactwith the insulation.
Press the read button andread the reference scale .Take readings atvarious depths ofpenetrationto
determine the location of moisture.
Fill holes with putty at the conclusion ofthe test.
Meters are not calibrated toread the percentageof moisture contentin theinsulation.Instead,meterreadings provide a qualitative
indication ofthe presence of moisture.The metermayalso help identifythe pattern of distribution to helpyou determine ifmoisture
is due to condensationorleakage.
The presence ofmoisturein insulating materialgreatlyreduces its insulatingproperties.Ifthe insulatingmaterial absorbsmoisture,
water willdisplacetheairinthe material.
Water also transmits heatata fasterrate than air,thus reducingtheefficiency of the insulation.
Sidewalls androof insulation pickup moisturebecause ofleaks inthe roof or structuraldefects in the sidewalls whichallowstorm
water to be forcedin duringhigh wind.
Duringperiods ofcoldweather, vapor will condenseas it approaches the colderareanearthe outside walls,and resultingmoisture
willremain trappedinthe insulation.
Cork, woodfiber boards,andothercellulosematerialread lowerthanglass woolandotherinorganictypes ofinsulation.
CARE OF YOUR METER
To keep your meterin good workingorder:
Store your meterina clean, dry place.The protective carrying caseprovidedis anidealstorageplacewhen the meteris
not in use.
Change the9-Volt battery as needed.Continuedusewitha low battery maycausethemeterto goout of calibration.
Change contactpins as needed.Keeppin retainers hand tightened.
Cleanthe meterandcontact pins withany biodegradable cleaner.Usethecleaner sparinglyandon external parts only.
Keep cleanerout of the external connector .
Removethe battery if the meterwill not beusedfor one monthor longer.
SERVICE FOR YOUR METER
Pack your metersecurely.Enclose apurchaseorderorletter with a brief descriptionofthe problem.
There is no need to call us for a returnauthorizationnumberif you are withinthe EU. Customers outside the EU mustcontact
us for morespecificinstructionspriortoreturninga meter.
Include your name,address,daytimephoneandfax numbers ore-mail address.If you believe the meteris underwarranty,
pleaseprovidetheoriginalsales sliporinvoice.
Shipvia UPS, Express Mail, PriorityMail or any overnightcourierwho provides promptservice.Donot usestandardparcel
post.
Insureyour instrumentforits full value and shipprepaid.Weare not responsiblefordamage intransit.
We do not accept COD shipments orcover anyincoming freightorduty charges onreturned merchandise
Turnaround timeonrepairs is approximatelytwoweeks.
We willcall you with an estimateifyou specificallyrequestone,orif we determinethatthe metermaybe too costlyto repair.
Non-warrantyrepairs willbereturned via UPS/COD unless youhave alreadyestablishedotherpaymentterms.Thereis no
COD serviceoutsidetheEU.
Payments have to be madeby Banktransferprior to the return shipment.Aproformainvoice willbe raised inadvance.
Warranty repairs willbereturned at no chargeif shippedwithinthe EU via GLS Ground Service. Freight charges forexpedited
services (i.e., Federal Express, UPS/2 Day, UPS/1 Day, etc.) are the customer’s responsibility and will be charged as per the
above terms.